Testing the alarms

Important!
bump or span calibration test. If any alarm does not
respond, check the Alarm Settings in Config Mode to
make sure all alarms are enabled (Config Mode › Alm
Setting › Alarm Settings › Both Enabled). If any alarms
are enabled but are not functional, do not use the
instrument. Under normal operation mode and non-
alarm conditions, the buzzer, LED and backlight can
be tested anytime by pressing the Right Key once.

Normal Mode

After warm-up is complete, the unit enters normal
reading mode and displays the instantaneous gas
concentration continuously.

Configuration Mode

In Config Mode, the user can perform a sensor
calibration, change the alarm limits and set up other
parameters.
Test the alarms prior to performing a
Other values like STEL, TWA,
and temperature, etc. can be
checked by pressing the Up/
Down Key to cycle through
these screens one by one.