NI PXIe-4136 Calibration Manual

NI PXIe-4136 Calibration Manual

Calibration procedure, single-channel system source measure unit (smu)
Hide thumbs Also See for PXIe-4136:

Advertisement

Quick Links

CALIBRATION PROCEDURE
NI PXIe-4136
Single-Channel System Source Measure Unit (SMU)
This document contains the verification and adjustment procedures for the NI PXIe-4136
(PXIe-4136). Refer to
Contents
Required Software.....................................................................................................................2
Related Documentation.............................................................................................................2
Password................................................................................................................................... 2
Calibration Interval................................................................................................................... 2
Test Equipment..........................................................................................................................3
Test Conditions..........................................................................................................................3
Safety Guidelines for System Operation...................................................................................4
As-Found and As-Left Limits................................................................................................... 5
Calibration Overview................................................................................................................ 5
Verification................................................................................................................................6
Self-Calibrating the PXIe-4136........................................................................................ 6
Testing the Safety Interlock.............................................................................................. 7
Verifying Voltage Measurement and Output.....................................................................8
Verifying Remote Sense Voltage Offset............................................................................9
Verifying Voltage Remote Sense.......................................................................................9
Verifying Current Offset..................................................................................................11
Verifying Load Regulation..............................................................................................12
Verifying 1 μA and 10 μA Current Measurement and Output........................................13
Verifying 1 A Current Measurement and Output............................................................17
Adjustment.............................................................................................................................. 18
Adjusted Specifications...................................................................................................19
Initiating the Adjustment Session................................................................................... 19
Voltage and Current Output............................................................................................ 19
Residual Offset Voltage...................................................................................................23
Closing the Adjustment Session..................................................................................... 23
Alternative to Performing Adjustment Procedures......................................................... 24
Reverification..........................................................................................................................24
Worldwide Support and Services............................................................................................ 24
ni.com/calibration
for more information about calibration solutions.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PXIe-4136 and is the answer not in the manual?

Questions and answers

Summary of Contents for NI PXIe-4136

  • Page 1: Table Of Contents

    CALIBRATION PROCEDURE NI PXIe-4136 Single-Channel System Source Measure Unit (SMU) This document contains the verification and adjustment procedures for the NI PXIe-4136 (PXIe-4136). Refer to ni.com/calibration for more information about calibration solutions. Contents Required Software........................2 Related Documentation......................2 Password........................... 2 Calibration Interval........................2 Test Equipment..........................3...
  • Page 2: Required Software

    Required Software Calibrating the PXIe-4136 requires you to install the following software on the calibration system: • NI-DCPower. The PXIe-4136 was first supported in NI-DCPower 15.1. • Supported application development environment (ADE)—LabVIEW or ™ ™ LabWindows /CVI • Supported operating system—Windows.
  • Page 3: Test Equipment

    Vishay PTF563K0000BYEB Remote sense 0.1% 250 mW accuracy Test Conditions Follow the setup and environmental information below to ensure the PXIe-4136 meets the published specifications. Test limits in this document are based on the June 2015 edition of the NI PXIe-4136 Specifications. •...
  • Page 4: Safety Guidelines For System Operation

    ±40 V DC, and is the internal device temperature recorded by the PXIe-4136 at the completion of the last self- calibration. Call the niDCPower Get Self Cal Last Temp VI to query T from the PXIe-4136.
  • Page 5: As-Found And As-Left Limits

    The as-left limits are equal to the published NI specifications for the device, less guard bands for measurement uncertainty, temperature drift, and drift over time. NI uses these limits to determine whether the device will meet the device specifications over its calibration interval.
  • Page 6: Verification

    You must complete all verification procedures in the specified order. You do not need to separately verify both measurement and output. The architecture of the PXIe-4136 ensures that if measurement is accurate, then output is as well, and vice versa. Related Information...
  • Page 7: Testing The Safety Interlock

    Disconnect or disable all connections to the PXIe-4136. Allow the PXIe-4136 30 minutes to warm up with the PXI chassis fans set to HIGH. Initialize an NI-DCPower session. Call the self-calibration function. Close the NI-DCPower session. Testing the Safety Interlock In order to ensure safe operation of the PXIe-4136, test the safety interlock for proper functionality before completing any verification procedures.
  • Page 8: Verifying Voltage Measurement And Output

    0.025% + 20 mV 0.0081% + 10 mV 200 V Set the first specified level range, limit range, and limit on the PXIe-4136. Set the niDCPower Sense property or NIDCPOWER_ATTR_SENSE attribute to Local. Measure the internal device temperature and perform self-calibration if necessary.
  • Page 9: Verifying Remote Sense Voltage Offset

    If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. If more than one level range is specified, repeat the previous steps using the values specified in each level range.
  • Page 10 Sense HI Load1 Sense LO Set the first specified level range, limit range, and limit on the PXIe-4136. Measure the internal device temperature and perform self-calibration if necessary. If the internal device temperature exceeds T ±1 °C, wait up to five minutes for the temperature to stabilize to within T ±1 °C.
  • Page 11: Verifying Current Offset

    Load2 Sense LO Verifying Current Offset Remove all connections from the PXIe-4136 and confirm the current measured by the PXIe-4136 at 0 V falls within the test limits. Refer to the following table as you complete the following steps. Complete this procedure only after successfully completing all previous verification procedures.
  • Page 12: Verifying Load Regulation

    Note Although load regulation is listed as a typical specification for the PXIe-4136, verification is required. If the PXIe-4136 fails the load regulation verification procedure, discontinue use of the device and contact an authorized NI service representative to request a Return Material Authorization (RMA).
  • Page 13: Verifying 1 Μa And 10 Μa Current Measurement And Output

    Take a voltage measurement using the PXIe-4136. Verifying 1 μA and 10 μA Current Measurement and Output Compare a set of measured currents reported by the PXIe-4136 to the currents measured by a DMM. Refer to the following table as you complete the following steps.
  • Page 14 Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the PXIe-4136. Set the first specified level range, limit range, and limit on the PXIe-4136. Measure the internal device temperature and perform self-calibration if necessary. If the internal device temperature exceeds T ±1 °C, wait up to five minutes for the...
  • Page 15: Verifying 100 Μa To 100 Ma Current Measurement And Output

    13. If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. 14. If more than one level range is specified, repeat the previous steps using the values specified in each level range.
  • Page 16 Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Current for the PXIe-4136. Set the first specified level range, limit range, and limit on the PXIe-4136. Measure the internal device temperature and perform self-calibration if necessary. If the internal device temperature exceeds T ±1 °C, wait up to five minutes for the...
  • Page 17: Verifying 1 A Current Measurement And Output

    If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. If more than one level range is specified, repeat the previous steps using the values specified in each level range.
  • Page 18: Adjustment

    Verify that the calculated DMM Measured Current value falls within the test limits. If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. Adjustment This section describes the steps needed to adjust the PXIe-4136 to meet published specifications.
  • Page 19: Adjusted Specifications

    Note You do not need to separately adjust both measurement and output. The architecture of the PXIe-4136 ensures that if measurement is accurate, then output is as well, and vice versa. Initiating the Adjustment Session After completing verification, wait a minimum of five minutes for the internal device temperature to stabilize.
  • Page 20 If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. Update the output calibration constants by configuring and calling the niDCPower Cal Adjust Voltage Level VI or niDCPower_CalAdjustVoltageLevel function.
  • Page 21 If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. Adjusting the 100 µA and 1 mA level ranges automatically adjusts the following ranges: 1 μA, 10 µA, 10 mA, and 100 mA.
  • Page 22 Adjusting 1 A Current Output and Measurement Compare a set of measured currents reported by the PXIe-4136 to the currents measured by an external DMM. Refer to the following table as you complete the following steps.
  • Page 23: Residual Offset Voltage

    If more than one test point per level range is specified, repeat the previous steps for each test point, from setting the level to the test point on the PXIe-4136 up to this step. Update the output calibration constants by configuring and calling the niDCPower Cal Adjust Current Limit VI or niDCPower_CalAdjustCurrentLimit function.
  • Page 24: Alternative To Performing Adjustment Procedures

    NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For telephone support in the United 24 | ni.com | NI PXIe-4136 Calibration Procedure...
  • Page 25 United States, visit the Worldwide Offices section of ni.com/ niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. NI PXIe-4136 Calibration Procedure | © National Instruments | 25...
  • Page 26 . You can find information about end-user license agreements (EULAs) ni.com/patents and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/ for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other legal/export-compliance import/export data.