Download Print this page
Triumph SP12 Quick Start Manual

Triumph SP12 Quick Start Manual

Syringe pump
Hide thumbs Also See for SP12:

Advertisement

Quick Links

Triumph SP12 Syringe
Pump Quick Start Guide
This guide provides a quick
reference on operating the
Triumph SP12 Syringe Pump.
Please refer to the Triumph SP12
Syringe Pump user manual @
WWW.TRIUMPHMED.COM
for
detailed operation and settings
.

Advertisement

loading

Summary of Contents for Triumph SP12

  • Page 1 Triumph SP12 Syringe Pump Quick Start Guide This guide provides a quick reference on operating the Triumph SP12 Syringe Pump. Please refer to the Triumph SP12 Syringe Pump user manual @ WWW.TRIUMPHMED.COM detailed operation and settings...
  • Page 2 Install Pole/Cage Clamp on back of pump with provided hex wrench and bolt. Ability to install vertically or horizontally. (See Figure A and B) If attaching to cage it is necessary to install vertically with “split” side of clamp on the top. Also necessary to have clamp handle all the way out (see figure C) Figure A Figure B...
  • Page 3: Basic Operation

    Press and Hold Power Button for 2 Seconds to Power on Select “Yes” or “No” to Clear last treatment. Selecting “Yes “will clear all infusion parameters from previous infusion. If no selection is made within 30 seconds all parameters are cleared automatically Pull out barrel clamp and rotate either direction.
  • Page 4 Occlusion Current Touchscreen Syringe Pressure Infusion Lock/Unlock Battery Size Syringe Brand/Alarm Indicator Rate Indicator During Start/Stop Infusion Infusion Alarm Indicator Programmable Bolus AC Power Indicator Manual Bolus Interface Power Button Settings VTBI Volume to Be Infused Time Left for Infusion Volume Infused Completion...
  • Page 5 To Access Modes Press The Bottom of Home Screen or Select Menu then Modes Syringe Size Rate Range ml/hr 0.1-100 Rate Mode: Set VTBI and Rate (ml/hr) 5/6ml 0.1-150 VTBI Range : 0.01 - 9999.99 ml 10/12ml 0.1-300 Time Mode: Set VTBI and Time 20ml 0.1-600 30/35ml...
  • Page 6 Purge: While pump is idle Press Purge on screen and select Yes or Press and Hold the Purge Softkey. (see Figure 1) Programmed Bolus: During infusion Press Bolus on screen and set Bolus VTBI, Bolus rate, Bolus time, and then select start (see Figure 2 and Figure 3) Manual Bolus: During infusion press and hold Bolus softkey (see Figure 2) - Audible tone every 1 ml bolus infused...
  • Page 7 To Enter additional Settings Select the Menu button on Touchscreen or Softkey while pump is idle Modes: Change current Infusion Mode Settings: Change current Syringe Brand, drug library, default KVO Rate (0.01 ml/hr – 5.0 ml/hr), default Manual Bolus rate, Occlusion Pressure (3 Levels 300 mmhg –...