Summary of Contents for National Instruments PXIe-7865
Page 1
This document contains the verification and adjustment procedures for NI PXIe-7865/7866. For more information about calibration solutions, visit ni.com/calibration. Software Calibrating the NI PXIe-7865/7866 requires installation of the correct software version on the calibration system. • LabVIEW 2019 SP1 or later •...
R Series devices. NI CB-68LP, NI TBX-68, The CB-68LP, CB-68LPR, and TBX-68 are NI CB-68LPR low-cost termination accessories with 68 screw terminals for easy signal connection to the R Series devices. 2 | ni.com | NI PXIe-7865 and NI PXIe-7866 Calibration Procedure...
As-Found and As-Left Limits The as-found (1-Year) limits are the published specifications for the NI PXIe-7865/7866. NI uses these limits to determine whether the NI PXIe-7865/7866 meets the specifications when it is received for calibration. Use the as-found limits during initial verification.
Page 4
Verification The following performance verification procedures describe the sequence of operation and provide test points required to verify the NI PXIe-7865/7866. The verification procedures assume that adequate traceable uncertainties are available for the calibration references. Analog Input Accuracy Verification Complete the following procedure to determine the As-Found status of the NI PXIe-7865/7866.
Page 6
Counter Verification Complete the following procedure to determine the As-Found status of the NI PXIe-7865/7866. Connect the NI R Series device to the counter as shown in the following figure. 6 | ni.com | NI PXIe-7865 and NI PXIe-7866 Calibration Procedure...
Page 7
If the NI PXIe-7865/7866 device successfully passed each of the verification procedures within the 24-hour test limits, then an adjustment is not required. • If the NI PXIe-7865/7866 device is not within the 1-year test limits, then adjustment is recommended to guarantee the published specifications for the next year.
Page 8
. You do not need to input values for Comments or Current Calibrator Value. Determine the value of the NI PXIe-7865/7866 device onboard reference by completing the following steps. Call . Set the Action to nirio78xxCalibrationAdjustExtCal.vi...
Page 10
NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the file on your media, or the National Instruments Patent Notice at . You can find patents.txt ni.com/patents...
Need help?
Do you have a question about the PXIe-7865 and is the answer not in the manual?