NeuLog NUL-205 Manual page 5

Oxygen logger sensor

Advertisement

Add 1 mL of the "DO Filling Solution" directly into the tip-end which was just
3.
unscrewed.
4.
Carefully reattach the tip-end of the probe by screwing it back into the probe.
5.
Do not replace the blue rubber cap until after testing.
Storage:
Short-term (less than 24 hours):
For short term storage, fill a beaker with enough distilled water to cover the membrane
cap of the oxygen probe. With the membrane cap still attached to the probe, place its
tip side down in the distilled water. This easy method is ideal for short term storage,
between trials for up to 24 hours.
Long-term (more than 24 hours):
1. Unscrew the membrane cap from the oxygen probe.
2.
Discard the DO filling solution from the membrane cap.
3.
Rinse the cap with distilled water and gently shake to air dry.
4.
Rinse and dry the exposed metal electrode with distilled water; it can be dried
with lab wipes.
5.
For storage purposes: Loosely reattach the membrane cap to the oxygen probe.
Maintenance and storage:
Never submerge the Neulog sensor's plastic body in any liquid.
Do not allow liquid into the Neulog plastic body.
After using the probe, wipe off all excess material, liquid or residue from the
oxygen probe body. Do not leave the membrane cap with filling solution for more
than two days.
Store in a box at room temperature out of direct sunlight.
Do not put the plastic case or cord of the oxygen probe in an acid or base.
Store the oxygen sensor's probe upright with the blue rubber tip pointing
downward.
NEULOG OXYGEN LOGGER SENSOR GUIDE
Do not use the sensor in acid or alkaline solutions with a molarity greater than
1.0.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship. The
warranty is for a period of 3 years from the date of purchase and does not cover
damage of the product caused by improper use, abuse, or incorrect storage. Sensors
with a shelf life such as ion selective probes have a warranty of 1 year. Should you
need to act upon the warranty, please contact your distributor. Your sensor will be
repaired or replaced.
Thank you for using NeuLog!