Command: ZERO
Purpose:
Response:
Command String:
Description:
Response String:
Description:
Example:
Manual 19854 Rev. D
ECN 50523
Zeros the output of a 443B102 when in long time constant charge mode.
The module responds with a "0" to tell the calling process the command has
been received.
XYC02ZERO
X—Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C02— Module type for the target. Only type "C02" (443B102) can accept this
command since the 443B101 does not have long time constant.
ZERO—Commands the module to reset the charge stage.
"0"
This return indicates that the command was received by the card.
Command String:
"04C02ZERO"
Response String:
"0"
The 443B102 in rack 0 and slot 4 is commanded to zero its output.
443B101/443B102
22