Page 1
Operating Instructions Photometer 4040 ROBERT RIELE GmbH & Co KG Software Version 2.2 Documentation Version 08.2017...
Page 2
Directive 98/79/EC on in vitro diagnostic medical devices (IVDD). Furthermore, the Photometer 4040 is manufactured according to the special safety requirements for IVD medical devices stated in EN 61010. ROBERT RIELE GmbH & Co KG Page 2 GENERAL NOTES...
Page 3
If a sample is spilled on the device, wipe up immediately and apply disinfectant. Waste Handle liquid waste properly, according to legislation on water pollution, and on the treatment of drainage and waste matter. ROBERT RIELE GmbH & Co KG Page 3 GENERAL NOTES...
Page 4
+49 (0)30 404 05 29 E-mail: info@riele.de www.riele.de QUALITY MANAGEMENT SYSTEM ROBERT RIELE GmbH & Co KG maintains a quality management system according to ISO 13485, certified by mdc medical device certification GmbH. ROBERT RIELE GmbH & Co KG Page 4 GENERAL NOTES...
Page 6
SHORT SPECIFICATIONS ......................68 11.5 TECHNICAL SPECIFICATIONS ..................... 69 ACCESSORIES AND SPARE PARTS ............71 METHOD LIST .................... 72 13.1 BASIC METHODS.......................... 72 13.2 LIST OF USER SPECIFIC METHODS ..................73 ROBERT RIELE GmbH & Co KG Page 6 CONTENTS...
Page 7
FLASH MEMORY technology: The operating system can be updated with program novelties and/or improvements comfortably, without having to open the equipment (please ask distributor for further information). ROBERT RIELE GmbH & Co KG Page 7 1 - INTRODUCTION TO PHOTOMETER 4040...
Insert printer paper reel into the axis guide. Push the printer paper through the slot in the printer cover and close the printer with the cover. ROBERT RIELE GmbH & Co KG Page 9 2 - INSTALLATION...
4040+e_22.docx / 06.10.17 OPERATING ELEMENTS FRONT printer cover working area touchscreen BACK ROBERT RIELE GmbH & Co KG Page 10 3 - OPERATING ELEMENTS...
Trigger a normal measuring by [RESULT]. 3.5.2 Reducing the minimum measuring volume The minimum measuring volume can be reduced from 500 µl to 250 µl. Spacers can be inserted into the measuring cells. ROBERT RIELE GmbH & Co KG Page 11 3 - OPERATING ELEMENTS...
METHOD NEW /CHANGE / COPY (chapter 4.3 - Method editor). The transmission of a method collection is possible by PC with special software. Further information: Application sheets of reagent manufacturers ROBERT RIELE GmbH & Co KG Page 12 4 - PROGRAM SELECTION...
1 0 : 1 4 Line feed [LF] Pressing [LF] in the main menu triggers a line feed in the case of activated printer. Several lines advanced continuous pressure on [LF]. ROBERT RIELE GmbH & Co KG Page 14 4 - PROGRAM SELECTION...
When the [Z] button returns to blue the result button [R] may be pushed. ROBERT RIELE GmbH & Co KG Page 15 4 - PROGRAM SELECTION...
Page 16
If R^2 is activated, the term LINEAR or NON- LINEAR is shown in the upper right corner. By [PRINT] a graphical printout is generated. By [LIST] all data points are shown sequentially. ROBERT RIELE GmbH & Co KG Page 16 4 - PROGRAM SELECTION...
For a homogeneous solution the input of a delay before a measuring is possible at all methods. All delay times can be aborted by pressing the touchscreen at any place for a long time. ROBERT RIELE GmbH & Co KG Page 17 6 - CALCULATION PROCEDURES...
The bichromatic wavelength might be not included in the standard set of filters. The bichromatic wavelength can be set after calling a method (chapter 6 METHOD EDITOR Fig. 6.5). ROBERT RIELE GmbH & Co KG Page 18 6 - CALCULATION PROCEDURES...
Only if MAX. VALUE is set and the sign of the measured value is not equal to the sign of the entered MAX. VALUE a message RANGE +/- is shown. The parameters MIN. VALUE, MAX. VALUE and MIN. R^2 are deactivated entering a zero value. ROBERT RIELE GmbH & Co KG Page 19 6 - CALCULATION PROCEDURES...
Page 20
[LIST] all data points are shown sequentially. Sequential View of data points after confirming [DIAG] and [LIST] By [NEXT] for each data point the numeration the time [s] and the extinction is shown. ROBERT RIELE GmbH & Co KG Page 20 6 - CALCULATION PROCEDURES...
Page 21
0.0365 deltas these are marked with a “x”. If this procedure does not lead to an improvement deltas reconsidered calculation procedure and none of the deltas are marked. ROBERT RIELE GmbH & Co KG Page 21 6 - CALCULATION PROCEDURES...
Stored results can be output through serial interface (chapter 8.2.8 - Stored results). When memory is full oldest test results will be overwritten or you can send all results through the serial interface and then delete them. ROBERT RIELE GmbH & Co KG Page 22 6 - CALCULATION PROCEDURES...
TRANSM., T ..Transmission in % ......At Fixed Time: incubation time in seconds ......At Fixed Time: reaction time in seconds ......At Kinetic: time per delta in seconds ROBERT RIELE GmbH & Co KG Page 23 6 - CALCULATION PROCEDURES...
CP-No......Number of the calculation procedure (chapter 7 - METHOD EDITOR) Characteristic ....Name of the calculation procedure (chapter 13.1 - BASIC METHOD) Calculation formula ..Calculation basis of basic method ROBERT RIELE GmbH & Co KG Page 24 6 - CALCULATION PROCEDURES...
Switch printer ON/OFF / printout of current time by [TIME] [DETAIL] Display/printout of detailed results [M-STD] Multi standard functions [E1/E2] Change to measuring E2 (chapter 6.4.14) [RETRY] Repeat a measurement [RETURN] Return to normal functions ROBERT RIELE GmbH & Co KG Page 25 6 - CALCULATION PROCEDURES...
Detail print-out of ABS.1, ABS.2 and deltas. A detail print-out of ABS.1, ABS.2 and deltas can be made after each measurement with [MODE] [MODE] [DETAIL] (see chapter 6.1.7 Fundamental to the Kinetic…). ROBERT RIELE GmbH & Co KG Page 36 6 - CALCULATION PROCEDURES...
Insert / measure sample R^2: 0.9997 A detail print-out of ABS.1, ABS.2 and deltas can be made after each measurement with [MODE] [MODE] [DETAIL] (see chapter 6.1.7 Fundamental to the Kinetic…). ROBERT RIELE GmbH & Co KG Page 37 6 - CALCULATION PROCEDURES...
T3: Delay, optional : manual process A: Start of time controlled measuring E1: Start of first course (E1) E2: Start of second course (E2) Fig. 6.4.14.1: time controlled measuring ROBERT RIELE GmbH & Co KG Page 39 6 - CALCULATION PROCEDURES...
Page 40
Measure all samples E2 (maximum 25) 0.000 1.530 0.000 1.384 RESULT Results based on the differences of the measured samples 0.446 0.392 0.307 Show the results by [MODE] [MODE] [DETAIL] ROBERT RIELE GmbH & Co KG Page 40 6 - CALCULATION PROCEDURES...
- - - - - - - - - - - - Insert / measure zero solution MEASURE ZERO µmol/L mg/dl The result will be displayed in two units of measurement 4.859 7.83 5.865 9.45 Insert / measure sample ROBERT RIELE GmbH & Co KG Page 41 6 - CALCULATION PROCEDURES...
3 7 . 0 ° C i 3 7 . 0 ° C 0 7 / 2 4 / 2 0 1 7 1 0 : 1 4 info line Fig. 5.4.16.2: screen during the method ROBERT RIELE GmbH & Co KG Page 42 6 - CALCULATION PROCEDURES...
Page 43
Insert / measure samples with reagent 2 0.285 0.165 The user is guided through the measuring procedure on the 0.165 0.165 screen. RESULT The result will be displayed -0.116 -0.116 -0.116 ROBERT RIELE GmbH & Co KG Page 43 6 - CALCULATION PROCEDURES...
This function is very useful if a basic method with setting parameters successfully tested. These parameters can be saved as a new method starting from no [EXIT] Return to main menu ROBERT RIELE GmbH & Co KG Page 44 7 - METHOD EDITOR...
Page 45
8 - M A X . V A L U E then also all data and reserved memory of this method in the quality control are deleted! P 3 / 3 ROBERT RIELE GmbH & Co KG Page 45 7 - METHOD EDITOR...
Page 46
In order to set the value to zero enter e.g. “0.0”. [ESC] End input and save curve data For measuring in a multi-standard method at least 2 bases with A and C must be defined! ROBERT RIELE GmbH & Co KG Page 46 7 - METHOD EDITOR...
8.2.10 User name 8.2.11 Error list 8.2.12 Key signal ON / OFF 8.2.13 Touchscreen adjustment 8.2.14 Date / Time 8.2.15 Language 8.2.16 8.2.17 ADC counts (Optic) Service tools 8.2.18 ROBERT RIELE GmbH & Co KG Page 47 8- UTILITY PROGRAMS...
P L E A S E W A I T program level. Monthly executed the optic adjust- ment compensates possible deviations of measuring accuracy environmental influences. ROBERT RIELE GmbH & Co KG Page 48 8- UTILITY PROGRAMS...
[PRINT CURVE] After inquiry of the curve number print-out on the internal printer or via the serial interface. For measuring in a multi-standard method at least 2 bases with A and C must be defined! ROBERT RIELE GmbH & Co KG Page 49 8- UTILITY PROGRAMS...
- - - - - - - - - - - - When printer ON via [MODE] [PRN] 0.675 19.8 after a measurement the current time can be TIME: 11:21:32 printed out by [TIME]. ROBERT RIELE GmbH & Co KG Page 50 8- UTILITY PROGRAMS...
Save setting temporarily up to next switch- off of the device by [EXIT] Table 8.2.5.1 Serial Method ID-no. Sample Result Temperature User Date Time control 2250 12345 15.5 [user name] 09/08/09 09:30:47 ROBERT RIELE GmbH & Co KG Page 51 8- UTILITY PROGRAMS...
Except the basic methods all methods can be connected with a quality control. The method-typical data of a control serum are entered via the method editor (see chapter 7 METHOD EDITOR). ROBERT RIELE GmbH & Co KG Page 52 8- UTILITY PROGRAMS...
Page 53
Print-out of daily memory for serum 1: [PRINT] Print all readings *** DAILY MEMORY ***S1 * [EXIT] Return to QC menu 4040+ PHOTOMETER # 1001 V2.2a dd/mm/yy D LAB.: RIELE BERLIN DATE: 07/17/17 ROBERT RIELE GmbH & Co KG Page 53 8- UTILITY PROGRAMS...
Page 54
Delete all measuring data of the monthly memory of the selected method and confirm again (e.g. at change of serum) [MORE] Change to output dialog [EXIT] Return to QC menu ROBERT RIELE GmbH & Co KG Page 54 8- UTILITY PROGRAMS...
1: ENGLISH 2: GERMAN KEY SIGNAL ON M.COUNTER : T.COUNTER : 334:45 S.COUNTER : ADJUST OPTIC POSITION : P3-1 3850 7246 D2/3 3500 D1/2 3396 PROGR. METHODS STORED RESULTS ROBERT RIELE GmbH & Co KG Page 56 8- UTILITY PROGRAMS...
[] : delete character [] : blank E S C A / 1 E N T [ESC] : finish input without storage [ENT] : finish input with storage ROBERT RIELE GmbH & Co KG Page 57 8- UTILITY PROGRAMS...
Save setting temporarily up to next switch- off of the device by [EXIT]. E X I T S E L E C T The deeper signal tone for error messages remains active in any case. ROBERT RIELE GmbH & Co KG Page 58 8- UTILITY PROGRAMS...
After some seconds a deep signal tone sounds and the text message “TOUCHSCREEN ADJUSTMENT” will be shown at the first line of the greeting screen. Release the touchscreen within one second. Execute the adjustment of the touchscreen as described above. ROBERT RIELE GmbH & Co KG Page 59 8- UTILITY PROGRAMS...
E X I T S E L E C T Save setting permanently by [OK]. The setting is temporarily saved up to next switch-off of the device by [END]. ROBERT RIELE GmbH & Co KG Page 60 8- UTILITY PROGRAMS...
CALIBRATING MEASURING SYSTEM At doubtful measurement results an optic adjustment has to be carried out corresponding to chapter 8.2.1. REPLACEMENT OF PAPER ROLL See chapter 2.4 - LOADING PRINTER PAPER ROBERT RIELE GmbH & Co KG Page 62 9 - MAINTENANCE...
(T) voltage rating: 250 V rated current: 1.6 A marking: T 1.6 A H mains switch fuse holder mains cable connector Connector for Incubator T20 ROBERT RIELE GmbH & Co KG Page 63 9 - MAINTENANCE...
(chapter 8.2.1) ROBERT RIELE GmbH & Co KG Page 64 10 - ERROR MESSAGE / CORRECTION...
Page 65
2 switch off/on device NAK at reception from module 2 switch off/on device filter position out of tolerance contact service partner ROBERT RIELE GmbH & Co KG Page 65 10 - ERROR MESSAGE / CORRECTION...
Page 66
(chapter 8.2.8) BCC error at sending result string send results through serial interface and delete results (chapter 8.2.8) ROBERT RIELE GmbH & Co KG Page 66 10 - ERROR MESSAGE / CORRECTION...
Short-term changes of the operational behavior do not influence the overall function of the device. 11.3 TYPE PLATE For installation pay attention to the specifications on the type plate. ROBERT RIELE GmbH & Co KG Page 67 11 - TECHNICAL DATA...
Mains Supply Range: 100 V up to 240 V at 50/60 Hz Dimensions Length 33 cm x width 34 cm x height 18 cm Weight 5.3 kg Marking ROBERT RIELE GmbH & Co KG Page 68 11 - TECHNICAL DATA...
Model: 4040 Instruction for use: Photometer 4040 , user manual Manufacturer ROBERT RIELE GmbH & Co KG Kurfuerstenstrasse 75-79 13467 Berlin Germany DECLARATION OF CONFORMITY: The above mentioned absorption photometer is in conformity with the following metrological description. Berlin, November 2016 ROBERT RIELE GmbH &...
Page 70
Photodiode The optical path in the UV range is directed from the back to the front of the device. Insert standard cuvette accordingly. Lense Cuvette Filter Lense Photodiode ROBERT RIELE GmbH & Co KG Page 70 11 - TECHNICAL DATA...
4040-034 Spacer for 250 µl measuring volume 4040-035 Spacer for 500 µl measuring volume 1707574001 Top cover small for printer Incubator T12 Incubator T16 Incubator T20 ROBERT RIELE GmbH & Co KG Page 71 12 - ACCESSORIES AND SPARE PARTS...
20 - 250 ..up to 231 user specific methods (chapter 13.2 - LIST OF USER SPECIFIC METHODS as copy master / to be filled out by the user) 13.1 BASIC METHODS ROBERT RIELE GmbH & Co KG Page 72 13 - METHOD LIST...