5.3.2
Sampling method
The sample is taken by a vertical move of the bottle inside the fluid.
The bottle is linked with a graduated tape. A reading window allows monitoring the bottle location.
For complete explanation of sampling procedures, please refer to §6 "Operation".
5.3.3
Liquid transfer
After sampling, the liquid can be transferred into a laboratory bottle through a transfer valve.
The transfer of the liquid from the sampling bottle to a laboratory bottle occurs by gravity.
The opening of the bottle valve is realized by lowering the sampling bottle until its sitting on the ball
of the valve.
A pump can be used to accelerate and complete the transfer of the fluid.
50470/SGTX/0909
11
Industrial HERMetic Sampler GTX Chem