Measurement Computing USB-2416 User Manual page 30

Usb-based 24-bit, isolated, high-channel-count multifunction daq device
Table of Contents

Advertisement

USB-2416 User's Guide
Table 11. T type thermocouple accuracy specifications, including CJC measurement error.
A/D data rate
Sensor temperature
range
3750 S/s
-200 °C
0 °C
400 °C
2000 S/s
-200 °C
0 °C
400 °C
1000 S/s
-200 °C
0 °C
400 °C
500 S/s
-200 °C
0 °C
400 °C
100 S/s
-200 °C
0 °C
400 °C
60 S/s
-200 °C
0 °C
400 °C
50 S/s
-200 °C
0 °C
400 °C
25 S/s
-200 °C
0 °C
400 °C
10 S/s
-200 °C
0 °C
400 °C
5 S/s
-200 °C
0 °C
400 °C
2.5 S/s
-200 °C
0 °C
400 °C
To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after
the initial power on. The accuracies listed above are only guaranteed if the USB-2416 is housed in its plastic
enclosure.
All specifications are (±).
Accuracy error,
maximum
2.821 °C
1.050 °C
0.957 °C
2.821 °C
1.050 °C
0.957 °C
2.821 °C
1.050 °C
0.957 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
2.644 °C
0.979 °C
0.912 °C
30
Specifications
Accuracy error,
Tempco
typical
(°C/°C)
1.676 °C
0.027
0.558 °C
0.595 °C
1.676 °C
0.027
0.558 °C
0.595 °C
1.676 °C
0.027
0.558 °C
0.595 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C
1.505 °C
0.027
0.487 °C
0.551 °C

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the USB-2416 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents