Maintaining the Defibrillator

Cleaning and Sterilizing the Internal Paddles

o
Soap and water
o
Chlorine bleach (30 ml/l water)
o
Ammonia-based cleaners
o
Keep the outside of the instrument clean and free of dust and dirt. Clean the
paddles thoroughly to prevent build-up of dried electrolyte paste.
o
Do Not allow any fluids to penetrate the instrument case. Avoid pouring fluid
on the unit while cleaning.
o
Do Not use abrasive cleaners, or strong solvents such as acetone, or acetone-
based compounds.
o
Clean the display screen carefully. It is especially sensitive to rough handling
and subject to scratching.
o
Do not steam sterilize the monitoring leads, submerge them for prolonged
periods, or heat them above 50ºC. If metallic surfaces become oxidized, clean
them with a very light abrasive (toothpaste). Do not use highly abrasive
cleaners such as steel wool or silver polish.
o
Do not steam or gas sterilize the external paddle set.
Cleaning and Sterilizing the Internal Paddles
8
o
Clean the electrode surface and handle with standard hospital solution.
o
Do not use acetone and ammonia-based cleaners.
o
Use a small, soft brush with cleaning solution to clean any contamination from
the electrode surface and edges.
o
Before sterilizing, clean any excessive residue which accumulates on the
handles or electrode surfaces.