Specifications - Gentec S-LINK User Manual

Pc-baced power & energy monitor
Table of Contents

Advertisement

S-LINK User's Manual
Unpacking
Each Gentec-EO S-Link is thoroughly tested and calibrated prior to shipment.
Visually inspect every S-Link unit after removing it from the shipping containers. If you see any damage,
retain all packaging materials and shipping receipts. Any damage claim should be made promptly to the
transportation company. Notify the nearest Gentec-EO representative concerning the claim, so that any
repair or replacement can be arranged as soon as possible.
You can download the latest software version for the PC S-Link (and install it on your PC) and for the S-
Link (and install it on the S-Link) with the serial interface anytime from our website www.gentec-eo.com.
You can download the PC S-Link program from our website. Go to the Downloads section of our
website: www.gentec-eo.com. Click on the file name and download it to your PC. The specific actions
necessary vary between browser and browser settings. After it is transferred, open the file on your PC
and follow the instructions to uncompress and install it. You may also consult the download instructions
document that is also in the download section of our website.

1.2 Specifications

The following specifications are based on a one-year calibration cycle, an operating temperature of 18 to
28ºC (64 to 82ºF) and a relative humidity not to exceed 80%.
Power Range
Power Scales
(thermal head)
Resolution (digital)
Wattmeter mode Accuracy
Response Time
2
(accelerated)
Sampling Frequency
Statistics
1
The 2.5µV bias can introduce an error into low power measurements with low sensitivity detectors. It is
essential to use the Zero Offset to rezero the S-Link before making a measurement in these conditions. It
is always good practice to use the Zero Offset. See section 2.2.
2
Varies with detector head.
Revision 2.8
Table 1-1 List of Specifications
Power meter specifications S-Link
15 scales:
300uW, 1 mW, 3mW, 30mW, 100mW,
300mW, 1W, 3W, 10W, 30W, 100W, 300W, 1kW,
3kW, 10kW
Normal mode: Current scale/2048
Ascii mode: Current scale/32768
±0.75 % 2.5 µV from 10% to full scale
Current value, Max, Min, Average, Std Dev., RMS
stability, PTP stability, Time
1 uW to 10 kW
1
1 sec
100 Hz
2

Advertisement

Table of Contents
loading

Table of Contents