II. Clinical Considerations
The GO
VENT™ provides short term, pressure cycled, and constant flow
2
ventilatory support for either breathing or non-breathing patients. This allows
the patient to receive consistent and reliable ventilatory support. Because the
GO
VENT™ is pressure cycled, changes in the patient's lung compliance will
2
cause a change in the patient's breathing rate. The GO
sensitive. Final adjustments should be made with the GO
secured operating position. The GO
and exhalation (PIP and PEEP) which minimizes the possibility of gas trapping.
During inhalation, exhalation will not start until PIP is reached. During
exhalation, inhalation will not begin until pressure drops to PEEP. For the
spontaneously breathing patient, the rate dial of the GO
baseline pressure is above the intrinsic PEEP allowing the patient to initiate
inhalation by drawing the baseline pressure down to the set PEEP. Because the
GO
VENT™ is a constant flow pressure cycled device, changes in patient
2
compliance will result in changes in the respiratory rate (stiffer or smaller
compliances produce faster rates). The advantage of this minimizes the danger
of barotrauma. It should be emphasized that the GO
by trained personnel who continuously monitor the patient. The GO
not an ICU stand-alone ventilator with multiple monitoring features.
Setup and use of the GO
III on page 7). Set desired flow (Q), adjust PIP pressure dial to obtain desired
inspiratory time (t
Table 1). The gas flow, patient's lung compliance, and PIP settings control the
inspiratory time and tidal volume. Then adjust rate dial to obtain desired
breathing rate.
Table 1 – Estimated Tidal Volume (mL) Delivered at
Various Flow (L/min) and Inspiratory Time (Seconds)
Flow
(L/min)
15
20
25
30
35
40
VORTRAN
®
GO
2
VENT™ User's Guide
VENT™ is simple (refer to Setup Instructions in Section
2
) to attain tidal volume (TV = Q x t
insp
Inspiratory Time (Seconds)
0.5
1
125
250
167
333
208
417
250
500
292
583
333
667
Page 3
VENT™ is pressure cycled on inhalation
2
VENT™ is to be used only
2
1.5
2
2.5
375
500
625
500
667
833
625
833
1042
750
1000
1250
875
1167
1458
1000
1333
1667
5/26/2017
VENT™ is positional
2
VENT™ in its
2
VENT™ is set so the
2
VENT™ is
2
see Tidal Volume
insp
3
750
1000
1250
1500
1750
2000