Replacing The Battery; Cleaning Exterior Surfaces - HP CodeMaster XL+ M1722A User Manual

Defibrillator/monitor
Table of Contents

Advertisement

Maintaining the Defibrillator

Cleaning Exterior Surfaces

8
9
The battery can be permanently damaged if left uncharged for a prolonged period.
CAUTION

Replacing the Battery

To install a new battery, perform the following steps:
1
2
3
4
5
6
7
8
Cleaning Exterior Surfaces
8
The CodeMaster XL+ and its accessories are chemically resistant to common
hospital cleaning solutions and non-caustic detergents. The following list includes
some approved cleaning solutions.
o
Fully recharge the battery by plugging the instrument into AC power for 18
hours. Verify that the
AC
The instrument is now ready to be returned to service.
Remove AC power from the instrument.
Place the instrument upside down on a workbench.
Turn the two battery compartment quarter-turn screws counter-clockwise.
Lift out the battery door.
Disconnect the battery connector and remove the battery.
Connect the new battery and slide it into the compartment.
Replace the battery door and secure it by turning the two retaining screws a
quarter-turn clockwise.
Perform the Battery Capacity Check above before placing instrument in service.
90% Isopropyl alcohol (except adaptors and patient cable)
and
BATT
lights are on.
8-5