Page 3
Terms Used in this Manual This guide is designed to give key concepts on safety precautions. WARNING A WARNING label advises against certain actions or situations that could result in personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure.
Chapter 2 Introduction........................6 2.1 ECG Sampling Box Appearance .................... 7 2.2 Features ..........................10 Chapter 3 Assembling VE-1010 VET PC ECG System ............11 Chapter 4 Installing VE-1010 VET PC ECG Software ............12 4.1 System Running Environment ..................... 12 4.1.1 Requirements on the Hardware of the PC ..............
Page 5
6.5.2.2 Editing the HRV Waveform on the Waveform Interface ........34 6.5.2.3 Previewing HRV..................... 34 6.5.3 Printing ECG Reports....................35 6.5.4 Saving ECG Reports ..................... 36 6.6 Sampling STAT ECG ......................37 Chapter 7 Processing Pet Records ....................38 7.1 Searching Pet Records......................38 7.2 Modifying Pet Records......................
Page 6
Chapter 9 Hint Information......................58 Chapter 10 Accessories ........................ 59 Chapter 11 Cleaning and Maintenance..................60 11.1 Cleaning and Maintaining the Patient Cable and Reusable Electrodes......60 11.2 Disinfection ........................61 Chapter 12 Warranty and Service....................62 12.1 Warranty ..........................62 12.2 Contact information......................
VE-1010 VET PC ECG is a PC-based diagnostic tool intended to acquire, process and store ECG signals from pets undergoing resting test. VE-1010 VET PC ECG is intended to be used only in pet hospitals by trained veterinary surgeons or technicians or drug researchers. The cardiogram recorded by VE-1010 VET PC ECG can help users to analyze and diagnose heart disease.
Page 8
VE-1010 Veterinary PC ECG User Manual Safety Guidance WARNING Connecting to other devices may decrease the antistatic gradation of this device during operation. EXPLOSION HAZARD - Do not use the system in the presence of flammable anesthetic mixtures with oxygen or other flammable agents.
Page 9
VE-1010 Veterinary PC ECG User Manual Safety Guidance WARNING requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our technical service department or your local distributor. 23. The summation of leakage current should never exceed leakage current limits while several other units are used at the same time.
Page 10
VE-1010 Veterinary PC ECG User Manual Safety Guidance CAUTION e) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohm. f) Test the enclosure leakage current according to IEC/EN 60601-1: Limit: NC 100 μA, SFC 500 μA. g) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10 μA, d.c.
VE-1010 Veterinary PC ECG User Manual Safety Guidance 1.3 List of Symbols Equipment or part of CF type with defibrillator proof Caution Consult Instructions for Use Recycle Part Number Serial Number Date of Manufacture Manufacturer Authorized Representative in the European Community...
With high performance, reliability and easy operation which are suitable for ECG diagnosis of medical institutes of all kinds and useful for the heart disease analysis, the VE-1010 VET PC ECG can help to reduce doctors’ workload greatly.
VE-1010 Veterinary PC ECG User Manual Introduction WARNING The system should be installed by a qualified service engineer. Do not power on the system until all cables are properly connected and verified. 2.1 ECG Sampling Box Appearance ECG Sampling Box Appearance...
Page 14
VE-1010 Veterinary PC ECG User Manual Introduction WARNING 1. When the computer connected to the USB cable is powered on, do not connect or disconnect the USB cable to the ECG sampling box. 2. It is not necessary or recommended to regularly disconnect the USB cable from the ECG sampling box.
Page 15
VE-1010 Veterinary PC ECG User Manual Introduction Patient Cable Socket : Applied part of type CF with defibrillator proof : Caution Definitions of corresponding pins: Signal Signal Signal C2 / V2 F / LL C1 / V1 C3 / V3...
VE-1010 Veterinary PC ECG User Manual Introduction WARNING 1. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
VE-1010 Veterinary PC ECG User Manual Assembling VE-1010 VET PC ECG System Chapter 3 Assembling VE-1010 VET PC ECG System WARNING 1. Use a special grounded socket to get accurate voltage and current. 2. When using a notebook computer with a two-prong plug, please connect a grounded printer to avoid power interference.
VE-1010 Veterinary PC ECG User Manual Installing VE-1010 VET PC ECG Software Chapter 4 Installing VE-1010 VET PC ECG Software 4.1 System Running Environment 4.1.1 Requirements on the Hardware of the PC Pentium P4, Celeron D 310 or above CPU:...
Environment Detection interface shows that a certain component needs to be installed, please install it manually. Click on the Help button to see the installation guide. For details on installing VE-1010 VET PC ECG software, please refer to VE-1010 Veterinary PC ECG Installation Guide. - 13 -...
VE-1010 Veterinary PC ECG User Manual Preparations Before Operation Chapter 5 Preparations before Operation WARNING The performance and electric shock protection can be guaranteed only if original EDAN patient cable and electrodes are used. 5.1 Connecting the Patient Cable to Electrodes...
Page 21
VE-1010 Veterinary PC ECG User Manual Preparations Before Operation The identifier and color code of electrode connectors used complies with IEC requirements. In order to avoid incorrect connections, the electrode identifier and color code is specified in Table 5-1. Moreover the equivalent code according to American requirements is given too.
Double-click on the shortcut icon on the desktop to display the Initial interface. Note: Do not use other software when using VE-1010 VET PC ECG software. Figure 6-1 Initial Interface The toolbar contains five buttons. From left to right, they are New Pet, STAT ECG, Data Manager, System Setting and Exit.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 6.1 Selecting a Pet Record to Start a New Test Click on the Data Manager button on the Initial Interface (Figure 6-1) to open the Data Manager interface (Figure 6-2).
Page 24
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 3. Or click on Advanced Search to display the Search Condition window. Enter the search conditions, and click on the Search button, and all the pet records which meet the conditions will be displayed in the pet information list.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Figure 6-3 Pet Information Interface Note: Click on an option in the pet information list, such as ID, name, etc, and then all the pet records will be arranged in sequence.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Then you need to input the pet’s related information. 1. Input basic information, such as pet ID, pet name, sex, age, etc. You can enter the physician name or the technician name directly in the Physician or Technician text box, and click on the OK button to save it on the Pet Information interface.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Figure 6-4 Pre-Sampling Interface Before sampling, if you do not connect the PC to the ECG sampling box, the following hint will pop up. 6.4.1 Specifying Display Mode 6-lead display mode There are two display modes including 6*1 and 3*2 in the 6 Leads mode.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG When the display mode is set to 6*1, 6-channel ECG waves are displayed on one screen. When the display mode is set to 3*2, 6-channel ECG waves are displayed in 2 groups of 3 on one screen.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 6.4.4 Specifying Lowpass Filter Lowpass Filter restricts the bandwidth of input signals. The cutoff frequency can be set to 25Hz, 35Hz, 45Hz, 75Hz, 100Hz, or 150Hz. The input signals whose frequency is higher than the set cutoff frequency will be attenuated.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 6.4.6 Stopping Sampling Data After clicking on the Start button, there are two ways to stop sampling data. 1. The system will stop sampling ECG data and display the ECG analysis interface automatically after the ECG sampling time is over.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Click on Exit to return to the ECG sampling interface. Note: Only 6*1 and 7*1 display modes can be displayed on the Wave Review interface. 6.5 Analyzing ECG Data You can open the ECG analysis interface in one of the following three ways: 1.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 6.5.1.1 Viewing the Waveform Click on the Waveform tab on the Normal ECG Analysis Interface to open the Waveform interface (Figure 6-6). Figure 6-6 Normal ECG - Waveform Interface You can choose a speed, a gain and a display mode for the displayed waves from the pull-down lists.
Page 33
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Click on the Re-Sample button and then the system can re-sample ECG data. Click on the Re-Diagnosis button and then the system can re-diagnose the 10s ECG data on the screen automatically.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG To Edit Diagnosis Result on the Waveform Interface To Edit the Diagnosis Result, 1. Enter your own opinions in the diagnosis textbox, and then click on the Save button.
Page 35
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Figure 6-7 Normal ECG - Average Template Interface You can press a lead button in to display magnified average templates of this lead. When you press ALL, magnified average templates of all leads will be overlapped with the same central axis.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 6.5.1.3 About the Detail Information Interface Click on the Detail information tab on the normal ECG analysis interface to open the detail information interface. This interface displays lead parameter values as Figure 6-8 shows.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Figure 6-9 Normal ECG - Rhythm Wave Interface You can click on Previous Page or Next Page to display the waves of the previous or next page. Click on one point on the wave, and then drag the mouse to another point. The section of wave you select will be printed.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 4. Click on the Zoom Out button on the toolbar to minify the preview page. 5. Click on the Close button to close the normal ECG preview interface and return to the previous interface.
Page 39
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Figure 6-11 Analysis Interface of HRV As follows: Designation Definition Sampling time Set sampling time Total beat number during the measuring course Total Beat Heart Rate Average RR interval...
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG User can edit diagnosis result on the Waveform interface. For details, please refer to Section 6.5.1.1. 6.5.2.2 Editing the HRV Waveform on the Waveform Interface Click on the Auto Diagnosis Result tab to enter Auto diagnosis result interface.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG Figure 6-12 HRV Preview Interface 6.5.3 Printing ECG Reports 1. Choose Start > Printers and Faxes, and then right-click on the icon of the printer used, and select Set as Default Printer. Then close the Printers and Faxes interface.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG 2. Click on the Print button on the analysis interface to print an ECG report. 3. Or click on the Print button on the preview interface to print an ECG report.
VE-1010 Veterinary PC ECG User Manual Operation Instructions for Resting ECG If you select Send, the sampled data will be sent by email (outlook express) when it is saved to the designated directory. During the saving and sending course, the system will give the hint information.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records Chapter 7 Processing Pet Records Click on the Data Manager button on the Initial Interface (Figure 6-1) to open the Data Manager interface (Figure 7-1). Figure 7-1 Data Manager Interface Click on a pet record in the pet information list, and then all the examination records of the pet will be displayed in the examination record list.
Page 45
VE-1010 Veterinary PC ECG User Manual Processing Pet Records Note: Click on an item in the pet information list, such as ID, name, etc, and then all the pet records will be arranged in sequence. 2. Or select a search item in the pull-down list, enter the corresponding information in the right textbox, and then click on the Search button.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records 4. Or you can click on Physician or Examination Record, and choose the physician name or examination types, all the pet records which meet the conditions will be displayed in the pet information list.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records 7.3 Deleting Records Note: The deletion of records is permanent, and you can’t restore the records deleted. Please use this operation cautiously. 7.3.1 Deleting Pet Records Click on a pet record in the pet information list on the Data Manager interface, and then click on the Delete button to delete the pet record from the pet information list.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records 7.5 Merging Examination Records Select one or multiple pet records in the pet information list on the Data Manager interface, and click on the Merge/Assign button to assign the selected records. Input the pet ID and click on the OK button to designate the pet information of the selected records to the .pet.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records 7.6 Comparing Two Examination Records Press the Ctrl button on the keyboard and select two examination records, and then click on the Compare button to display the Compare interface. Note: Please select two records to compare only in Resting ECG.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records 7.7 Importing ECG Data into the Data Manager Interface Click on the Import button on the Data Manager interface (Figure 7-1) to open the following window. Select the record or the folder to be imported and click on the Select button to import the record or the folder into the Data Manager interface.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records When the import is successful, the following hint will pop up. If the imported folder includes no valid data, the following hint will pop up. Note: Only ECG data in DAT format created by VET PC ECG software can be imported.
VE-1010 Veterinary PC ECG User Manual Processing Pet Records When the export is successful, the following hint will pop up. 7.9 Viewing an Examination Record Click on a pet record in the pet information list, and then all the records of the pet will be displayed in the examination record list.
VE-1010 Veterinary PC ECG User Manual Configuring the System Chapter 8 Configuring the System Click on the System Setting button on the Initial Interface (Figure 6-1) to open the System Setting interface. There are four tabs on the System Setting interface: Basic Information, Sample Setting, Device, Print Setting and Others.
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.1.1 Setting Basic Information Enter information in the Hospital Name, Physician Name, Technician Name, or User-Defined 1/2/3/4 textbox on the Basic Information setup interface (Figure 8-1). When you fill in the User-Defined 1/2/3/4 textbox, the corresponding items on the Pet information interface will change into what is filled.
VE-1010 Veterinary PC ECG User Manual Configuring the System Note: Click on the New Pet button on the Initial Interface to open the Pet information interface as the figure (a) shows. Click on the Data Manager button on the Initial...
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.1.3 Setting Language You can set the language to Chinese or English. Note: To validate the language setup, after setting, you should exit the system and open it again. 8.1.4 Specifying the Storage Path of the ECG Data Click on the Browse button on the basic information setup interface (Figure 8-1) to assign the storage path.
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.2.1 Setting Sample Select a sampling device from the Sampling Device pull-down list on the Sample Setting interface (Figure 8-2). Sampling Port can be set to a value from COM0 to COM9.
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.2.3 Setting Sampling Time You can enter the normal ECG sampling time manually. The range is 10~600s. You can enter the HRV sampling time manually. The range is 1~15min. You can set the HRV analysis lead to one of the 6 leads (І, П, Ш, aVR, aVL, aVF) or 7 leads (І, П, Ш, aVR, aVL, aVF, V).
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.3 Setting Printer Figure 8-3 Printer Setting Interface 8.3.1 Choosing Pet Information to be Printed The default item of the pet information is Owner. You can also select the additional information, such as weight, BP and species.
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.3.3 Setting Rhythm Lead The rhythm lead can be one of 6 standard leads or 7 standard leads. When the printing mode is set to 3×2+1 or 3×3+1, the rhythm lead selected in the Rhythm list box will be printed out.
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.4 Setting Others Figure 8-4 Other Setup Interface 8.4.1 Setting Unit Set the weight unit to Kg or Pound. Set the BP unit to kPa or mmHg. Set the time mode to 24Hour or 12Hour.
VE-1010 Veterinary PC ECG User Manual Configuring the System 8.5 Modifying the Glossary Click on Edit Diagnosis Template in the Tool (F) pull-down list on the Initial Interface (Figure 6-1), and then the Edit Diagnosis Template window appears. 1. Adding an item Enter a diagnosis item in the textbox, such as aa, and then click on the Add button.
Page 63
VE-1010 Veterinary PC ECG User Manual Configuring the System 2. Adding a sub-item Click on the item you wanted to add a sub-item, enter a diagnosis sub-item, such as bb in the textbox, and then click on the Add button. The added sub-item will be displayed under aa.
VE-1010 Veterinary PC ECG User Manual Hint Information Chapter 9 Hint Information Hint information and the corresponding causes provided by the system are listed as follows. Table 9-1 Hint Information and Causes Hint Information Causes Electrodes fall off the pet or the patient cable Lead off: X falls off the ECG sampling box.
VE-1010 Veterinary PC ECG User Manual Accessories Chapter 10 Accessories WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. Table 10-1 Standard Accessory List...
VE-1010 Veterinary PC ECG User Manual Cleaning and Maintenance Chapter 11 Cleaning and Maintenance CAUTION Turn off the system power and drag the power cable out from the socket before cleaning or disinfection. 11.1 Cleaning and Maintaining the Patient Cable and...
VE-1010 Veterinary PC ECG User Manual Cleaning and Maintenance 11.2 Disinfection To avoid permanent damage to the equipment, disinfection can be performed only when it is considered as necessary according to your hospital’s regulations. Before disinfection, clean the equipment first. Then wipe the surfaces of the unit and the patient cable with hospital standard disinfectant.
VE-1010 Veterinary PC ECG User Manual Warranty and Service Chapter 12 Warranty and Service 12.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period.
VE-1010 Veterinary PC ECG User Manual Recommended Optional Accessories Chapter 13 Recommended Optional Accessories Electrical Outlet: Power Consumption: no less than 4500VA Special use for medical equipment Printer: Model: HP2035, HP5568 Manufacturer: HP Company, USA Model: CANON3500, CANON1800 Manufacturer: CANON Company, Japan WARNING 1.
Guidance and manufacture’s declaration - electromagnetic emission VE-1010 VET PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of VE-1010 VET PC ECG should assure that it is used in such an environment.
Page 74
Guidance and manufacture’s declaration - electromagnetic immunity VE-1010 VET PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of VE-1010 VET PC ECG should assure that it is used in such an environment.
Page 75
Guidance and manufacture’s declaration - electromagnetic immunity VE-1010 VET PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of VE-1010 VET PC ECG should assure that it is used in such an environment.
Page 76
VE-1010 VET PC ECG is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of VE-1010 VET PC ECG can help prevent electromagnetic interference by maintaining a minimum distance between...
VE-1010 Veterinary PC ECG User Manual Abbreviation Appendix 3 Abbreviation Abbr English Electrocardiograph/Electrocardiogram Heart Rate Variability Heart Rate P-wave Duration P-R Interval QRS Complexes Duration QT/QTc Q-T Interval of the Current Lead / Normalized QT Interval Dominant Direction of the Average Integrated ECG...
Page 78
VE-1010 Veterinary PC ECG User Manual Abbreviation Right Right Arm Right Leg Identification Alternating Current Universal Serial Bus - 72 -...