Riele 5010 V5+ Operating Instructions Manual
Hide thumbs Also See for 5010 V5+:
Table of Contents

Advertisement

Operating Instructions
Photometer 5010
V5+
ROBERT RIELE GmbH & Co KG
Software Version 5.13
Documentation Version 01.2010

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 5010 V5+ and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Riele 5010 V5+

  • Page 1 Operating Instructions Photometer 5010 ROBERT RIELE GmbH & Co KG Software Version 5.13 Documentation Version 01.2010...
  • Page 2 IVD medical devices stated in DIN EN 61010, testified by TÜV Rheinland Group. The Photometer 5010 fulfils the EMC immunity requirements for laboratory use equipment according to the EMC standard EN 61326. ROBERT RIELE GmbH & Co KG Page 2 GENERAL NOTES...
  • Page 3 If a sample is spilled on the device, wipe up immediately and apply disinfectant. Waste Handle liquid waste properly, according to legislation on water pollution, and on the treatment of drainage and waste matter. ROBERT RIELE GmbH & Co KG Page 3 GENERAL NOTES...
  • Page 4 QUALITY MANAGEMENT SYSTEM ROBERT RIELE GmbH & Co KG maintains a quality management system according to EN ISO 9001:2008, and EN ISO 13485:2003+AC:2007 certified by TÜV Rheinland Product Safety GmbH. ROBERT RIELE GmbH & Co KG Page 4...
  • Page 5: Table Of Contents

    Calculation procedure 11 (KIN/F/Rb)..................34 5.4.12 Calculation procedure 12 (KIN/S/Rb)..................35 5.4.13 Calculation procedure 13 (TRANSMISSION) ................36 5.4.14 Calculation procedure 14 (C/F Delta) ..................37 5.4.15 Calculation Procedure 15 (C/F 3 WL) ..................39 METHOD EDITOR..................40 ROBERT RIELE GmbH & Co KG Page 5 CONTENTS...
  • Page 6 MINIMAL OPERATION QUALITY ....................67 10.3 TYPE PLATE ..........................67 10.4 SHORT SPECIFICATIONS......................68 10.5 CERTIFICATES ..........................69 ACCESSORIES AND SPARE PARTS............72 METHOD LIST ....................73 12.1 BASIC METHODS...........................73 12.2 LIST OF USER SPECIFIC METHODS ..................74 ROBERT RIELE GmbH & Co KG Page 6 CONTENTS...
  • Page 7: Introduction To Photometer 5010

    Photometer 5010 is future-proof by FLASH MEMORY technology: The operating system can be updated with program novelties and/or improvements comfortably, without having to open the equipment (chapter 7.2.5 - Serial interface – DOWNLOAD). ROBERT RIELE GmbH & Co KG Page 7 1 - INTRODUCTION TO PHOTOMETER 5010...
  • Page 8: Installation

    5 0 1 0 out. V 5 . 1 3 a 0 2 / 0 9 / 0 9 w w w . r i e l e . d e ROBERT RIELE GmbH & Co KG Page 8 2 - INSTALLATION...
  • Page 9: Loading Printer Paper

    Press [LF] several times for line-feed until the paper has a length of about 5 cm. • Insert printer paper reel into the axis guide. • Push the printer paper through the slot in the printer cover and close the printer with the cover. ROBERT RIELE GmbH & Co KG Page 9 2 - INSTALLATION...
  • Page 10: Operating Elements

    5010e_513.doc / 22.01.10 OPERATING ELEMENTS FRONT printer cover working area touchscreen sipping lever [P] aspiration tube BACK ROBERT RIELE GmbH & Co KG Page 10 OPERATING ELEMENTS...
  • Page 11: Touchscreen

    WORKING AREA pump tube with joints REF 5010-050 outlet tube cuvette REF 5010-066 waste tubing REF 1704834001 aspiration tubing REF 5010-065 joint inlet tube cuvette REF 1707175001 ROBERT RIELE GmbH & Co KG Page 11 3 - OPERATING ELEMENTS...
  • Page 12: Cuvettes And Cuvette Adaptor

    To work in optimized volume mode use the function Volume optimized ON / OFF (chapter 7.2.4.3.2). This function makes it possible, for example, to pump two consecutive times 500 µl from 1000 µl sample volume. ROBERT RIELE GmbH & Co KG Page 12...
  • Page 13: Working With Standard Cuvettes

    Switch on the pump by the function Pump ON / OFF (chapter 7.2.4.1). Switch on or off bubble detector by the function Bubble detector ON / OFF (chapter 7.2.4.3). After the installation execute a Pump calibration (chapter 7.2.4.2). ROBERT RIELE GmbH & Co KG Page 13 3 - OPERATING ELEMENTS...
  • Page 14: Program Selection

    METHOD NEW /CHANGE / COPY (chapter 4.3 - Method editor). The transmission of a method collection is possible by PC with special software. Further information: Application sheets of reagent manufacturers ROBERT RIELE GmbH & Co KG Page 14 PROGRAM SELECTION...
  • Page 15: Measurement With Basic Methods

    E X I T Further information: 3 7 . 0 5 C 0 9 / 0 8 / 0 9 1 1 : 3 9 Chapter: 6 - METHOD EDITOR ROBERT RIELE GmbH & Co KG Page 15 4 - PROGRAM SELECTION...
  • Page 16: Utility Programs

    U T I L I T I E S L A M P 3 7 . 0 5 C 0 9 / 0 8 / 0 9 1 1 : 3 9 ROBERT RIELE GmbH & Co KG Page 16 4 - PROGRAM SELECTION...
  • Page 17: Calculation Procedures

    For a homogeneous solution the input of a delay before a measuring is possible at all methods. • All delay times can be cancelled by pressing the aspiration tube [P] for a long time. ROBERT RIELE GmbH & Co KG Page 17 CALCULATION PROCEDURES...
  • Page 18: Fundamental To The Methods With Standard

    The bichromatic wavelength might be not included in the standard set of filters. The bichromatic wavelength can be set after calling a method (chapter 6 METHOD EDITOR Fig. 6.5). ROBERT RIELE GmbH & Co KG Page 18 5 - CALCULATION PROCEDURES...
  • Page 19: Fundamental To The Kinetic

    Only if MAX. VALUE is set and the sign of the measured value is not equal to the sign of the entered MAX. VALUE a message RANGE +/- is shown. The parameters MIN. VALUE, MAX. VALUE and MIN. R^2 are deactivated entering a zero value. ROBERT RIELE GmbH & Co KG Page 19 5 - CALCULATION PROCEDURES...
  • Page 20: Fundamental To The Methods With Reagent Blank

    Stored results can be output through serial interface (chapter 7.2.19). • When memory is full oldest test results will be overwritten or you can send all results through the serial interface and then delete them. ROBERT RIELE GmbH & Co KG Page 20 5 - CALCULATION PROCEDURES CALCULATION PROCEDURES...
  • Page 21: Abbreviations

    TRANSM., T..Transmission in % ......At Fixed Time: incubation time in seconds ......At Fixed Time: reaction time in seconds ......At Kinetic: time per delta in seconds ROBERT RIELE GmbH & Co KG Page 21 5 - CALCULATION PROCEDURES...
  • Page 22: Survey Of The Methods

    CP-No......Number of the calculation procedure (chapter 6 - METHOD EDITOR) Characteristic ....Name of the calculation procedure (chapter 12.1 - BASIC METHOD) Calculation formula ..Calculation basis of basic method ROBERT RIELE GmbH & Co KG Page 22 5 - CALCULATION PROCEDURES...
  • Page 23: Description Of Method Procedures

    Quality Control functions. [PRN] Switch printer on / off. [DETAIL] Print detailed results on kinetic measurement. [LAMP] Lamp Protection (chapter 4.5). [M-STD] Multi standard functions. [RETURN] Return to normal functions. ROBERT RIELE GmbH & Co KG Page 23 5 - CALCULATION PROCEDURES...
  • Page 24: Calculation Procedure 1 (C/F)

    - - - - - - - - - - - - Insert / measure sample 0.843 24.8 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 24 5 - CALCULATION PROCEDURES...
  • Page 25: Calculation Procedure 2 (C/F/Rb)

    - - - - - - - - - - - - Insert / measure sample 1.340 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 25 5 - CALCULATION PROCEDURES...
  • Page 26: Calculation Procedure 3 (C/F/Sb)

    - - - - - - - - - - - - Insert / measure sample blank 1.033 4.23 Insert / measure sample Sb[A]: 0.702 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 26 5 - CALCULATION PROCEDURES...
  • Page 27: Calculation Procedure 4 (C/F/Sbrb)

    - - - - - - - - - - - - Insert / measure sample blank 0.646 Insert / measure sample Sb[A]: 0.497 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 27 5 - CALCULATION PROCEDURES...
  • Page 28: Calculation Procedure 5 (C/S)

    - - - - - - - - - - - - Insert / measure sample 1.357 6.92 - - - - - - - - - - - - Insert / measure sample 1.582 8.07 ROBERT RIELE GmbH & Co KG Page 28 5 - CALCULATION PROCEDURES...
  • Page 29: Calculation Procedure 6 (C/S/Rb)

    - - - - - - - - - - - - Insert / measure sample 1.118 149.6 - - - - - - - - - - - - Insert / measure sample 2.006 281.2 ROBERT RIELE GmbH & Co KG Page 29 5 - CALCULATION PROCEDURES...
  • Page 30: Calculation Procedure 7 (C/S/Sb)

    - - - - - - - - - - - - Insert / measure sample blank 2.223 197.2 Insert / measure sample Sb[A]: 0.193 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 30 5 - CALCULATION PROCEDURES...
  • Page 31: Calculation Procedure 8 (C/S/Sbrb)

    - - - - - - - - - - - - Insert / measure sample blank 1.394 8.39 Insert / measure sample Sb[A]: 0.329 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 31 5 - CALCULATION PROCEDURES...
  • Page 32: Calculation Procedure 9 (Ftk/F/Rb)

    Insert / measure sample DELTA [A]: 0.410 - - - - - - - - - - - - 0.829 1381.2 Insert / measure sample DELTA [A]: 0.502 ROBERT RIELE GmbH & Co KG Page 32 5 - CALCULATION PROCEDURES...
  • Page 33: Calculation Procedure 10 (Ftk/S/Rb)

    Insert / measure sample DELTA [A]: 1.103 - - - - - - - - - - - - 0.329 12.84 Insert / measure sample DELTA [A]: 1.310 ROBERT RIELE GmbH & Co KG Page 33 5 - CALCULATION PROCEDURES...
  • Page 34: Calculation Procedure 11 (Kin/F/Rb)

    A detail print-out of ABS.1, ABS.2 and deltas can be made after -0.0867 each measurement with [MODE] [MODE] [DETAIL]. -0.0899 -0.0871 -0.0874 -0.1087 DELTA REG:-0.0908 - - - - - - - - - - - - ROBERT RIELE GmbH & Co KG Page 34 5 - CALCULATION PROCEDURES...
  • Page 35: Calculation Procedure 12 (Kin/S/Rb)

    - - - - - - - - - - - - 0.175 81.8 Insert / measure sample R^2: 0.9997 A detail print-out of ABS.1, ABS.2 and deltas can be made after each measurement with [MODE] [MODE] [DETAIL]. ROBERT RIELE GmbH & Co KG Page 35 5 - CALCULATION PROCEDURES...
  • Page 36: Calculation Procedure 13 (Transmission)

    UNIT: Method procedure: Insert / measure zero solution MEASURE 100% ABS. RESULT Insert / measure sample 0.329 46.9 Insert / measure sample 1.004 Insert / measure sample 2.020 ROBERT RIELE GmbH & Co KG Page 36 5 - CALCULATION PROCEDURES...
  • Page 37: Calculation Procedure 14 (C/F Delta)

    T3: Delay, optional : manual process A: Start of time controlled measuring E1: Start of first course (E1) E2: Start of second course (E2) Fig. 5.4.14.1: time controlled measuring ROBERT RIELE GmbH & Co KG Page 37 5 - CALCULATION PROCEDURES...
  • Page 38 Sb[A] S[A]E2 0.000 1.458 0.000 1.530 0.000 1.384 Results based on the differences of the measured samples RESULT 0.446 Show the results by [MODE] [MODE] [DETAIL] 0.392 0.307 ROBERT RIELE GmbH & Co KG Page 38 5 - CALCULATION PROCEDURES...
  • Page 39: Calculation Procedure 15 (C/F 3 Wl)

    The result will be displayed in two units of measurement µmol/L mg/dl 4.859 7.83 Insert / measure sample - - - - - - - - - - - - 5.865 9.45 ROBERT RIELE GmbH & Co KG Page 39 5 - CALCULATION PROCEDURES...
  • Page 40: Method Editor

    E X I T L A S T with new setting parameters was successfully tested. These parameters can be saved as a new method starting from no 20. [EXIT] Return to main menu ROBERT RIELE GmbH & Co KG Page 40 METHOD EDITOR...
  • Page 41 8 - M A X . V A L U E method in the quality control are deleted! P 3 / 3 ROBERT RIELE GmbH & Co KG Page 41 6 - METHOD EDITOR D EDITOR...
  • Page 42 “0.0”. [ESC] End input and save curve data For measuring in a multi-standard method at least 2 bases with A and C must be defined! ROBERT RIELE GmbH & Co KG Page 42 6 - METHOD EDITOR D EDITOR...
  • Page 43: Utility Programs

    User name 7.2.12 Error list 7.2.13 Key signal ON / OFF 7.2.14 Touchscreen adjustment 7.2.15 Date / Time 7.2.16 Language 7.2.17 ADC counts (Optic) 7.2.18 Send results 7.2.19 Service tools 7.2.20 ROBERT RIELE GmbH & Co KG Page 43 UTILITY PROGRAMS...
  • Page 44: Description Of Utility Programs

    P L E A S E W A I T Monthly executed the dark adjustment compensates possible deviations measuring accuracy due to environmental influences. ROBERT RIELE GmbH & Co KG Page 44 7- UTILITY PROGRAMS...
  • Page 45: Multi-Standard Functions

    For measuring in a multi-standard method at least 2 bases with A and C must be defined! ROBERT RIELE GmbH & Co KG Page 45 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 46: Printer On / Off

    Save setting permanently by [OK] Save setting temporarily up to next switch-off of the device by [EXIT] E X I T S E L E C T ROBERT RIELE GmbH & Co KG Page 46 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 47: Pump Menu

    W A S H although the pump tube is connected and 1000 µl were sucked in, execute a media test (see 7.2.4.4). Afterwards the pump calibration can be repeated. ROBERT RIELE GmbH & Co KG Page 47 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 48 For example, this function makes it possible to pump two consecutive times 500 µl from a 1000 µl sample volume. ROBERT RIELE GmbH & Co KG Page 48 7- UTILITY PROGRAMS...
  • Page 49 Then increase the sensitivity level by one further step. Empty the aspiration tube by [Sipping lever P]. The display changes from “WATER” to “AIR”. The adjustment is permanently saved by [E]. ROBERT RIELE GmbH & Co KG Page 49 7- UTILITY PROGRAMS...
  • Page 50: Serial Interface

    Deactivate remote control by sipping lever [P]. E X I T S T A R T By [EXIT] the program returns to the superordinate menu item. ROBERT RIELE GmbH & Co KG Page 50 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 51: Quality Control

    1/2 [QC EDP] - not implemented - 3 7 . 0 5 C 0 9 / 0 8 / 0 9 1 1 : 3 9 ROBERT RIELE GmbH & Co KG Page 51 7- UTILITY PROGRAMS...
  • Page 52 [EXIT] Return to QC menu PHOTOMETER 5010 #7000 V5.13a 02/09/09 D LAB.: RIELE GMBH+CO KG DATE: 09/08/09 TIME: 08:30:03 25 GLUCOSE 13.03 21 HDL-C 27 UREA COL 197.2 29 CK-MB 1128.1 31 GOT 189.9 ROBERT RIELE GmbH & Co KG Page 52 UTILITY PROGRAMS...
  • Page 53 Delete all measuring data of the monthly memory of the selected method and confirm again (e.g. at change of serum) [MORE] Change to output dialog [EXIT] Return to QC menu ROBERT RIELE GmbH & Co KG Page 53 7- UTILITY PROGRAMS...
  • Page 54 02.04.06 12.13 -2s 02.03.06 12.65 -2s 02.02.06 13.11 -1s 02.01.06 13.88 +1s 01.31.06 13.51 -1s 01.30.06 13.24 -1s 01.29.06 12.50 -2s 01.28.06 12.74 -2s 01.27.06 13.30 -1s ROBERT RIELE GmbH & Co KG Page 54 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 55: Settings Printout

    MS AIR MS/50ul B-DET. VOLUME OPTIMIZED OFF BATTERY: VDC LAMP 12.12V ADC CORRECTION STACK TEST BOOST nm (STX-ETX-BCC) LANUAGE 1: ENGLISH 2: GERMAN KEY SIGNAL ON PROGR. METHODS ROBERT RIELE GmbH & Co KG Page 55 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 56: Data Logging

    5/ 3 stands for 3 stored readings for ID no 5. DATE: 09/08/09 METHOD GLUCOSE 12.17 mmol/l [EXIT] Return to the utilities. Report of logging memory : **** LOGGING REPORT **** PHOTOMETER 5010 #7000 DATE: 09/08/09 ROBERT RIELE GmbH & Co KG Page 56 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 57: Temperature On / Off

    Enter the password “5010”. Enter the actual temperature in TEMP. ACTUAL four-digit in °C (e.g. 36.90) and confirm by [E]. ROBERT RIELE GmbH & Co KG Page 57 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 58: Laboratory Name

    [ ] : blank E S C A / 1 E N T • [ESC] : finish input without storage • [ENT] : finish input with storage ROBERT RIELE GmbH & Co KG Page 58 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 59: Error List

    Save the adjustment by [OK]. Reject the adjustment by [EXIT]. S A V E A D J U S T M E N T E X I T ROBERT RIELE GmbH & Co KG Page 59 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 60: Date / Time

    E X I T S E L E C T Save setting permanently by [OK]. The setting is temporarily saved up to next switch-off of the device by [END]. ROBERT RIELE GmbH & Co KG Page 60 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 61: Adc Counts (Optic)

    S E R V I C E T O O L S Stop function by [ESC]. P A S S W O R D E S C ROBERT RIELE GmbH & Co KG Page 61 7- UTILITY PROGRAMS Y PROGRAMS...
  • Page 62: Maintenance

    Put in the ribbon cassette. Pay attention that the paper is to be taken through inked ribbon and cassette. Take printer paper through the crack of the printer cover and close the printer cover. ROBERT RIELE GmbH & Co KG Page 62...
  • Page 63: Replacement Of Paper Roll

    (T) • voltage rating: 250 V • rated current: 1.6 A • marking: T 1.6 A H mains switch fuse holder mains cable connector ROBERT RIELE GmbH & Co KG Page 63 ERROR MESSAGE / CORRECTION...
  • Page 64: Error Message / Correction

    / lamp; check blank mathematical overflow, at measurement calculation check filter; check standard; check measuring solution ROBERT RIELE GmbH & Co KG Page 64 ERROR MESSAGE / CORRECTION...
  • Page 65 2 switch off/on device check sum error at reception from module 2 switch off/on device NAK at reception from module 2 switch off/on device ROBERT RIELE GmbH & Co KG Page 65 9 - ERROR MESSAGE / CORRECTION...
  • Page 66 (chapter 7.2.19) BCC error at sending result string send results through serial interface and delete results (chapter 7.2.19) ROBERT RIELE GmbH & Co KG Page 66 9 - ERROR MESSAGE / CORRECTION...
  • Page 67: Technical Data

    Short-term changes of the operational behavior do not influence the overall function of the device. 10.3 TYPE PLATE For installation pay attention to the specifications on the type plate. ROBERT RIELE GmbH & Co KG Page 67 10 - TECHNICAL DATA...
  • Page 68: Short Specifications

    Range: 90 V up to 264 V at 50/60 Hz Mains Supply Length 33 cm x width 34 cm x height 18 cm Dimensions 6 kg Weight Marking ROBERT RIELE GmbH & Co KG Page 68 10 - TECHNICAL DATA...
  • Page 69: Certificates

    ROBERT RIELE GmbH & Co KG Kurfuerstenstrasse 75-79 D-13467 Berlin Germany ROBERT RIELE GmbH & Co KG declares under sole responsibility that the product: Product name: Photometer 5010 to which this declaration relates is in conformity with the following standards or other normative documents:...
  • Page 70 Photometer 5010 Model: 5010 Instruction for use: Photometer 5010, user manual Manufacturer ROBERT RIELE GmbH & Co KG Kurfuerstenstrasse 75-79 D-13467 Berlin Germany DECLARATION OF CONFORMITY: The above mentioned absorption photometer is in conformity with the following metrological description. Berlin, March 2006 ROBERT RIELE GmbH &...
  • Page 71 Diaphragm 1 Diaphragm 2 Diaphragm 3 Diaphragm 4 Lens 1 Lens 2 Lens 3 Lens 4 Halogen lamp Glass for heat absorption Photodiode Cuvette Cuvette adaptor Filter ROBERT RIELE GmbH & Co KG Page 71 10 - TECHNICAL DATA TECHNICAL DATA...
  • Page 72: Accessories And Spare Parts

    Secondary calibration standards, four-piece, certified 1704800001 Standard cuvette adaptor 090-063 Thermistor 1707574001 Top cover small for printer 1704834001 Waste tubing, 2 pcs Incubator T12 Incubator T16 ROBERT RIELE GmbH & Co KG Page 72 11 - ACCESSORIES AND SPARE PARTS...
  • Page 73: Method List

    20 - 250..up to 231 user specific methods (chapter 12.2 - LIST OF USER SPECIFIC METHODS as copy master / to be filled out by the user) 12.1 BASIC METHODS ROBERT RIELE GmbH & Co KG Page 73 12 - METHOD LIST...
  • Page 74: List Of User Specific Methods

    5010e_513.doc / 22.01.10 12.2 LIST OF USER SPECIFIC METHODS ROBERT RIELE GmbH & Co KG Page 74 12 - METHOD LIST...

Table of Contents

Save PDF