Q.series
Gantner Instruments Test & Measurement GmbH
field length is calculated including the decimal point, but
without any commas displayed for the thousands.
After closure of the dialog the number of transferred places is
displayed in the Format/Adjustment column, e.g. ff,fff.f for
an output with a total of seven characters including the deci-
mal point and one post decimal place.
If negative numbers occur, the display in this example is lim-
ited to -9,999.9 (seven characters without the comma).
6. Click in the Range/Error column to define the output scaling
and the reaction in the case of an error (optional).
7. Click in the Additionals column to specify the signal source
(variable) used for the output.
8. Save the changes in a file and close the module once you have
made all the changes:
The file is created within the project folder and the file name
is generated automatically. The project folder bears the same
name as the project. The generated file name contains the
address of the Test Controller through which the module is
connected and an identifier for the relevant module.
IMPORTANT
If the modules are connected through a Test Controller, new
module settings must also be updated there. Select
> Write Project (Update) in the test.commander. The project
file is in this case automatically updated.
Configuration → Defining analog outputs
5
or File > Save To File And Exit.