Fluke RUSKA 2482 User Manual page 25

Differential pressure piston gauge
Table of Contents

Advertisement

To zero the Line sensor: Set-up → Calibrate → Line
1. Click
.
2. Release pressure from the line sensor or apply a known low pressure.
3. Enter the Actual pressure.
4. Click
. The C0 coefficient will be adjusted to give the correct pressure
reading.
5. Click
.
To calibrate the line sensor: Set-up → Calibrate → Line
1. Click
.
2. Apply the first known pressure.
3. Enter the Actual pressure.
4. Click
. The dialog for the second point will be displayed.
5. Apply the second known pressure.
6. Enter the Actual pressure.
7. Click
. C0 and C1 will be adjusted to calibrate the line pressure sensor.
8. Click
.
Figure 20. Calibrate Line Pressure
Differential Pressure Piston Gauge
Software Setup
gms21.bmp
17