If there’s anything abnormal, immediately stop using the equipment and contact Triumph Medical Service Department. Additionally, the adhesion or intrusion of fluid/drug may possibly cause the equipment fault and malfunction. Therefore, please clean the equipment after use, and store it correctly.
Page 4
It is not allowed to disassemble or refit this equipment or use it for other purposes except normal infusion. No one is allowed to repair this equipment except Triumph Medical Services. To avoid risk of electric shock, this equipment must only be connected to a supply mains with protecti...
2. Specifications Safety Classification Electric protection Class I Type Electric protection Defibrillation proof type CF applied Part Level IP34 (Prevent solid objects larger than 2.5mm in diameter and water intrusion Ingress Protection from splashing in all directions) Working mode Continuous Classification Portable equipment, non-portable syringe pump Specification Parameters...
Page 6
Time Range 1min-99hrs59min Fuse Type slow fuse 2A 250V Dimensions 242.5(W)*111(D)*126.5(H) mm Weight 1.7kg Power Supply AC power supply 100-240V 50/60Hz Input power 50VA DC power supply DC 12V Specification: 7.4V 2500mAh Charging time: 5h. (under OFF status) Battery Working time: over 9h. (after completely charging the new battery, when the Specifications environment temperature is 25℃...
Page 7
Safety Standard IEC 60601-1:2005+A1:2012 Medical Electrical Equipment, Part 1: General Requirements for basic safety and essential performance IEC60601-2-24:2012 Medical electrical equipment – Part 2-24: Particular requirements for the safety of syringe pumps and controllers IEC60601-1-8:2006 +A1:2012 Main Safety Medical electrical equipment –Part 1-8: General requirements for basic safety Standards and essential performance –Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrica...
3. Appearance FRONT VIEW ① Handle Control syringe pump push-pull sliding box and clip. ② Slider box ③ Pressure sensor Detect the pressure of the syringe ④ Syringe clip Clamp the syringe plunger ⑤ Lead-screw ⑥ Syringe fixture lever Pull forward then turn right, install the syringe into the slot. ⑦...
Page 9
OPERATION PANEL ① Touch Screen ② 【Power】 Pump power switch, press and hold for 3 seconds to turn pump on, pump power off, and standby selection button. ③ AC indicator light When connecting with AC power supply, AC indicator lights on. ④...
Page 10
Display Screen The display screen interface layout composes of title bar and typical interface. Title Bar Title bar displays real-time information and is not touchable, the left upper corner displays the name of current editing parameter. Title Bar Icon Icon Paraphrase Description Syringe apparatus...
Page 11
Typical Interface Icon Paraphrase Icon Paraphrase Description Start Click this icon, start infusion Stop Click this icon, infusion stop 1. During infusion, it is a 『Bolus』 function, click it to start bolus infusion Bolus/Purge 2. Before starting infusion, it is a『Purge』function, click it to exhaust air from the syringe Home Click this icon, return to the main interface...
Page 12
Rear View ① USB Port Port only for software upgrade ② DC Input Port External 12V DC power supply ③ Handle ④ A/C Adapter Port External 100-240V 50/60Hz AC power supply ⑤ Pole Clamp Using for fixing the equipment to an infusion stand ⑥...
All devices must be connected according to the valid version of IEC60601-1-1 system. The technician who takes charge of connecting to additional devices with the equipment interface is responsible for meeting the IEC60601-1-1 standard. Please contact Triumph Medical if you have any question.
Page 14
Install the syringe Pump Pole Clamp (1) Install pole clamp/cage mount using provide hex wrench and screw. (2) If using as a cage mount then leave rotary knob off of pole clamp and install the mount vertically as shown below leaving the “split” side of pole clamp at the top.
5.Basic Operation Operation Flow 1. ¤ Mount the syringe pump on the IV Pole or Cage 2. ¤ Power on: press two seconds 3. ¤Install syringe 4. ¤ Confirm syringe brand and size: or select syringe brand 5. ¤Remove air bubble in the line 6.
After passing the self-test, pump enters into rate mode interface. Warning: ● If self-test failed, pump cannot operate properly and cannot be used for patient infusion, please contact Triumph Medical. Install Syringe (1) Hold the clutch and pull the slider to the right side.
Page 18
Set Infusion Parameters Remove Air bubble Under the parameters setting interface, Press 【Bolus】button and hold on, or touch the purge icon on the product panel, eliminate the air bubble in the line. The purge total volume is not calculated in the Total Volume Infused. Cautions: ●...
Page 19
Bolus Application In operation, Bolus functions have two operation modes: Manual bolus and Automatic bolus. (1) Manual bolus: press and hold the 【Bolus】 button on the product panel. Pump will work at the max flow rate of current syringe size, or set max bolus rate under setting interface. Release the button, pump will go back to the previous set infusion rate.
Page 20
Power OFF or Standby 【Power】Button till the screen is OFF. Method 1: hold the 【Power】Button to enter into OFF interface. Method 2: press the ⑴ Turn off the equipment: click 『Power off』icon, the equipment is turned OFF. ⑵ Standby: click 『Standby』 icon to enter into standby time setting interface and set the standby time.
Page 21
Set Infusion Parameters Introduction to Infusion Parameters Setting (1) The drug information can be displayed in the infusion running interface only when the drug library is under active state. (2) Click『Settings』 icon in the main interface to enter sub-menu, find 『Drug Library』 menu item, click to enter then select drug.
Page 22
Infusion Mode Setting After starting the equipment and self-test, the equipment automatically enters into the rate mode parameters setting interface. To select other modes, click『Menu』icon to enter into the main interface. Click『Modes』icon to enter into the mode selection menu interface, and select preset infusion mode.
7. System Settings Settings Click『Settings』icon in the main interface to enter into parameters setting interface. Drug Library Click on the preset drug name, the selected drug will be reflected in infusion mode parameters. To remove drug name from infusion mode parameters, select “none” in drug library. To Edit drugs, go to Settings –...
Page 24
Warning: When adopting fluid/drug of a high viscosity and the occlusion pressure is set at low level, it is possible that the system will report occlusion alarm even when the line is not obstructed. Under this condition, please carefully observe the pressure indication icon in the display screen and infusion line, and raise the occlusion pressure if needed.
Page 25
Finish Pre-alarm Time for pre-alarm refers to the time of activating near completion alarm when the fluid/drug infused volume is nearly reaching the preset value. Click 『Finish pre-alarm』to enter into the time for pre-alarm setting interface. Click the preset time option to set the finish pre-alarm time. Micro Mode Click 『Micro mode』...
Page 26
Brightness Click『Brightness』to enter into display brightness setting interface. The brightness has 10 levels. The equipment has the function of automatic brightness adjustment if external power supply is unavailable. When there is no external power supply, and the power is supplied by battery, if it is not operated within 3min, the system will automatically adjust the brightness to the lowest level.
Page 27
Last Therapies Click 『Records』 then 『Last therapies』 to enter last therapies interface. System will store 20 of the previous most recent therapies to select. 8. Alarm Prompt and Troubleshooting Introduction to Alarm Levels During infusion preparation and infusion, this equipment will alarm when reaching or exceeding the set alarm threshold value and prompt with sound, light and text.
Page 28
Please eliminate the fault alarm according to the prompt. If it can’t be eliminated, please stop the equipment, contact Triumph Medical to repair and test the equipment. Do not put it into operation before the equipment has passed the inspection, otherwise, it may possibly cause unpredictable harm if it works with a fault.
Page 29
9. Maintenance If pump displays the following: “Scheduled Maintenance Recommended Contact Triumph Medical” this means your pump is do for its yearly calibration. Please proceed to Triumphmed.com and fill out the Veterinary RMA form located at the top of the home page.
Page 30
Warning: ● Please do not adopt Cidex OPA orthophthalaldehyde, methyl ethyl ketone or similar solvent, otherwise, it may corrode the equipment. Repair Please contact Triumph Medical (888-388-3344 ext 3) to repair if there’s any fault. Do not disassemble and repair the equipment. Alarm and Solution...
Page 31
【 Start 】 button to continue system occlusion level is set too infusion 3. The pressure sensor is damaged Please contact Triumph Medical 1. When power is supplied only alarm automatically with the built-in battery, under low...
Page 32
System Error High alarm still exists, please contact exception Triumph Medical Note: When alarm rings, click the 『Mute』 icon on the screen to temporarily stop sound alarm for 2min. Warranty Warranty for both parts and labor is 12 Months from date of purchase. See extended warranty card...