7. Sensor Setup

7.1
General
In order to adjust the sensor, the complete system including the
ATOS software needs to be installed and the sensor must be
equipped with a set of lenses matching the measuring volume and
the camera position.
For each measuring volume (MV), the set of lenses are factory-
preadjusted.
If you have the respective set of lenses for each measuring volume,
setting up the sensor is easy as you do not need to carry out all the
steps referring to adjusting the focus and aperture!
You always need to perform a complete sensor setup if, for example,
the corresponding set of lenses has never been adjusted before or is
decalibrated.
7.2

Measuring Volume

Use the table in section 3.1 to determine a measuring volume.
You create larger measuring volumes with camera positions 400 and
800 and smaller measuring volumes with camera position SO (see ta-
ble 3.2).
Please check whether the cameras are in the correct position. How to
change the camera positions is described in section 7.3.
7.3

Changing the Camera Positions (Step by Step)

In order to change the camera distances, you need to open the sensor
housing. Therefore, disconnect all cables from the sensor.
It is not possible to operate the sensor without the upper part of the
housing!
In the following, we describe how to change the camera distance from
400 to SO.
When you are changing the distance from SO to 400 or when you
are changing to camera position 800, please proceed corresponding-
ly.
Sensor Setup