Using the depicted button function, the head or foot low posi-
tion can be adjusted.
Caution ! When using the rotation function,
ensure the safety of the patient! No foreign
bodies must be in the movement area of the
lifting mechanism. These functions may only
be used by professional personnel and must be
locked on the back of the remote by the profes-
sional personnel. Maladjustment may lead to
permanent damage to the resident.
Blocking function on manual
control unit
The electrical equipment is based on state-of-the-art technolo-
gy and first failure safety.
The blocking function provides an additional safety measure.
This function is located on the rear side of the hand control unit
and can be operated by personnel using a key switch. In the
event of a malfunction of the electric drive, the functions can
be stopped using the appropriate rotary switch.
20
|
(Operating buttons Trend/Anti-Trend)
Operation:
The operation of the lock function is further described on page
14/15 (hand switch symbols).

Braking and transporting

Evolution 100
The lying surface is in the lowest positi-
on. Hold down switch S8 on the hand
control unit and at the same time
press the roller button on the hand
control unit to release the castors.
(Operating buttons release castors)
The Evolution 100 model is now ready for transport. Use the S7
button on the hand control unit – height adjustment of lying
surface to retract the castors. The bed is now resting on the
supporting feet.
Caution ! On the Evolution 100 model (trans-
portable in lowest position), the S8 hand
control switch has a double function (only ICS).
(Operating buttons retract castors )