balance will automatic calibrate (XXX should be the weight of standard weights. For
instance, "C---20" indicates to put 20 kg test weights). At the same time, only place
calibrating weights on the scale platform. After stabilizing, the scale displays the weight of
the calibrating weights and the symbol "kg"; till now the calibration is completed and the
scale can be normally used. If press "CAL", but display "C----F", which indicates the Zero
point is not stable. Then re-press "TARE" to make "zero" point stable; and re-press the
"CAL" to carry out calibration.
"
"compound key:The main function for conversion ,counting and print.Default
4.
setting is the conversion.
4.1 convert units: Press
will lighten.
4.2 Count
i.
Choose samples. To precisely count the articles, the quantity of samples for
counting should be chosen in according with the weights of the articles. The
quantity of samples can be chosen from "1-10-20-50-100". For a small number of
articles, the larger quantity of samples, the more precise.
ii.
Place the samples on the pan, the balance will show the total weights of samples.
Then press "Count", balance will show "1", and unit will show "pcs", which
indicates that balance is now counting. And all the samples are one unit. Then
press "kg/lb", the displaying will show between 1-10-20-50-100. The quantity of
articles must be the same with the quantity of samples, and place the similar
articles. The total amount of objects is showed. Press the "Count" again back to
the normal situation.
Note:For counting keys and convert the unit key, print for the same key buttons "
in the use of counting functions must first to balance setting, please refer to the setting
method of section 6 5 points.
4.3 Print : put the balance connect printer, Press"
print.

5. Displaying Information

5.1 Display window
F----1 data beyond scope
F----2 AC/DC conversions
F----3 buttons
F----5 sensor faults
F----L lower zero position of sensor signal
F----H overload
C----F calibration error, instability or not at zero point during calibrating
, the units will be changed between "g" and "lb". The indicator
"
"
" ,the balance set A data,the printer
" ,so