Page 2
The BioniCare Knee System involves the use of a therapeutic electrical current that mimics the signal of a healthy knee to treat osteoarthritis. It takes daily use over several weeks to feel a reduction in pain and an improvement in NOT FEEL functioning.
Page 3
STEP TAKE THESE ITEMS STEP STEP Press battery door INSERT lead wire into up to open and lift off. INSERT the BioniCare device battery into the device. Slide door to close. LEAD WIRE PORT LEAD WIRE 1 Rechargeable 1 BioniCare Device Battery SIDE VIEW THIS SIDE UP...
Page 4
STEP STEP Replace thigh liner by removing the standard Apply thigh electrode to thigh liner. OActive 2 liner from the brace and replacing it with the Peel WHITE backing off thigh electrode and adhere electrode to the center of the thigh liner with white one included in the BioniCare kit.
Page 5
STEP Apply knee electrode. Peel clear plastic backing from knee electrode. With knee at 45°, adhere electrode to knee by first placing center section over kneecap with wire extending down as shown. Center the round area in the middle of the electrode over the kneecap.
Page 6
STEP STEP Press middle button to turn on device. Answer the following questions. Press the unlock icon on screen to unlock. Your device will periodically ask 4 questions to chart your progress. Press the middle button to answer and the side ...
Page 7
BIONICARE KNEE WRAP SYSTEM QUICK START GUIDE STEP TAKE THESE ITEMS STEP STEP Press battery door INSERT lead wire into up to open and lift off. INSERT the BioniCare device battery into the device. Slide door to close. LEAD WIRE PORT LEAD WIRE 1 Rechargeable 1 BioniCare Device...
Page 8
STEP TAKE THESE ITEMS STEP Lay Wrap on table with 1 KNEE and 1 THIGH electrode. 2 Thigh Electrodes 1 BioniCare Wrap 2 Knee Electrodes STEP STEP Peel CLEAR backing off each electrode. Peel WHITE backing off knee electrode and apply within outlines on BioniCare Wrap.
Page 9
STEP STEP Apply BioniCare Wrap to knee. Wrap top and bottom straps around leg. WIRE KNEE THIGH SIDE VIEW KNEE CALF WIRE Line up circle of knee electrode over your kneecap and thigh Wires should extend downward electrode over your thigh with electrodes sticking to skin. and upward from the BioniCare Wrap.
Page 10
STEP STEP Answer the following questions. Press play button to start treatment. Your device will periodically ask 4 questions to chart your progress. Press the middle button to answer and the side buttons to scroll through answer options. STEP Increase past 2.0 volts then continue to Recommended treatment time is a minimum of 8 hours a increase until you feel tingling.
Never use bleach, fabric softener, or harsh detergents. • If you experience symptoms of skin irritation while using the BioniCare system, discontinue use immediately and contact VQ OrthoCare’s Patient Care FOR QUESTIONS OR TROUBLESHOOTING PLEASE CALL Department at 800.452.7993, or consult your BioniCare provider or healthcare 800.452.7993 TO SPEAK TO A BIONICARE REPRESENTATIVE...
Page 12
State in which • Do not use the BioniCare Knee System for any electrode place- • Patients with demand style cardiac pacemakers should he/she practices to use or order the use of the device.