Triumph IP12 Infusion
Pump Quick Start Guide
This guide provides a quick
reference on operating the
Triumph IP12 Infusion Pump.
Please refer to the Triumph IP12
Infusion Pump user manual @
WWW.TRIUMPHMED.COM
for
detailed operation and settings
.

Advertisement

loading

Summary of Contents for Triumph IP12

  • Page 1 Triumph IP12 Infusion Pump Quick Start Guide This guide provides a quick reference on operating the Triumph IP12 Infusion Pump. Please refer to the Triumph IP12 Infusion Pump user manual @ WWW.TRIUMPHMED.COM detailed operation and settings...
  • Page 2 Install Pole/Cage Clamp on back of pump with provided hex wrench and bolt. Ability to install vertically or horizontally. (See Figure A and B) If attaching to cage it is necessary to install vertically with “split” side of clamp on the top. Also necessary to have clamp bolt all the way out (see figure C) Figure A Figure B...
  • Page 3: Basic Operation

    Press and Hold Power Button for 2 Seconds to Power on Select “Yes” or “No” to Clear last treatment Open door by pulling door handle; lift free flow clamp (see figure 1 and 2); install IV Admin Set following flow direction; secure line all the way across line guide by pressing line in to both air sensor holders (see figure 3);...
  • Page 4 Current Drop Occlusion Touchscreen Battery Infusion IV Brand/Alarms Indicator Pressure Lock/unlock Indicator Rate Indicator Start/Stop Infusion During Programmable Alarm Bolus Indicator Infusion AC Power Manual Bolus Indicator Interface Power Button Settings VTBI Volume to Be Infused Time Left for Infusion Volume Infused Completion...
  • Page 5 To Access Modes Press The Bottom of Home Screen or Select Menu then Modes Rate Mode: Set VTBI and Rate (ml/hr) VTBI Range : 0.01 - 9999.99 ml Rate Range : 0.1 ml/hr – 1500 ml/hr Time Mode: Set VTBI and Time VTBI Range: 0.01 - 9999.9 ml Time Range: 0 –...
  • Page 6 Purge: While pump is idle Press Purge on screen and select Yes or Press and Hold the Purge Softkey. (see Figure 1) Programmed Bolus: During infusion Press Bolus on screen and set Bolus VTBI, Bolus rate, Bolus time, and then select start (see Figure 2 and Figure 3) Manual Bolus: During infusion press and hold Bolus softkey (see Figure 2) - Audible tone every 1 ml bolus infused...
  • Page 7 To Enter additional Settings Select the Menu button on Touchscreen or Softkey Modes: Change current Infusion Mode Settings: Change current IV admin Brand, drug library, default KVO(Keep Vein Open) Rate (0.01 ml/hr – 5.0 ml/hr), default Bolus rate, Occlusion Pressure (3 Levels 300 mmhg – 900 mmhg), Bubble sensor size (7 Levels 50 ul –...