5. Information displayed on the counting weighing apparatus
5.1. The weight field displays the following information:
F----1
Data exceeds the range of display.
F----2
Error in A/D conversion circuit
F----3
Indicating that keys are available
F----5
Indicating fault in the sensor
F----L
Zero of sensor signal is too low
F----H
Weight exceeds the range
C----F
Error in calibration. It is not at zero or the data are not stable during the calibration.
C----H
Error in calibration. Article is placed on the weighing platform or the zero of the sensor
is too high during the calibration.
E-----
Error in storing the calibration data. Service is possible only after re-calibration.
5.2. The unit weight field displays the following information:
-CAL-
You are in the calibration state
≡ X≡
The displayed value is the number of operations for accumulation. The weight field
displays the total weight and the number-of-pieces field the total weight.
H
Meaning the preset quantity is exceeded and an audible warning is given.
5.3. The quantity field displays the following information:
F---1
The value of the number-of-pieces exceeds the range of display
5.4. Indicators
kg(g)
lb
SAMP↑
UNIT↑
Under-voltage
Charging
Fully charged
Lights when the data are stable. The current weight unit is in kg(g)
Lights when the data are stable. The current weight unit is in lb.
Lights when the total sampling amount is less than the min. sampling value.
Now, add the number of samples till the indicator is out. Set by re-sampling to
ensure the accuracy.
Lights when the average or set unit weight is less than the min. unit weight
value. Now, the counting weighing apparatus remains serviceable but error may
happen during the counting. It is recommended that counting weighing
apparatus with high distinguishability is used.
Lights when the voltage of the battery is insufficient along with audible alarm,
charge the battery. Otherwise, the battery operates only 1-2 hours. The counting
weighing apparatus operates normally during the charge.
Lights when the counting weighing apparatus is in the charging state. It takes
about 8 hours for one charge.
Slowly lights when the counting weighing apparatus is under charge. When the
charging lamp dies out, it indicates that it is fully charged. Now, it allows further
charging.