Chematest 35
Troubleshooting
Measurement
after adding
reagents fails
36
Error message "Operation not possible! Signal out of range.
Consult the manual." is displayed.
Possible cause
Extraneous light
Cuvette dirty
Concentration or pH value
outside the measuring
range of the instrument
Reagents expired
Measured value below
zero value
Corrective action
Make sure that the cuvette cap is
properly closed.
Clean and rinse the cuvette.
Consult
Instrument
Specifications, p.
Repeat measurement.
Use new reagents.
Avoid formation of bubbles when
filling the sample into the cuvette.
Repeat measurement.
A-96.730.121 / 160920
50.