3nh HG Series Operation Manual

3nh HG Series Operation Manual

Gloss meter
Hide thumbs Also See for HG Series:

Advertisement

Quick Links

GLOSS METER
OPERATION MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for 3nh HG Series

  • Page 1 GLOSS METER OPERATION MANUAL...
  • Page 2: Table Of Contents

    Content Introduction Cautions 1 External Structure Description 2 Operating Instruction 2.1 Turning On/Off 2.2 Calibration 2.3 Measurement 2.4 Data Save 2.5 Connecting to PC 2.6 Print 3 Function Description of QC Software 4 Routine Maintenance 5 Technical Parameters...
  • Page 3: Introduction

    Safety Instructions The instrument is a safe device. Before operating, please read the safety instructions and strictly comply with the following terms to avoid unexpected damages. We shall not be liable for any losses by incorrect operation. Battery The device configures a built-in battery. Please use the original one.
  • Page 4 c. Aesthetic design perfectly combined with ergonomics structure. d. Three measurement angles (20°, 60°, 85°),can measure simultaneously (except the single-angle gloss meter). e. QC software with powerful extending function (except certain models). f. Flexible choice of auto-calibrating when power on, easy to operate. g.
  • Page 5 1. External Constructions Figure 1 External Constructions Button Function Instruction LCD Screen: Display measuring data and instrument operation navigation. Switch/Measure button: Long-press the button 3 seconds to turn on or turn off the gloss meter. Short press the button is to measure. Indicator Light: When turning on, it will show green light.
  • Page 6: Operating Instruction

    Figure 2 Separation Method 2. Operating Instruction 2.1 Turn on/off Long-press the switch button 3 seconds to turn on the instrument. The LCD screen will display boot logo. After a few seconds, it will enter into the measurement interface automatically, as shown in figure 3. Long-press the switch button 3 seconds again, it will be turn off.
  • Page 7: Measurement

    power-off is manually, it will calibrate automatically when turn on next time. If the last power-off is automatic after 5 minutes backlight off, then it will not calibrate automatically when turn on next time. When operating environment change (such as the rapid change of temperature, altitude and humidity), it must do calibration.
  • Page 8: Data Save

    Figure 4 Multi-angle Measurement “T005” at the upper left means the record number of the last measurement. “16:12” and “2015.10.23” are time and date. “T001-T005” is the record number of five measurements. (Some models only display three records.) “T102316” is a default name of measuring record. It is making up of “T”+”month”+”day”, “T”...
  • Page 9: Routine Maintenance

    a) Check Status (Instrument basic information, such as model, serial number) b) Calibrate c) Change Calibration Value (It’s better to handle by manufacturer of qualified metrological institutes.) d) Data Management (Check record, delete record, export record, print report) e) Set Time and Data f) Set Language g) Select Angles (only for tri-angle gloss meter) h) Set Auto Save or Not-save Function.
  • Page 10 5. Technical Specifications Measurement Angle 20°/60°/85°/20°60°85° Standard ISO 2813, GB/T 9754, ASTM D 523, ASTM D 2457 Measuring Area(mm) 20°: 10x10, 60°: 9x15, 85°: 5x36 Measuring Range Multi-angle Models Single Models Some Single Models 20°: 0-2000GU 60°: 0~300GU 60°: 0~200GU 60°: 0~1000GU 85°: 0~160GU Division Value...