Download Print this page

Triumph IP12 User Manual page 5

Infusion pump
Hide thumbs Also See for IP12:

Advertisement

2. Product Specification
Safety Classification
Electric
protection
Class I
Type
Electric
protection
Defibrillation proof type CF applied Part
Level
IP34 (Prevent solid objects larger than 2.5mm in diameter and water intrusion
Ingress Protection
from splashing in all directions)
Working mode
Continuous
Classification
Portable equipment, non-portable infusion pump
Specification Parameters
Infusion
line
10-60 drops
specification
±5%
System Accuracy
Drip Infusion Rate
±5% or ±1drops/min
Accuracy
Infusion Rate
0.10-1500ml/h
Drip mode range
1~500drops/min
Bolus Rate
0.1-1500ml/h
Bolus preset
0.1-50ml
KVO Rate
0.1-5.00ml/h
Micro mode
100-1500ml/h
setting range
Minimum flow rate
0.01ml/h
increment
VTBI
0-9999.99ml, minimum step is 0.01ml
Total Volume
0-9999.99ml, minimum step is 0.01ml
Infused
Time Range
1min-99hrs59min
Fuse Type
slow fuse 2A 250V
Dimensions
199(W)*111(D)*126.5(H) mm
Weight
1.4kg
Power Supply
AC power supply
100-240V 50/60Hz
Input power
50VA
5

Hide quick links:

Advertisement

loading