3.2.5. Adjusting the CALIBRATION

It is possible to adjust the calibration of the device. You can do so by applying a Correction Factor to the
measurements. To access the Calibration menu, simply press the CAL button
displayed is the multiplication factor that will be applied to the measurements. A factor of 1.000 keeps the
calibration unchanged. You can enter any number between 0.500 and 2.500.
The 4 digits of the correction factor are changed individually. For example, if the Correction Factor is
1.000 and you want to change it to 1.200, just press the first 0 (make sure it turns blue, like this
and then increment it twice with the Up button.
Initial Setting
The last measured
value (145.3 W) is
displayed between
the Up/Down buttons
Once the value is entered, press the On/Off/Settings button to exit the menu and automatically save your
new settings.

3.2.6. Getting Information ABOUT the Device

Relevant information about the device are stored in the About menu
useful when returning the device for repairs or recalibration. These include, the Model Name, The Serial
Number, the Firmware version and the Last Calibration Date.
Press the On/Off/Settings button to exit the menu and go back to the measurement screen.