Calibration Using Calibration Panel CP 40
Sensor Calibration
Calibration object at measuring distance to sensor
9.3.4
Calibration Process
After that the software will guide you through the individual calibration
steps. In the first step, you are asked to place the sensor and the cali-
bration object in the center of the measuring volume.
Enable
the
Automatic exposure
time.
In exceptional cases it might happen that the function
Automatic ex-
posure time
cannot work correctly.
In this case, disable the function and use the exposure time wheel to
adjust the optimum exposure time for the camera image.
Exposure time adjustment for calibration
Page 35 (44)