Error Messages - HEIDENHAIN ND 261 Series Manual

Working with the measured value display unit

Advertisement

Display Freeze by Measured Value Output Signal
The effect of the signal for measured value output on the display is defined in user
parameter P23.
Concurrent display: No display freeze. The unit shows the current measured
value (
Frozen display: The display is frozen and is updated with each signal for meas-
ured value output (
Frozen/concurrent display: The display freezes only as long as the signal is
present (

Error Messages

To clear error message
When you have removed the cause of the error,
press CL.
Message
If all decimal points light up, the measured value is too large or too small.
In this case, set a new datum or retract.
If all sorting indicators light up, this means that the upper sorting limit is less
than the lower limit.
1)
These errors are significant for a connected device. The error signal (pin 19) at the
EXT D-sub connection is active.
).
).
).
Cause and effect
Last measured value not yet latched
External device not ready for data transfer
(
Data interface:
Parity error or wrong format
Incorrect input value
Overflow caused by external preset
Overflow, trigger limit 1
Overflow, trigger limit 2
Overflow, lower sorting limit
Overflow, upper sorting limit
Encoder signal too weak
Input frequency too high for encoder input
(will occur for example when traverse speed too high)
Internal counter overflow
Error while crossing over reference marks
To clear the error message: Switch off the display unit.
Should any of these error codes recur, contact your HEIDENHAIN
service agency
Offset compensation values for encoder signals have been
erased: contact your HEIDENHAIN service agency.
Erase the operating parameters.
only appears once)
1)
1)
(encoder may be contaminated)
1)
1)
1)
1)