Troubleshooting
* Configuration Failure
Event Number: 301
Event Classification: Severe/Fatal Event
Action: Download a new configuration to the Monitor Module. If the problem
still exists replace the Monitor Module immediately.
Monitor Module will stop alarming.
Configuration Failure
Event Number: 301
Event Classification: Potential Problem
Action: Download a new configuration to the Monitor Module. If the problem
still exists replace the Monitor Module as soon as possible.
* Module Entered Cfg Mode
Event Number: 302
Event Classification: Typical Logged Event
Action: No action required.
Monitor Module will stop alarming.
Software Switches Reset
Event Number: 305
Event Classification: Potential Problem
Action: Download the software switches to the Monitor Module. If the
software switches are not correct, replace the Monitor Module as soon
as possible.
Internal Cal Reset
Event Number: 307
Event Classification: Severe/Fatal Event
Event Specific:
Action: Replace Monitor Module immediately.
Monitor TMR PPL Failed
Event Number: 310
Event Classification: Potential Problem
Action: Replace the Monitor Module.
Monitor TMR PPL Passed (Monitor TMR Proportional value Passed)
Event Number: 311
Event Classification: Potential Problem
Action: Replace the Monitor Module.
Module Reboot
Event Number: 320
Event Classification: Typical Logged Event
Action: No action required.
200
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
(Module Entered Configuration Mode)
(Internal Calibration Reset)
Ch pair x
(Monitor TMR Proportional value Failed)
3500/42 Operation and Maintenance
Need help?
Do you have a question about the PROXIMITOR 3500/42 and is the answer not in the manual?