III. Protocol: Setup Instructions - GO
4.4.3 STEP [1]: Set flow to 10-25 L/min
Note: For better flow control, a flowmeter capable of 40 L/min is
Note: If using an orifice-type flow regulator that is common to most
Note: The GO
Note: The duration of an "E" cylinder when using the GO
 
VORTRAN
®
GO
2
VENT™ User's Guide
Remove the GO
VENT™ from package and
2
connect the supply tubing to either an
appropriate cylinder or wall source. A good
starting point is 10 L/min. Adjust as
needed. See ENTRAINED FLOWCHART
for "Total delivered flow" requirements. The
GO
VENT™ is designed to automatically
2
deliver 40 L/min when connected directly to
a 50 PSIG gas source.
preferred. The flow controls the inspiratory time – the higher
the flow, the shorter the i-time; the lower the flow, the longer
the i-time.
cylinders, you will only be able to provide as much flow as the
regulator indicates. If the regulator being used has a high flow
port connection and you connect the GO
you will automatically get 40 L/min.
VENT™ is completely gas driven, requiring no
2
electrical power and will deliver 100% oxygen to a patient.
depend on the flow. An "E" cylinder contains 625 L of gas. At
40 L/min, 625 L will last up to 15 minutes; at 20 L/min, 625 L
will last up to 30 minutes. 15 L/min orifice type flowmeters
used on many "E" cylinders will not be able to deliver more
than 15 L/min. When clinicians decide that 15 L/min is not
sufficient flow, the GO
that has a high flow port (50PSIG) to deliver 40 L/min. The
length of use for various sizes of compressed oxygen tank (D,
E, M & H) is a function of supplied oxygen flow from 6 to 40
L/min to GO
VENT™ (see Table 1 below).
2
TABLE 1 - LENGTH OF USE FOR COMPRESSED OXYGEN TANKS
Tank
(Liters)
Flow (L/Min)
6
8
10
12
15
20
25
30
35
40
Page 9 
VENT™ (continued)
2
VENT™ can be attached to a regulator
2
D
E
387
662
Length of use (minutes)
65
100
50
80
40
60
30
50
25
40
20
30
15
25
13
20
11
18
10
16
5/26/2017
VENT™ to this port,
2
VENT™ will
2
M
H
3028
6905
500
1150
380
860
300
690
250
570
200
460
150
340
120
270
100
230
80
190
70
170