  • EN

Available languages

  • ENGLISH, page 12
5. Select "Test functions >> Readjustment with
test gas".
? The "CO" dialog window is opened.
i
Numerical values must be entered separated by
a dot.
6. Enter the sample gas value for CO and confirm
with <Ok>.
? The "CO2" dialog window is opened.
Robert Bosch GmbH
Start-up | 1 687 001 866 / 1 682 666 309 | 19
1 687 001 866 / 1 682 666 309 | 19
7. Enter the sample gas value for CO and confirm
with <Ok>.
? The "Propane" dialog window is opened.
8. Enter the sample gas value for propane and confirm
with <Ok>.
? Zero position compensation is implemented.
Set stable flow
The message
Re-adjustment
1. Connect the test hose to the calibrating gas inlet and
set a flow rate of ≥ 1 l/min.