Photometer 5010 V5+ Version 5.x > SN 7000 Service Manual...
Page 2
BLOCK DIAGRAM (up to SN 8001) RIELE BERLIN Photometer 5010 V5+ 3.1.1 29.08.2007...
Page 3
BLOCK DIAGRAM (from SN 8002) RIELE BERLIN Photometer 5010 V5+ 3.1.2 29.08.2007...
Page 4
S E R V I C E T O O L S E X I T 3 7 . 0 1 C 2 2 . 1 1 . 0 7 1 3 : 0 5 RIELE BERLIN Photometer 5010 V5+ 3.2.1 12.03.2009...
Page 5
7.2.11 User name 7.2.12 Error list 7.2.13 Key signal ON / OFF 7.2.14 Touch panel adjustment 7.2.15 Date / Time 7.2.16 Language 7.2.17 ADC counts (Optic) 7.2.18 Send results 7.2.19 Service tools 7.2.20 RIELE BERLIN Photometer 5010 V5+ 3.2.2 12.03.2009...
Page 6
T O O L S P A S S W O R D Confirm with [E]. The service level will be opened until to the next shutdown of the device. E S C RIELE BERLIN Photometer 5010 V5+ 3.3.1 11.03.2009...
Page 7
ADC COUNTS (OPTIC) ADC COUNTS (DARK ADJ.) BOOST WAVELENGTH CALIBRATION 12V ADC DISTRIBUTOR TEXT INIT AUTOMATIC RESULTS BCC OF METHOD / CURVE CLEAR FACTORS + RB CLEAR ERROR LIST PRINTER TEST STACK TEST RIELE BERLIN Photometer 5010 V5+ 3.3.2 11.03.2009...
Page 8
Resulting factors of methods based on standard not affected Error list not affected Laboratory name not affected User name not affected Data logging not affected Quality control not affected Multi-standard functions not affected Bubble detector RIELE BERLIN Photometer 5010 V5+ 3.3.3 11.03.2009...
Page 9
(chapter 6.1 – Communication protocol). The action ends with a long beep. E X I T D E L S E T [EXIT] leaves the function without any changes. RIELE BERLIN Photometer 5010 V5+ 3.3.4 11.03.2009...
Page 10
D E L . M E M O R Y The action ends with a long beep. [EXIT] leaves the function without any changes. E X I T S T A R T RIELE BERLIN Photometer 5010 V5+ 3.3.5 11.03.2009...
Page 11
S E R I A L N O . 7 0 0 1 A new number can be entered with maximum five digits. The input has to be confirmed with [E]. E S C RIELE BERLIN Photometer 5010 V5+ 3.3.6 11.03.2009...
Page 12
The new value of the wavelength has to be entered. N E W W A V E L E N G T H The input has to be confirmed with [E]. E S C RIELE BERLIN Photometer 5010 V5+ 3.3.7 11.03.2009...
Page 13
• the stability value • the current value of ADC With [SELECT] the temperature level can be chosen. With [OK] the temperature regulation is started. [EXIT] leaves the function without any changes. RIELE BERLIN Photometer 5010 V5+ 3.3.8 11.03.2009...
Page 14
The cursor of the current correction value is blinking. With [NEW] a correction value can be edited. Enter correction value. Confirm with [E]. N E W V A L U E E S C RIELE BERLIN Photometer 5010 V5+ 3.3.9 11.03.2009...
Page 15
OFF) the time slot. The functions F0 to F9 place the E S C filter wheel into the positions 0 to 9. The position 0 corresponds to the filter wheel position at the dark level adjustment. Stop function by [ESC]. RIELE BERLIN Photometer 5010 V5+ 3.3.10 11.03.2009...
Page 16
W A V E L E N G 3 9 0 n m displayed on the right upper corner of the window. The default value is 390 nm. The input has to be confirmed with [E]. E S C RIELE BERLIN Photometer 5010 V5+ 3.3.11 11.03.2009...
Page 17
I N I T [EXIT] leaves the function without any changes. A new distributor text can be set by software “Serial Monitor 5010” (chapter 4.9). E X I T S T A R T RIELE BERLIN Photometer 5010 V5+ 3.3.12 11.03.2009...
Page 18
M E T H O D / C U R V E BCC of a method can be determined. With [M] the BCC of a method can be tested. With [C] the BCC of a curve can be tested. E X I T RIELE BERLIN Photometer 5010 V5+ 3.3.13 11.03.2009...
Page 19
P R I N T E R T E S T The printout is started by invoking [START]. [EXIT] leaves the function without any changes. E X I T S T A R T RIELE BERLIN Photometer 5010 V5+ 3.3.14 11.03.2009...
Page 20
ON = only for development purposes • OFF = normal position The setting has to be confirmed with [OK]. E X I T S E L E C T [EXIT] leaves the function without any changes. RIELE BERLIN Photometer 5010 V5+ 3.3.15 11.03.2009...
Page 21
Pcb RR-121 motor peristaltic pump • ⇒ motherboard pcb RR-120 Pcb RR-126 cuvette adaptor • ⇒ motherboard pcb RR-120 Pcb RR-126 cuvette adaptor • ⇒ Terminal AC / DC power supply Pcb printer controller RIELE BERLIN Photometer 5010 V5+ 3.4.1 22.11.2007...
Page 22
• dimensions [mm] : 5 * 20 • standard: IEC 60127-2/V • time-current characteristic: time lag (T) • voltage rating: 250 V • rated current: 1.6 A • marking: T 1.6 A H RIELE BERLIN Photometer 5010 V5+ 4.1.1 18.08.2010...
Page 23
MAINS POWER CIRCUIT power input 90 V~ - 264 V~ gnye gnye fuses Fa and Fb ground connection mains switch line filter AC / DC power supply REF 5010-016 (terminal block with 7-pins) RIELE BERLIN Photometer 5010 V5+ 4.1.2 18.08.2010...
Page 24
MAINS POWER CIRCUIT power input 90 V~ - 264 V~ gnye gnye fuses Fa and Fb ground connection Mains switch line filter AC / DC power supply REF 5010-086 (terminal block with 9-pins) RIELE BERLIN Photometer 5010 V5+ 4.1.3 18.08.2010...
Page 25
Function Description The unit connects the power input with the AC/DC power supply. Location Backside of Photometer 5010 V5+, base plate mains switch fuses Fa and Fb power input 90 V~ - 264 V~ RIELE BERLIN Photometer 5010 V5+ 4.2.1 13.11.2006...
Page 26
Separating top case and bottom case • see chapter 5.6 Power supply • see chapter 4.4 Cable power supply / printer Installation For installation proceed in reverse order. See also exploded view, chapter „Dismantling“ Terminal block RIELE BERLIN Photometer 5010 V5+ 4.3.1 27.08.2010...
Page 27
AC / DC POWER SUPPLY REF 5010-016/-086 Cable AC / DC power supply Capacitor 10000 µF/25 V RIELE BERLIN Photometer 5010 V5+ 4.3.2 27.08.2010...
Page 28
Separating top case and bottom case • see chapter 5.6 Power supply • see chapter 4.4 Cable power supply / printer Installation For installation proceed in reverse order. See also exploded view, chapter „Dismantling“ Terminal block RIELE BERLIN Photometer 5010 V5+ 4.3.3 27.08.2010...
Page 29
AC / DC POWER SUPPLY REF 5010-016/-086 Cable AC / DC power supply Cable CN2 printer controller / power supply Capacitor 10000 µF/25 V RIELE BERLIN Photometer 5010 V5+ 4.3.4 27.08.2010...
Page 30
Devices up to SN 8002 with pin header. • After installing AC / DC power supply it is strongly recommended to update to the newest operating system version (see Operating Instructions Photometer 5010, chapter 7.2.5.4). RIELE BERLIN Photometer 5010 V5+ 4.3.5 27.08.2010...
Page 31
Beginning with SN 8002 all devices are equipped with this cable at factory, thus no replacement is required. Location The printer controller cable supply should be placed as shown here. RIELE BERLIN Photometer 5010 V5+ 4.4.1 29.08.2007...
Page 32
In case of servicing a device previous to SN 8002 the connection has to be removed: Cable PCB motherboard P23 / PCB printer controller CN2. Remove the plastic cover of the terminal. Unscrew the two lens head screws labeled with V1 and GND. RIELE BERLIN Photometer 5010 V5+ 4.4.2 29.08.2007...
Page 33
This will prevent the cable to interfere with the fan operation after mounting / dismounting the device. Plug the cable with CN2 at the printer controller. RIELE BERLIN Photometer 5010 V5+ 4.4.3 29.08.2007...
Page 34
The Photometer 5010 V5+ has to be dismantled: • see chapter 5.1 Opening • see chapter 5.2 Separating top case and bottom case • see chapter 5.3 Base plate Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.10.1 11.05.2006...
Page 35
Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier Installation For installation proceed in reverse order. Filter wheel with nine filter positions called F1 to F9 Motor mounting plate Filter wheel equipped with interference line filters of standard wavelengths RIELE BERLIN Photometer 5010 V5+ 4.11.1 09.04.2009...
Page 36
The filter wheel is turned by hand so that the corresponding hole is at the cut-out of the motor mounting plate (see page 4.11.1). Tool B With the tool kit filter exchange ( RIELE REF 5010-001 ) a filter exchange can be easily carried out.
Page 37
In case of using filters with a diameter 10 mm an additional spacer ring is needed. Spacer ring for Optical filter Complete unit: optical filter with 10 mm diameter with 10 mm diameter Optical filter with spacer ring RIELE BERLIN Photometer 5010 V5+ 4.11.3 09.04.2009...
Page 38
The Seeger ring can loosely be inserted into the hole in the filter wheel. With the tool B (left illustration) and a pipe wrench the filter and the Seeger ring can be fixed. RIELE BERLIN Photometer 5010 V5+ 4.11.4 09.04.2009...
Page 39
Select amplification level V0 and boost OFF. Stop function by [ESC]. The ADC counts of the new wavelength should be within a range of 15.000 and 600.000. Finish the operation with [ESC]. RIELE BERLIN Photometer 5010 V5+ 4.11.5 09.04.2009...
Page 40
The Photometer 5010 V5+ has to be dismantled (see chapter 5.1 – Opening – item 3 and item 5). Unscrew the cylinder head screw. Remove the lamp shield. Replacement Unscrew the knurled head screw. Remove the holding plate. RIELE BERLIN Photometer 5010 V5+ 4.12.1 12.05.2007...
Page 41
The halogen lamp may be hot. Use tweezers for removing. The picture shows the lamp socket. Attention: The lamp socket is adjusted in the factory. The two screws (A) are not to be solved without a special reason. RIELE BERLIN Photometer 5010 V5+ 4.12.2 12.05.2007...
Page 42
The plastic bag (B) of the halogen lamp can be used for inserting of the lamp. Do not touch the glass of the halogen lamp with fingers. The picture shows the plastic bag, the halogen lamp, the holding plate and the lamp shield. RIELE BERLIN Photometer 5010 V5+ 4.12.3 12.05.2007...
Page 43
Check one after another the used positions of the filter wheel. The ADC counts should be within a range of 20.000 and 600.000 (see chapter 4.40.6 – Adjustment of amplifier stage). Finish the operation with [ESC]. RIELE BERLIN Photometer 5010 V5+ 4.12.4 12.05.2007...
Page 44
The photodiode is covered by a black label. Location The photodiode is situated at the optical unit. The plug P7 is connected with the socket ST7 on the mother board RR-120. Unit photodiode Plug P7 RIELE BERLIN Photometer 5010 v5+ 4.13.1 10.05.2006...
Page 45
Opening – item 1 and item 3 Unplug the unit photodiode from the mother board RR-120. Unscrew the two screws attaching the unit photodiode at the optical unit. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 v5+ 4.13.2 10.05.2006...
Page 46
Separating top case and bottom case • see chapter 5.3 Base plate Unscrew the two cylinder head screws of the PCB RR-126. Unscrew the cylinder head screw of the transistor. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.14.1 10.05.2006...
Page 47
5.3 Base plate • see chapter 5.4 Optical compartment, peristaltic pump, heating unit Two clamp screws are glued. Heat conducting paste Heat sink Mica washer Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.15.1 27.04.2006...
Page 48
To fit the step motor the cover plate of the working area needs to be modified. To remove the cover plate see chapter 5.4.2 – ASSEMBLING / DISMANTLING. Modified cover plate is part of the step motor spare part. RIELE BERLIN Photometer 5010 V5+ 4.16.1 23.07.2013...
Page 49
Separating top case and bottom case • see chapter 5.3 Base plate • see chapter 5.5 Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.18.1 10.05.2006...
Page 50
Opening – item 1, item 2, item 3, item 5 • see chapter 5.5 Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier Adjusting procedure of the light barrier: Loosen the cylinder head screw. Hole on the mounting plate RIELE BERLIN Photometer 5010 V5+ 4.19.1 11.05.2006...
Page 51
Check one after another the used positions of the filter wheel. Finish the operation with [ESC]. RIELE BERLIN Photometer 5010 V5+ 4.19.2 11.05.2006...
Page 52
Open the lid. Disconnect the bubble detector. Withdraw the aspiration tubing from the metal tube. Don’t bend the apiration tubing. Replace the bubble detector. Installation For installation proceed in reverse order. For further information see OPERATING INSTRUCTIONS, chapter 7.2.4 – Pump / bubble detector. RIELE BERLIN Photometer 5010 V5+ 4.20.1 11.05.2006...
Page 53
Loosen clamp screw, open tube connection, remove adaptor. Installation For installation proceed in reverse order. After installation of a new cuvette adaptor a dark level adjustment has to be done by [UTILITIES] [DARK LEVEL ADJUSTMENT] [START]. RIELE BERLIN Photometer 5010 V5+ 4.21.1 28.04.2006...
Page 54
For installation proceed in reverse order. After installation of a new cuvette adaptor a dark level adjustment has to be done by [UTILITIES] [DARK LEVEL ADJUSTMENT] [START]. Holding spring for disp. cuvette REF 1707159001 RIELE BERLIN Photometer 5010 V5+ 4.22.1 28.04.2006...
Page 55
Opening – item 1, item 2, item 3, item 5 Loosen milled nut and remove the connector bubble detector. Remove the green lead with crimped connector belonging to plug P2 pin 4 (type of adaptor). Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.23.1 11.05.2006...
Page 56
Opening – item 1, item 2, item 3, item 5 Remove halogen lamp. For further information see chapter 4.12 – Halogen lamp. Unplug P4. Unscrew the two cylinder head screws. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.24.1 12.05.2006...
Page 58
Separating the top case from the bottom case • see chapter 4.41 Touchscreen Note: When the plastic foil touchscreen (REF 5010-070) is removed from device it might not be reusable. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.30.1 15.05.2006...
Page 59
4.43 PCB RR-122 Display, printer, touch • see chapter 4.34 Plastic foil touchscreen Note: When the plastic foil touchscreen (REF 5010-070) is removed from device it might not be reusable. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.31.1 15.04.2009...
Page 60
Metal tube inlet Location Top case Replacement Remove aspirate tubing from metal tube inlet. Remove metal tube inlet by turning and pulling at the same time. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.32.1 15.05.2006...
Page 61
Disconnect plug J/ST3 of PCB RR-122 display, printer, touch. Remove microswitch by a screw driver and clean frame from remaining glue. Installation For installation proceed in reverse order. Use for example glue „Delomet Typ 2301“. RIELE BERLIN Photometer 5010 V5+ 4.33.1 15.05.2006...
Page 62
Replacement Remove the plastic foil carefully. Avoid scratches on the top case. Note: When the plastic foil is removed from device it might not be reusable. Installation For installation proceed in reverse order. RIELE BERLIN Photometer 5010 V5+ 4.34.1 12.05.2006...
Page 63
Remove top cover for printer, paper and ribbon. Disconnect 18-pin plug CN6 from printer controller 8NAD (REF 5010-053). Loosen 2 front screws in the hollows and remove the other 2 screws. Remove printer mechanism. RIELE BERLIN Photometer 5010 V5+ 4.35.1 15.05.2006...
Page 64
4.35 PRINTER MECHANISM 8NAD REF 5010-052 Installation For installation proceed in reverse order. Marked lead Solder printer connection cable to printer mechanism. Also take care for marked lead of pin 1. RIELE BERLIN Photometer 5010 V5+ 4.35.2 15.05.2006...
Page 65
5.2 Separating top case and bottom case • see chapter 5.3 Base plate • see chapter 5.7 Motherboard Installation For installation proceed in reverse order. Perform the installation instruction of this chapter. RIELE BERLIN Photometer 5010 V5+ 4.40.1 17.03.2009...
Page 66
4.40 PCB RR-120 MOTHERBOARD REF 5010-120 Component diagram (layout c) RIELE BERLIN Photometer 5010 V5+ 4.40.2 17.03.2009...
Page 68
12. Connect the positive lead of DVM to the battery BT1. (tolerance ± 10%). 13. Check the voltage + 3 V + 3 V ± 10% + 5 V ± 5% - 12 V ± 5% + 12 V ± 5% Ground connector RIELE BERLIN Photometer 5010 V5+ 4.40.4 17.03.2009...
Page 69
VCC (+ 5 V T 3.15 A / 250 V + 15 V / + 12 V T 6.3 A / 250 V Power supply printer + 5 V T 6.3 A / 250 V RIELE BERLIN Photometer 5010 V5+ 4.40.5 17.03.2009...
Page 70
• positive and <> 0, • set with potentiometer P4. The following diagram shows the ADC-counter values versus the photocurrent. ~ 600.000 offset photo current RIELE BERLIN Photometer 5010 V5+ 4.40.6 17.03.2009...
Page 71
1 MOhm 10. Switch the device on. 11. Run UTILITIES – ADC COUNTS (OPTIC). Turn the filter wheel to the dark position by selecting F0. 12. Note the ADC value e.g. 18015. RIELE BERLIN Photometer 5010 V5+ 4.40.7 17.03.2009...
Page 72
17. Remove the plug 1 MOhm and connect the photodiode plug with the socket ST7. 18. Do a dark level adjustment by applying utility U1 with both cuvette adaptors. (OM chapter 7.2.1 - Dark level adjustment). Ground connector RIELE BERLIN Photometer 5010 V5+ 4.40.8 17.03.2009...
The ADC converts the input voltage to a digital value between 0 and 32767 as shown on following diagram: 32767 ca. 17000 ca. 15200 ca. 13900 3,1 VDC 3,03 VDC 2,98 VDC input voltage of temperature sensor RIELE BERLIN Photometer 5010 V5+ 4.40.9 17.03.2009...
Page 74
3 ⇒ 1.3 V Involved components are the microcontroller unit IC3 (87C52PLCC), the OP-booster U1A (LM358), the transistor T4 (BC547), the transistor T5 (TIP132 mounted at the base plate) and the Peltier element. RIELE BERLIN Photometer 5010 V5+ 4.40.10...
Page 75
25 counts means a temperature change of 0.1 °C. • Service program S11 - Correction value is used for editing the value (chapter 3.3) • Utility programs U10 - Temperature adjustment adjusts the system automatically. (chapter OM 7.2.10) RIELE BERLIN Photometer 5010 V5+ 4.40.11 17.03.2009...
Page 76
The utility is working specifically to the current adaptor. An adjustment of the standard cuvette adaptor should be done later on. 14. Perform a media test by applying utility U4.4 (OM chapter 7.2.4.2 – Media test). RIELE BERLIN Photometer 5010 V5+ 4.40.12 17.03.2009...
Page 77
(see item 2 and chapter 3.3 – S17 BATTERY: Calibration 12 VADC) VOLUME OPTIMIZED OFF VDC LAMP 12.09V ADC CORRECTION STACK TEST BOOST nm EDP ON (STX-ETX-BCC) LANGUAGE 1: ENGLISH 2: GERMAN KEY SIGNAL ON PROGR. METHODS RIELE BERLIN Photometer 5010 V5+ 4.40.13 17.03.2009...
Page 78
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.14 04.08.2008...
Page 79
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.15 04.08.2008...
Page 80
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.16 04.08.2008...
Page 81
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.17 04.08.2008...
Page 82
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.18 04.08.2008...
Page 83
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120 33 R RIELE BERLIN Photometer 5010 V5+ 4.40.19 04.08.2008...
Page 84
Circuit diagram page 4.40.16 Potentiometer Circuit diagram page 4.40.19 D1, D2 Diode 1N4148 Resistor Resistor Resistor Header Transistor BC547 Attention! The following parts of the circuit diagrams are not specified. Circuit diagram page 4.40.16 Resistor RIELE BERLIN Photometer 5010 V5+ 4.40.20 04.08.2008...
Page 85
4.40 PCB RR-120 MOTHERBOARD Layout d REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.21 04.08.2008...
Page 86
4.40 PCB RR-120 MOTHERBOARD Layout d REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.22 04.08.2008...
Page 87
4.40 PCB RR-120 MOTHERBOARD Layout d REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.23 04.08.2008...
Page 88
4.40 PCB RR-120 MOTHERBOARD Layout d REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.24 04.08.2008...
Page 89
4.40 PCB RR-120 MOTHERBOARD Layout d REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.25 04.08.2008...
Page 90
4.40 PCB RR-120 MOTHERBOARD Layout d REF 5010-120 RIELE BERLIN Photometer 5010 V5+ 4.40.26 04.08.2008...
Page 91
500K. If the board which has to be replaced is equipped with 1M resistors: One of both R30 and R40 has to be removed by side cutting pliers. two 1Mohm resistors RIELE BERLIN Photometer 5010 V5+ 4.40.27 04.08.2008...
Page 92
5.1 Opening • see chapter 5.2 Separating top case and bottom case Installation For installation proceed in reverse order. Front view of the touchscreen J/ST9 J/ST8 J/ST1 Back view of the touchscreen RIELE BERLIN Photometer 5010 V5+ 4.41.1 11.03.2009...
Page 93
Connector for flexible cable with safety latch Flexible cable of J/ST8 Attention: In case of separating the PCB RR-122 from the touchscreen the flexible cable has to be removed after opening the safety latch. RIELE BERLIN Photometer 5010 V5+ 4.41.2 11.03.2009...
Page 94
Connector for flexible cable with safety latch OPEN Flexible cable of J/ST8 Attention: The safety latch has to be opened in case of unplugging the flexible cable. Touchscreen with PCB RR-122 inside of the top case RIELE BERLIN Photometer 5010 V5+ 4.41.3 11.03.2009...
Page 95
4.41 TOUCHSCREEN REF 5010-057 Mounting parts of the touchscreen: With the spacer the position of the touchscreen can be adjusted. The nut has to be unscrewed. Spacer Spring RIELE BERLIN Photometer 5010 V5+ 4.41.4 11.03.2009...
Need help?
Do you have a question about the Photometer 5010 V5+ and is the answer not in the manual?
Questions and answers