3 AE Specifications
3.1
DICOM Recording Specification
The DICOM Recording AE provides Standard Conformance to the DICOM Media Storage
Service, File Format Class and the Media Storage Application Profile.
Application Profile
Basic Cardiac X-ray
Angiographic Studies on
CD-R Media
1024 X-ray Angiographic
Studies on CD-R Media
3.1.1
File Meta Information for the Application Entity
The Application Entity Title is registered into the DICOM File Meta Information header.
•
Source Application Entity Title
3.1.2
Real-World Activity
3.1.2.1
Real-World Activity - Transfer of X-ray Examination
DICOM Recording AE acts as a FSC when chosen by the operator. This AE writes the SOP
instances onto the CD-R media, creates a DICOMDIR and writes it onto the CD-R media.
3.1.3
Media Storage Application Profile
DICOM Recording AE uses the Application Profile defined in Table 1.
3.1.4
Information Object Definition and DICOMDIR keys
Information Object Definition and DICOMDIR keys are described in Chapter 8.
3.2
DICOM Reader Specification
The DICOM Reader AE provides Conformance to the DICOM Media Storage Service,
File Format Class and the Media Storage Application Profile.
Application Profile
Basic Cardiac X-ray
Angiographic Studies on
CD-R Media
Extended Cardiac X-ray
Angiographic Studies on
CD-R Media
Table 1
Identifier
Real World Activity
Transfer of X-ray
STD-XABC-CD
Examination
Transfer of X-ray
STD-XA1K-CD
Examination
TM_XA_DCM_V12
Table 2
Identifier
Real World Activity
Review of X-ray
STD-XABC-CD
Examination
Review of X-ray
AUG-XABC-CD
Examination
- 6 -
No. MIIXR0002EAC
Role
SC Option
FSC
Interchange
FSC
Interchange
Role
SC Option
FSR
Interchange
FSR
Interchange