Chematest 35
Menu Explanations
Standard
Electrode
Temp.
Compensation
Filter time
constant
Factor TDS
Reset cell
constant
4.2 Identification
Attribute
ID List
List of Users
A-96.730.121 / 160920
Enter the mV value of the redox/ORP standard.
Range: 200–900 mV
Conductivity
Choose the compensation model which fits best to your application.
Temp. Compensation
None
Coefficient
Non-linear DIN
 None: No temperature compensation.
 Coefficient: Allows the entry of a coefficient for linear
temperature compensation. Range: 0.00%–3.00%.
 Non-linear DIN: the non-linear temperature compensation
should be set for the conductivity measurement of natural waters
(EN 27888, ISO 7888).
Used to damp noisy signals. The higher the filter time constant, the
slower the system reacts to changes of the measured value.
Range: 5–300 s
Factor for the calculation of TDS.
Range: 0.000– 10.000
Allows to display the cell constant determined during factory cali-
bration and to reset the cell constant to this value.
Optionally, each measurement can be assigned an ID (e.g. for
sampling points) and/or a user name. If activated, the ID and/or
user is queried at the end of each measurement.
Select which attributes are to be queried:
Attribute
None
ID
User
ID&User
Enter up to 10 IDs.
Enter up to 10 user names.
43