Vitea Care PARAPODIUM PD150 User Manual

Dynamic parapodium

Advertisement

Quick Links

Dynamic Parapodium
Model PD150, PD 180, PD 200
User's Manual
09.2016

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PARAPODIUM PD150 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Vitea Care PARAPODIUM PD150

  • Page 1 Dynamic Parapodium Model PD150, PD 180, PD 200 User’s Manual 09.2016...
  • Page 2: Table Of Contents

    CONTENTS MEDICAL INFORMATION ..........................................3 NTRODUCTION ORDS INESITHERAPY ISTURBANCES UNCTION NTERNAL RGANS ISABLED UBJECTS HRONIC ITTING OSITION ............................3 IDDING ......................... 4 DVANTAGES YNAMIC ARAPODIUM ............4 OMMON YSTEMIC ISTURBANCES IKELY CCUR ARAPLEGIA XEMPLARY RELIMINARY EHABILITATION ROGRAMME ERIOD DAPTATION RGANISM .
  • Page 3: Medical Information

    MEDICAL INFORMATION “It is easier to achieve the intended aim after getting acquainted with the subject-matter of the undertaking – it enables to predict possible consequences adequately and unifies the language of communication between the partners of the dialogue reducing unnecessary, unpleasant experiences”. Introduction, Key Words, Kinesitherapy Paraplegia and paraparesis can be of traumatic or pathologic origin and it most frequently occurs in the following diseases:...
  • Page 4: Advantages Of Dynamic Parapodium

    • increased level of stress associated with everyday activities, • disturbances of sensory perception, • social interaction and self-acceptation more difficult (depression), • increased stigmatisation, stereotypy, discrimination and lack of acceptation among the active members of the society, • reduced effectiveness and possibility of rehabilitation, which leads to secondary aggravation of the above disorders.
  • Page 5: Exemplary Preliminary Rehabilitation Programme For The Period Of Adaptation Of The Organism To The Device

    Efficiency of the locomotor system. Frequent dysfunction of the locomotor system accompanying the underlying disease prompt to analyse thoroughly the possible need of passive or dynamic correction, compensation, alleviation or stabilisation of the particular elements of the system by means of appropriate orthoses (collars, etc.), e.g. the cranio- cervical or thoracic segment of the vertebral column in the cases of subluxation symptoms or instability by means of corsets or belts applied for the cases of scoliosis or muscular insufficiency, orthoses for the upper extremities in the cases of contractures or special shoes or insoles –...
  • Page 6: User's Manual

    fatigue) the exercise can be prolonged by one minute. In case of any adverse effects associated with upright position, the exercise should be discontinued immediately. When the symptoms subside, the training should be resumed, starting the stage from the beginning. Leaving the parapodium and changing position to sitting on a chair, bed or wheelchair (the presence of an accompanying person mandatory during this exercise).
  • Page 7: General Information

    ATTENTION. This symbol is used to signal that particular attention to the content of the paragraph marked in this way is needed. WARNING. This symbol is used for the description of functions (activities), which performed incorrectly, i.e. according to the Manual, may lead to the damage of the device. DANGER.
  • Page 8 The parapodium consists of the following elements: Diagram 1 Principles of Function of Dynamic Parapodium Model PD150, PD180, PD200. Many years of experiments and observation of disabled people with paraplegia have proven, how difficult it is to design a simple, easy to operate device for a large group of users characterised by a large diversity of physical condition of the organism and complications involving the internal organs.
  • Page 9: Preparation Of The Parapodium To Use

    Alterations of the centre of gravity in the parallel to the intended direction of movement are obtained by slight bending of the patient’s trunk forth and back, possible owing to the special design of the lower joint and the corset hanger joint.
  • Page 10: Use Of The Parapodium

    Table 1. PD 150 PD 180 PD 200 No. Parameter (cm) MIN. MAX. MIN. MAX. MIN. MAX. Hip width (S) Hip depth (G) Height to the knees (D) Height to the trochanter (K) Weight Contracture of the knee (°) 30° 30°...
  • Page 11 ATTENTION Dynamic Parapodium is to be used only on a flat, horizontal and hardened surface. ATTENTION Any attempt to overturn the parapodium on purpose more than the maximum lateral tilt (10’) is contrary to the instructions for use of the device and may cause a threat to the patient’s life. While using the parapodium the patient must not be loaded more than 5 kg over his normal weight.
  • Page 12: Transport Of The Device

    Standing in the parapodium. The parapodium makes it possible for a disabled person to stand self-dependently and comfortably without the aid of the hands for the period of many hours. Even if the patient faints, upright position is maintained owing to continuous control of the child’s gravity centre. Walking in the parapodium.
  • Page 13: Storing, Cleaning And Maintenance Of The Parapodium

    In order to reassembly the parapodium follow these instructions: Put the corset assembly back (pos.1, pict.4) on the basis assembly (pos.2, pict.4), Adjust the previous height of the mounting of the corset assembly (1), Screw the screws with two pads again (pos.3, pict.4), Check the condition of equal dimensions X and Y (X=Y).
  • Page 14: Environmental Protection

    • The users of the device must not be under the influence of alcohol or drugs. • The predicted time of using the device is 4 years. After this period it can be used only after a positive control performed by the manufacturer. •...
  • Page 15: Warranty

    2.10 Warranty The Warranty Chart is a separate document attached to every new product. 2.11 Warranty instructions Failure to comply with the procedures presented below will result in the loss of warranty. • In case of any damage, the use of the device should be discontinued until the moment of repair. •...
  • Page 16 MANUFACTURER mdh Sp. z o.o 90-349 Łódź 22/24 Tymienieckiego Street tel.: (+48 42) 674 83 96 tel.: (+48 42) 676 32 54 fax.: (+48 42) 299 69 24 www.viteacare.com „Dynamic Parapodium Model PD150, PD180, PD200. User’s Manual“ is a property of mdh Sp. z o.o. All rights reserved.

This manual is also suitable for:

Parapodium pd 180Parapodium pd 200

Table of Contents