Advertisement

Quick Links

Kinetic Spectra
Instruction & Safety Manual
Please read this manual before operation, for
maintenance instructions and safe usage
The enabling power
of applied knowledge

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Kinetic Spectra and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Unicare Health Kinetic Spectra

  • Page 1 Kinetic Spectra Instruction & Safety Manual Please read this manual before operation, for maintenance instructions and safe usage The enabling power of applied knowledge...
  • Page 2: Table Of Contents

    Kinetic Spectra, Instruction & Safety Manual Index Defi nition ..........................2 Description ..........................3 Safety ..4 Electrical connection ......................5 Use of the hand control Select your language ....................6 Locking the hand control setting ................7 START/STOP/REVERSE function ................7 Quick Start .......................8 Possible values for each parameter .................8...
  • Page 3: Defi Nition

    Defi nition The KINETEC Spectra is a Knee PASSIVE mobilization device enabling the extension and fl exion movement from -10° to 120°. Indications • Knee replacement surgery. • Fractures (patellar, tibia plateau, femoral,...). • Arthrolysis • Hip surgery, including hip replacement, hip pinning, osteotomy,...). •...
  • Page 4: Description

    Kinetic Spectra, Instruction & Safety Manual Description KINETEC Spectra consists of the following components: Lower limb support. high support. Foot support and hand control location for transport. Hand control. Thigh support setting lock. Lower limb support setting lock . Foot support positioning setting lock.
  • Page 5: Safety

    Safety The physician defi nes the protocol and ensures that it is correctly implemented (adjustments, session duration and frequency of use). The patient must know the start/stop/reverse function on the control handle. Hand control must be accessible to patient at all times.
  • Page 6: Electrical Connection

    Kinetic Spectra, Instruction & Safety Manual Electrical Connection KINETEC Spectra is a type B class I device. Before connecting the device to the power supply, check that the mains voltage matches that shown on the identifi cation plate (100-240 V~ 50-60Hz).
  • Page 7: Use Of The Hand Control

    Use of the Hand Control Select your language Begining Keys to press Display Remarks Switch the Unit ON KINETIC Spectra V01.0 Check if the hand Please wait control is not blocked Kinetic (See page 7) STOP Press the 2 keys in Language: The display indicates the same time...
  • Page 8: Locking The Hand Control Setting

    Kinetic Spectra, Instruction & Safety Manual Use of the Hand Control Locking the hand control setting The hand control allows the patient to control the machine as appropriate. Simultaneously press the keys to lock the hand control, The display reads “LOCK”, you can not change the parameters, if you try the display reads “LOCK”.
  • Page 9: Quick Start

    Use of the Hand Control Quick Start Begining Keys to press Display Remarks Switch the Unit ON KINETIC Spectra V01.0 Check if the hand Please wait control is not blocked Kinetic (See page 7) STOP If the warm up mode WARMUP is selected, switch OFF STOP...
  • Page 10: How To Adjust The Basic Parameters Of The Movement

    Kinetic Spectra, Instruction & Safety Manual Use of the Hand Control How to adjust the basic parameters of the movement Begining Keys to press Display Remarks To stop the unit KINETIC Check if the hand STOP control is not blocked...
  • Page 11: How To Modify Programs Program Mode

    Use of the Hand Control How to modify programs PROGRAM MODE Begining Keys to press Display Remarks To switch off the unit Check if the hand control is not blocked (See page 7) Press the two keys at KINETIC Welcome text during the same time and Spectra V01.0 3 seconds...
  • Page 12 Kinetic Spectra, Instruction & Safety Manual Use of the Hand Control How to modify programs PROGRAM MODE (continued) Begining Keys to press Display Remarks To change the limit PROGRAM The new value of movement warmup blinks To validate the new...
  • Page 13 Use of the Hand Control Comments: When a program has been deleted, the display shows Program table: Program Treatment Flexion Extension Speed Force Flexion Extension Timer Number Mode Limit Limit Pause Pause 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
  • Page 14: Using Programs

    Kinetic Spectra, Instruction & Safety Manual Use of the Hand Control Using Programs The KINETEC Spectra allows you to store up to 16 programs, including the type of treatment, ROM, speed, load, pauses and timer. The original parameter values of the program are empty.
  • Page 15: Reading The Values Of A Program

    Use of the Hand Control Reading the values of a program: example SPEED Begining Keys to press Display Remarks To stop the unit PROGRAM Check of the hand STOP control is not blocked (See page 7) To access the program PROGRAM The program number mode...
  • Page 16: How To Use The Warm Up Key

    Kinetic Spectra, Instruction & Safety Manual Use of the Hand Control How to use the WARM UP key Begining Keys to press Display Remarks To stop the unit KINETIC Check of the hand STOP control is not blocked (See page 7)
  • Page 17 Use of the Hand Control How to defi ne the patient tolerance • At the start of a session This function, MODULATION Mode is a way to set within the tolerance of a patient at the beginning of a session. Set up the patient and proceed as below: Begining Keys to press...
  • Page 18 Kinetic Spectra, Instruction & Safety Manual Use of the Hand Control How to defi ne the patient tolerance • During the session Begining Keys to press Display Remarks The unit is running KINETIC Check of the hand control is not blocked...
  • Page 19: Session Time

    Use of the Hand Control Session Time This function shows the running time (in minutes) of the session (motor functioning). TIME 02H25 • It directly accessible by the key, the display shows 30 STOP 15 90 This counter is reset each time the unit is switched ON. 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
  • Page 20: Patient Pad Kit

    Kinetic Spectra, Instruction & Safety Manual Use of the Kinetec Patient Pad Kit How to defi ne the patient tolerance • During the session FOR OPTIIMAL HYGIIENE,, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIIENT.. Eacch ccovverr iiss prrovviided wiitth a llabell tto rreccorrd tthe pattiientt’’ss name CLEANING: Sterilization of the pads ( if necessary): Sterilized at 134 °C during 18 minutes.
  • Page 21: Setting Up The Patient

    Setting up the patient Place the KINETEC Spectra machine in a position that will be comfortable for the patient. Measure in cm or inches the length of the patient’s femur (L); adjust the thigh support to this measurement using knobs (5) To install the patient on the KINETEC Spectra machine.
  • Page 22: Product Information

    Kinetic Spectra, Instruction & Safety Manual Product information Maintenance After 2,000 hours of operation, KINETEC Spectra requires a few lubrication and maintenance operations (lubrication of the joints, pointer stops and ball screws). The need for maintenance is indicated by display of the message SERV. MOTOR when the system is switched on.
  • Page 23: Cleaning

    Product information Cleaning Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply. Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and metal components). In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.
  • Page 24: Symbols Used

    Kinetic Spectra, Instruction & Safety Manual Product information Symbols used TYPE B device (protection against Speed electric shocks Caution (consult the accompanying Timer documents) STOP (power off) Force ON (power on) Pause Start movement Increase Stop movement Decrease Program access...
  • Page 25: Maintenance Sheet

    Product information Maintenance sheet Serial Number: Purchasing Date: Date: ..............Operation Done: ............ Technician:............................Running Time (see page 22): ......................Exchange Parts: ........................................................... Date: ..............Operation Done: ............ Technician:............................Running Time (see page 22): ......................
  • Page 26 Kinetic Spectra, Instruction & Safety Manual 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
  • Page 27 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
  • Page 28 PO Box 82 Bentley WA 6982 Authorised Distributor T 08 9351 7500 F 08 9350 5299 info@unicarehealth.com.au www.unicarehealth.com.au...

Table of Contents