BIRD 4044 Series Operation Manual

Non-directional power sensor

Advertisement

Quick Links

Non-Directional Power Sensor
Model 4044 Series
Operation Manual
©Copyright 2019 by Bird Technologies, Inc
Instruction Book Part Number 920-4044 Rev. D
Thruline® and Termaline® are registered trademarks
of Bird Electronic Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4044 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BIRD 4044 Series

  • Page 1 Non-Directional Power Sensor Model 4044 Series Operation Manual ©Copyright 2019 by Bird Technologies, Inc Instruction Book Part Number 920-4044 Rev. D Thruline® and Termaline® are registered trademarks of Bird Electronic Corporation...
  • Page 2: Safety Precautions

    Safety Precautions The following are general safety precautions that are not necessarily related to any specific part or procedure, and do not necessarily appear elsewhere in this publication. These precautions must be thoroughly understood and apply to all phases of operation and maintenance. WARNING Keep Away From Live Circuits Operating Personnel must at all times observe general safety precautions.
  • Page 3: Safety Symbols

    Safety Precautions Safety Symbols WARNING Warning notes call attention to a procedure, which if not correctly performed, could result in personal injury. CAUTION Caution notes call attention to a procedure, which if not correctly performed, could result in damage to the instrument. The caution symbol appears on the equipment indicating there is important information in the instruction manual regarding that particular area...
  • Page 4: Safety Statements

    Directional Power Sensor Safety Statements USAGE ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION. EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
  • Page 5 Safety Precautions SERVICE SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE - TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED TO DO SO. SERVICIO LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO. PARA EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO REALICE NINGÚN SERVICIO A MENOS QUE ESTÉ...
  • Page 6: About This Manual

    The installation section provides an example connection diagram and steps required for a typical installation. Maintenance — This section provides contact information for the Bird Service Center and provides a list of the specifications for the 4044 Series Non- Directional Power Sensors.
  • Page 7: Table Of Contents

    Table of Contents Safety Precautions ......... i Safety Symbols .
  • Page 8: Chapter 1 Introduction

    The Bird 4044 Series Non-Directional Power Sensors are analog power sensors designed for use with the Bird Channel Power Monitor (CPM). The 4044 Series power sensors are typically used to measure forward average power on a 50 ohm RF transmission line with a maximum 100 watt forward power level in a frequency band with an overall range of 144 to 960 MHz.
  • Page 9: Chapter 2 Installation

    Chapter 2 Installation The Bird 4044 Series Non-Directional Power Sensors are designed to be used for continuous power monitoring in multiple Land Mobile Radio (LMR) systems. When used with the Bird CPM the 4044 can provide real-time RF power measurements.
  • Page 10: Power Sensor Installation

    Non-Directional Power Sensor Power Sensor Installation This procedure is an example of one possible cable configuration, based on the Connection Diagram WARNING Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being applied. Leaking RF energy is a potential health hazard.
  • Page 11: Chapter 3 Maintenance

    Chapter 3 Maintenance There is no user maintenance for the 4044 power sensor, contact the Bird Service Center in the event the power sensor requires maintenance. Customer Service Any maintenance or service procedure beyond the scope of those in this chapter should be referred to a qualified service center.
  • Page 12: Specifications

    Non-Directional Power Sensor Specifications Specification Value Power Requirements Input Voltage 15 VDC (supplied by CPM via RJ25 Jack) Operating Power < 5 mA Operating Characteristics 144 to 960 MHz, "Model Identification" on page 7 Frequency Range for specific frequency ranges of each model. Average Forward Power 100 W Max.
  • Page 13: Mismatch Uncertainty

    Maintenance Mismatch Uncertainty Mismatch uncertainty is a factor that can affect measurement accuracy. When a source and an antenna have different impedances, some signal will be reflected back to the source. The non-directional power sensor does not differentiate between forward and reflected power, so any mismatch will impact the measurement.
  • Page 14: Model Identification

    Non-Directional Power Sensor Model Identification...
  • Page 15: Limited Warranty

    Limited Warranty All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one (1) year, unless otherwise specified, from date of shipment and to conform to applicable specifications, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be other than as warranted;...