1.3. SPECIFICATIONS

The following specifications are based on a one-year calibration cycle, an operating temperature of 15 to
28ºC (59 to 82ºF) and a relative humidity not exceeding 80%. Monitors must be stored in an environment
o
o
between 10
C to 60
C (50 to 140
Interface
Display
Measurement Controls
Data Acquisition and Transfer
Screen Personalization
Saved Settings
Effective Aperture Diameter
Spectral Range
Calibrated Spectral Range
Power Noise Level
Response Time
(0 - 100 %)
Typical Sensitivity
Calibration Uncertainty
Linearity with Power
Repeatability (Precision)
Maximum Power Resolution
Minimum Power
Maximum Power
Pulsed Laser Damage Thresholds
1.064 µm, 360 µs, 5 Hz
1.064 µm, 7 ns, 10 Hz
532 nm, 7 ns, 10 Hz
266 nm, 7 ns, 10 Hz
Minimum Frequency
Maximum Number of Readings
before probe must be cooled
[for 25°C starting temp.]
Maximum Exposure Time before
probe must be cooled
[for 25°C starting temp.]
Maximum Device Temperature
Operating Conditions
Revision 2.2
o
F) and a relative humidity not exceeding 90%.
Color LCD 28 X 35 mm (128 x 160 pixels)
Wavelength Selection and User Calibration
Maximum of 50 000 Measurements
4 screen orientations and 4 brightness levels
(A maximum exposition of 6 sec is allowed at 250 W)
Max. Energy Density
9 J/cm
1.0 J/cm
0.6 J /cm
0.3J /cm
Maximum Relative humidity : 80%
Pronto-250
Touch Screen
On/Off Controls
Screen Orientation
Screen Brightness
Wavelength
User calibration
19 mm Ø
190 nm – 20 µm
0.248 – 2.5 µm and 10.6 µm
10 mW
5 s
0.23 mV/W
± 3 %
± 2 %
± 0.5 %
1 mW
0.5 W
250 W
Peak Power Density
2
2
2
2
15Hz
5 W: 50
25 W: 10
50 W: 5
125 W: 2
250 W: 1
5 W: 120 sec
25 W: 60 sec
50 W: 30 sec
125 W: 12 sec
250 W: 6 sec
65°C
Ambient temperature : 15-28°C
2
25 kW/cm
2
143 MW/cm
2
86 MW/cm
2
43 MW/cm
2