Download Print this page

Triumph IP12 User Manual

Infusion pump
Hide thumbs Also See for IP12:

Advertisement

User Manual
Triumph IP12 Infusion Pump

Advertisement

loading

Summary of Contents for Triumph IP12

  • Page 1 User Manual Triumph IP12 Infusion Pump...
  • Page 2: Table Of Contents

    Contents 1. Safety.……………..………………………………3-4 2. Specifications……………………………………...5-6 3. Appearance………………………………………7-11 4. Installation……………………………………...12-13 5. Basic Operation………………………………..14 6. Infusion Operation……………………………..15-19 7. Settings………………………………………….20-25 8. Alarms/Troubleshooting…………………….…26-27 9. Maintenance/Repair/Warranty…………..…..28-31...
  • Page 3 If there’s anything abnormal, immediately stop working and contact Triumph Medical’s service department. Additionally, the adhesion or intrusion of fluid/drug may possibly cause the equipment to fault and malfunction. Therefore, please clean the equipment after use, and store it correctly.
  • Page 4 It is not allowed to disassemble, refit, or use this equipment for other purposes except normal infusion.  No one is allowed to repair this equipment except Triumph Medical Services LLC.  To avoid risk of electric shock,the equipment must only be connected to a supply mains with protectiv...
  • Page 5 2. Product Specification Safety Classification Electric protection Class I Type Electric protection Defibrillation proof type CF applied Part Level IP34 (Prevent solid objects larger than 2.5mm in diameter and water intrusion Ingress Protection from splashing in all directions) Working mode Continuous Classification Portable equipment, non-portable infusion pump...
  • Page 6: Safety

    DC power supply DC 12V Specification: 7.4V 2500mAh Charging time: 5h.(under OFF status) Battery Working time: over 8h.(after completely charging the new battery, when the Specifications environment temperature is 25℃ and flow rate is 25ml/h, the constantly working time) Alarm When the sound is set at lowest level, alarm signal sound pressure level Alarm signal sound ≥50dB(A)
  • Page 7: Appearance

    3. Appearance Front View ① Tubing guide ② Pump door ④⑥⑩ Pressure Plate ③Pressure sensor-UPSTREAM (if applicable) ⑦ Pump tablets ⑨ Pressure sensor-DOWNSTREAM ⑤⑧Air–in-line sensor ⑪Anti-free flow clamp...
  • Page 8 Operation Panel ① Touch Screen ②【Power】 Pump power switch, press and hold for 3 seconds, pump power off. Standby selection button. ③ AC indicator light When connecting with AC power supply, AC indicator lights on. ④ Alarm indicator While pump alarms, indicator light glitter, different level different frequency and color, more information please refer to Chapter 10.1 ⑤...
  • Page 9 Display Screen The display screen interface layout composes of title bar and typical interface. Title Bar The Title bar displays real-time information and is not touchable, the left upper corner displays the name of current editing parameter. Icon Paraphrase Description Infusion line Infusion line indication icon...
  • Page 10 Typical Interface Icon Paraphrase Table3.3.2.1-1 Icon Paraphrase Description Start Click this icon, start infusion Stop Click this icon, infusion stop 1. During infusion, it is『Bolus』function, click it to start fast infusion Bolus/Purge 2. Before infusion starting, it is『Purge』function, click it to exhaust air from the IV set Home Click this icon, return to the main interface...
  • Page 11 Rear View ① USB Port Port for software upgrade and drop sensor ② DC Input Port External 12V DC power supply ③ Handle ④ A/C Adapter Port External 100-240V 50/60Hz AC power supply ⑤ Loudspeaker ⑥ Pole Clamp Using for fixing the equipment on the infusion stand ⑦...
  • Page 12: Installation

    IEC60601-1-1 system. The technician who takes charge of connecting to additional devices with the equipment interface is responsible for meeting the IEC60601-1-1 standard. Please contact Triumph Medical if you have any enquiry.
  • Page 13 Installing the Infusion Pump and Pole Clamp Install the pole clamp/cage mount using provided hex wrench and screw. (2) If using as a cage mount then leave rotary knob off of pole clamp and install the mount vertically as shown below leaving the “split” side of pole clamp at the top.
  • Page 14: Basic Operation

    5. Basic Operation Operation Flow ¤ Mount the infusion pump on the IV stand: ¤ Power on: press for two seconds, Power on equipment ¤ Install IV Set ¤ Confirm IV set brand name: Select infusion tube brand ¤ Remove air bubble from the line ¤...
  • Page 15: Infusion Operation

    CPU communication, alarm indicator. After passing the self-test, pump will enter into the rate mode interface. Warning: ● If self-test failed, the pump cannot be used, please contact Triumph Medical. Infusion Line Installation Connect the IV line with the infusion bottle/bag.
  • Page 16 ● Please confirm that the infusion line brand and specification displayed in the display screen is accordant with the actual one in use. ● If infusion accuracy is in question, the user is strongly suggested to contact Triumph Medical. 9) Install Drop sensor Please install.
  • Page 17 Changing the Rate During infusion Under the running interface, click rate number in touch screen and reset target infusion rate. Or 【Stop】 and enter the parameters setting interface, then reset target infusion rate. Note: ●Only the rate mode, drip mode and Body weight mode support online rate modification function during infusion without pressing【Stop】...
  • Page 18 Stop Infusion During infusion or after infusion, click , infusion stop. It will return to the parameter setting interface display Total Volume Infused and adjustable parameters. Remove the Infusion Line Disconnect the IV set’s extension line from the patient, then remove the IV set. Replace IV set Power OFF or Standby 【Power】Button till the screen is OFF, the equipment is OFF.
  • Page 19 Set Infusion Parameters Introduction to Infusion Parameters Setting (1) The drug information can be displayed in the infusion running interface only when the drug library is under active state. (2) Click『Settings』 icon in the main interface to enter sub-menu, find 『Drug Library』 menu item, click to enter then select drug.
  • Page 20 7. Infusion Mode Setting After starting the equipment and self-test, the equipment automatically enters into the rate mode parameters setting interface. To select the other modes, click『Menu』icon to enter into the main interface. Click『Modes』icon to enter into the mode selection menu interface, and select preset infusion mode.
  • Page 21: Settings

    Drip mode Under this mode, set the VTBI and drop rate, and the system will automatically calculate the infusion flow rate and time. Drip Sensor not included. Note: ● The flow rate under drip mode is calculated according to the specification of the current infusion line.
  • Page 22 IV Admin Brands For the built-in infusion line brand of the system, after installing the infusion line, click 『IV Admin Brands』to enter into the infusion line brand selecting interface, and click the preset brand option. ● The infusion line of different brands may possibly cause flow rate deviation. When using, please confirm if the displayed information in the interface is accordant with the actual infusion line in use.
  • Page 23 Unit Mark Unit Conversion 1 kPa=7.5mmHg=0.145psi=0.01bar 1psi=51.713mmHg=6.895kpa=0.069bar 1bar=787.5mmHg=15.225psi=105kPa Bubbles Size Click『Bubbles size』 to enter into air bubble size setting interface, move the long box to the preset level, confirm and then click『OK』. The air bubble detector has 7 levels, when the volume of single air bubble or the total air bubbles within 15min in the line reach the preset air bubble alarm threshold value, it will activate air bubble alarm.
  • Page 24 Note: ● The default state for drop sensor function system is OFF. It can be manually turned on by the user when the drop sensor should be adopted. If the function is ON when the drop sensor is not installed, then the system will report “Drop sensor connection” alarm. Drop sensor level The sensitivity of the drop sensor can be adjusted in three levels.
  • Page 25 operated within 3min, the system will automatically adjust the brightness to the lowest level. When it is touched or button is clicked by user or when there’s an alarm, it will automatically recover the brightness. Sound Click『Sound』 to enter into the sound parameters setting interface, the volume has 10 levels. The lowest volume is off, and the highest volume is ≤80 dB.
  • Page 26 8. Alarm Prompt and Troubleshooting Introduction to Alarm Level During infusion preparation and infusion, this equipment will alarm when reaching or exceeding the set alarm threshold value and prompt with sound, light and text. According to the importance of alarm information as well as the emergency and safety, the alarm is divided into three levels: high, middle and low.
  • Page 27 Please eliminate the fault alarm according to the prompt. If it can’t be eliminated, please stop the equipment, contact Triumph Medical to repair and test the equipment. Do not put it into operation before the equipment has passed the inspection, otherwise, it may possibly cause unpredictable harm if it continues to operate under fault contition.
  • Page 28 9. Maintenance If pump displays the following: “Scheduled Maintenance Recommended Contact Triumph Medical” this means your pump is do for its yearly calibration. Please proceed to Triumphmed.com and fill out the Veterinary RMA form located at the top of the home page.
  • Page 29 Warning: ● Please do not adopt Cidex OPA orthophthalaldehyde, methyl ethyl ketone or similar solvent, otherwise, it may corrode the equipment. Repair Please contact Triumph Medical (888-388-3344 ext 3) to repair if there’s any fault or damage. Do not disassemble and repair the equipment. Alarm and Solution...
  • Page 30 No battery Battery is removed external power supply, inserted contact Triumph Medical Under ON state, AC power supply alarm automatically No power supply is adopted, but the AC power wire eliminates after connecting is dropped during the process the external power supply.
  • Page 31 Triumph Medical Note: When alarm rings, click the 『Mute』 icon on the screen to temporarily stop sound alarm for 2min. Warranty Warranty for both parts and labor is 12 Months from date of purchase. See extended warranty card...