Content 1. General 4 Scope of manual 4 Intended use 4 General safety instructions 4 Standards and directives 5 Symbols 5 Sterilizer views 2. Description of the sterilizer 6 Control panel 6 Technical specification 6 Packing content 7 General conditions 3.
1 General Scope of Manual This manual contains information concerning the installation, operation and maintenance of the steam sterilizers. To ensure proper performance of the sterilizer, the instructions given in this manual should be thoroughly understood and followed. Keep the manual near the sterilizer in an accessible location for future reference. Intended Use The steam sterilizer described in this manual is intended for the sterilization in all medical, dental, beauty, vet and tattoo fields of the following types of instrument loads: solid, porous,...
Symbols For safe operation, please pay close attention to the alert symbols below which can be found on the sterilizer and throughout this manual. Important information (Caution) Hot surface Ground connection 2 Description of the sterilizer Sterilizer views Date of manufacture Manufacture number Manufacturer Name Manufacturer Address...
Door closed Door blocked Time Current internal temperature Up button Down button Enter button Menu button Technical specifications Model STE-8-D STE-12-D STE-18-D STE-23-D STE-29-D Chamber (mm) Φ170 x 320 φ200 x 360 φ247 x 350 φ247 x 450 φ247 x 625...
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used for the sterilizer is recommended. Never connect the device pin to reductions of any type. Location requirements and dimensions (mm) Model STE-8-D STE-12-D STE-18-D STE-23-D...
4 Setup Connect the power cord to an outlet of the appropriate voltage. Turn on the main power switch on the right side. Open the door to remove all of the inner contents for unpacking. After switching on, the machine turns on the LCD and shows the door position, water level, working program, date, time, etc.
Basic Set From the main menu,, select "Basic Set”. Program Date: 11-07-2019 Basic Set Time: 12:05:35 Report Language: ENG Label Counter: 00000 The “Basic Set” menu permits to set the following options: *Date *Time *Language Select the “Basic Set” from the main menu by pressing button.
Page 10
The “Parameter” menu permits to set the following options: *Holding time *Dry time Select “Parameter” from the menu by pressing button. Select the program by pressing then press button. Select the parameter by pressing button. Adjust it by pressing Press button to save and back.
Expiry date (Optional) To modify the expiration date of the labels, select “Expire date” from the menu by pressing button. Adjust it by pressing Press button to save and back. Parameter Unit Month: Preheat Expiry date Water quality (optional) If your sterilizer is equipped with a water quality sensor and you want to deactivate it, select “Water quality”...
Load Open the door then placed the trays inside the chamber by the tray handle. After the instruments are loaded, you may close the door. Select the program Enter to the main menu by pressing button, Select “Program”. Select the program by pressing then press to confirm program, in the screen will appear the information of...
Note: If the cycle is manually interrupted before it reaches the drying phase, the items inside the sterilizer must be considered not sterile. N20 will appear on the screen. (see Error code description). Caution: Depending on the phase of the cycle, steam and water can escape from the sterilization chamber when you open the door.
Data The internal memory will store the information of the last 9999 cycles. USB Flash memory (Optional) A USB drive can be used as a method of storing a report of the cycle. To do so, insert the USB drive into the slot located on the service door of the sterilizer.
Page 15
Sample of a printer report When reading printed data records, refer to the diagram below: ================================ ======================= Program: WRAPPED Program:Vacuum test Temperature: 134C Tp:1℃ Pressure: 206.0 kPa P1:-75.0kPa Drying Time: 08Min P2:-74.0kPa Holding Time: 4.0Min rate of pressure rise:0.10 -------------------------------- Start Time:08:22 Time Temp.
Print labels Select “Labels” from the main menu and press button to enter in the menu. Select the cycle number by pressing button. Choose the labels quantity by pressing then press button to print. Press button to exit. Program 00012 Basic Set 00011 QT: 01...
Clean door seal and mating surfaces with a damp cloth. Note: Do not use bleaching agents or any abrasive materials or substances in the chamber. Failure to comply may cause damage to the chamber and/or other components. Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the surface. Door adjustment Under normal circumstances, the chamber door does not require adjustments.
7 Troubleshooting Code Description Proposed solution Steam generator temperature sensor Power off & run a new cycle error. Contact your supplier if error persists. Power off & run a new cycle Inner temperature sensor error. Contact your supplier if error persists. Temperature sensor of the chamber Carefully ensure that the chamber wall is heated and wall error.
8 Transportation and storage Switch off the sterilizer before transportation or storage. Pull out the plug. Let the machine cool down. Drain the distilled water tank and the used water tank. Conditions for transport and storage Temperature: -20°C ~ +50°C Relative humidity: ≤...
Appendix 1 Water properties / Characteristics Description Feed water Condensate Evaporate residue ≤ 10mg/ I ≤ 1.0mg/kg Silicium oxide sio2 ≤ 1mg/ I ≤ 1.0mg/kg Iron ≤ 0.2mg/ I ≤ 0.1mg/kg Cadmiun ≤ 0.005mg/ I ≤ 0.05mg/kg Lead ≤ 0.05mg/ I ≤...
Appendix 2 Diagrams of the sterilization programs Programs Type (STE-8-D) 15-20 SOLID Unwrapped solid material 2.00 0.60 30-35 25-40 Unwrapped solid material 2.00 0.60 WRAPPED Single-wrapped solid or 45-50 1.50 0.50 hollow material Unwrapped porous material 0.50 0.15 30-45 Single-wrapped porous 0.35...
Page 22
Programs Type (STE-12-D) 15-20 SOLID Unwrapped solid material 3.00 1.00 30-40 30-45 Unwrapped solid material 3.00 1.00 WRAPPED Single-wrapped solid or 45-50 hollow 2.50 0.80 material Unwrapped porous material 0.75 0.25 30-45 Single-wrapped porous 0.50 0.15 material Dual-wrapped porous material 0.30 0.10 TEXTILE...
Page 23
Programs Type ( STE-18-D) 15-30 SOLID Unwrapped solid material 4.50 1.20 30-40 30-45 Unwrapped solid material 4.50 1.20 WRAPPED Single-wrapped solid or hollow 45-60 3.50 1.10 material Unwrapped porous material 1.00 0.30 35-50 Single-wrapped porous 0.80 0.25 material TEXTILE Dual-wrapped porous material 0.60 0.20 Single-wrapped hollow material...
Page 24
Programs Type (STE-23-D) 25-45 SOLID Unwrapped solid material 5.00 1.50 30-60 35-60 Unwrapped solid material 5.00 1.50 Single-wrapped solid or WRAPPED 35-65 hollow 4.00 1.20 material Unwrapped porous material 1.25 0.40 45-65 Single-wrapped porous 1.10 0.30 material Dual-wrapped porous material 0.75 0.25 TEXTILE...
Page 25
Programs Type (STE-29-D) SOLID 30-45 Unwrapped solid material Unwrapped porous material Single-wrapped porous 45-60 material Dual-wrapped porous material WRAPPED Single-wrapped hollow material 50-65 Dual-wrapped solid and hollow material Unwrapped porous material Single-wrapped porous material Dual-wrapped porous material PRION 55-70 Single-wrapped hollow material Dual-wrapped solid and hollow...