Carrier 69NT40-541-304 Operation, Service And Parts Supplement page 50

Dual temperature container refrigeration units
Table of Contents

Advertisement

A. Connection to Low Side of System
B. Connection to either: Refrigerant Cylinder or Oil Container
C. Connection to High Side of System
When the suction pressure hand valve is frontseated (turned all the way in), the suction (low) pressure can be
checked. When the discharge pressure hand valve is frontseated, the discharge (high) pressure can be checked.
When both valves are open (turned counter-clockwise all the way out), high pressure vapor will flow into the low
side. When the suction pressure valve is open and the discharge pressure valve shut, the system can be charged.
Oil can also be added to the system.
A R-134a manifold gauge/hose set with self-sealing hoses (see
covered within this manual. The manifold gauge/hose set is available from Carrier Transicold. (Carrier Transicold
P/N 07-00294-00, which includes items 1 through 6,
hose set, do the following:
a. Preparing Manifold Gauge/Hose Set For Use
If the manifold gauge/hose set is new or was exposed to the atmosphere, it will need to be evacuated to remove
contaminants and air as follows:
1. Back seat (turn counterclockwise) both field service couplings (see
valves.
2. Connect the yellow hose to a vacuum pump and refrigerant 134a cylinder.
3. Evacuate to 10 inches of vacuum and then charge with R-134a to a slightly positive pressure of 0.1 kg/cm
(1.0 psig).
4. Front seat both manifold gauge set valves and disconnect from cylinder. The gauge set is now ready for
use.
T-319
Figure 6.2 Manifold Gauge Set
SUCTION
PRESSURE
GAUGE
B
A
OPENED (Backseated )
HAND VALVE
Figure
6–2
DISCHARGE
PRESSURE
GAUGE
C
CLOSED (Frontseated)
HAND VALVE
Figure
6.3) is required for service of the model
6.3). To perform service using the manifold gauge/
Figure
6.3) and midseat both hand
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

69nt40-541-345

Table of Contents