Alarm Conditions - CAIRE FreeStyle Comfort User Manual

Oximedical respiratory
Hide thumbs Also See for FreeStyle Comfort:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Alarm Conditions

Priority #
Alarm
Description
1
System
Malfunction
Battery Over
Temp
2
System
Malfunction
High
Temperature
3
System
Malfunction
High
Pressure
4
No Breath
detected
5
Low Battery
Shut Down
6
Warm Up
7
Low
Concentration
Display
FreeStyle Comfort
Audible
More Details
Signal
1 beep
Unit will alarm when battery temperature is
every
approaching saftey shutdown threshold.
20 sec
If shutdown threshold is reached, unit will shut
down. If external power is connected, alarm
screen will display and audible alarm will
continue for 2 min.
1 beep
Unit will alarm when internal temperature is
every
approaching shutdown threshold.
20 sec
If shutdown threshold is reached, unit will shut
down and alarm screen will display and audible
alarm will continue for 2 min.
1 beep
If shutdown threshold is reached, unit will shut
every
down and alarm screen will display and audible
20 sec
alarm will continue for 2 min
1 beep
No breath detected for 15 seconds.
every
Activated after first breath or warm up screen
30 sec
is off, whichever comes sooner.
1 beep
Alarms until battery runs out of power and unit
every
shuts down
30 sec
none
Unit will start by auto-dosing until a breath is
detected or Warm up ends, whichever is sooner.
1 beep
every
30 sec,
for up to
5 min
PN MN232-C4 C | User Manual
ENG - 19