NeuLog NUL-205 Manual page 3

Oxygen logger sensor

Advertisement

computer using the WiFi-201. To engage operation in this fashion, press 3 times
on the panel key. Repeating this operation will return the unit to WiFi mode.
Operation with NeuLog viewer
Materials needed:
NUL-205 Oxygen Sensor
VIEW-101 Graphic Display Module
BAT-200 Battery Module
Procedure:
1.
Connect the oxygen sensor to the left side of the viewer.
2.
Connect the BAT-200 module to the right side of the viewer.
3.
The VIEW-101 will auto detect the sensor and you will see it appear on the left
side of the screen.
4.
Once the sensor appears, it will be monitoring data in real time.
5.
Remove the blue rubber cap on the end of the oxygen probe. Unscrew the
membrane cap, fill half of it with the filling solution and screw it back.
6.
Prior to initial use, the Neulog oxygen sensor needs to be offset, please review
the "Sensor Offset" section above.
7.
To run an experiment and collect data click the run button (little green person)
Off-line experiments
(Off-line experiments are when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-205 Oxygen Sensor
BAT-200 Battery module
Materials needed to configure your offline experiment:
USB-200 Module, WiFi-201 module or VIEW-101 Module
USB to mini USB cable (included with the USB-200 module)
NEULOG OXYGEN LOGGER SENSOR GUIDE
Procedure:
1.
Connect the oxygen sensor directly to the left side of the charged BAT-200
module
2.
Remove the blue rubber cap on the end of the oxygen probe. Unscrew the
membrane cap, fill half of it with the filling solution and screw it back.
3.
Prior to initial use, the Neulog oxygen sensor needs to be calibrated; please
review the "Sensor Offset" section above.
When ready to collect data, press the "Start/Stop" button on the oxygen sensor;
4.
a red light will turn on for the duration of the experiment.
5.
To change the experiment settings, first connect to your chosen device (PC,
Mac, tablet, smart device) and select the "Module setup" button. (For more
information on how to change experiment settings, view quick start procedure
section for your chosen device in this document.)
After the experiment has concluded (when the "Start/Stop" button is pressed
6.
again or the data collection period ends) connect the sensor to your chosen
device (you can find full instructions in each device's quick start procedure
section).
7.
Open the Neulog software.
Click the "Off-line experiment" button.
8.
Click the "Load data from sensors" button.
9.
10.
Select which experimental data to upload (5 experimental runs can be stored at
one time on the NUL-205 oxygen sensor).
Restoring sensor's factory default settings:
Procedure:
1.
Connect the USB-200 to a PC, Mac, tablet, or smart device.
2.
Connect the oxygen sensor to a USB-200 module (they directly plug together).
3.
Open the NeuLog software.
Click the "Tools" icon.
4.
Click "Restore sensor's factory defaults" icon.
5.
If prompted to clear the graph, click either the "Off-line experiment" or "On-line
6.
experiment" button (whichever you are using).
Click "Clear experiment results".
7.
8.
Resume from step 4.