ESR-Auto Plus
RAM Reset Procedure
For Model 504-V9,V10 and Model 505-V3
Sep 2014, 880004-4
Note: Performing a RAM reset will clear all existing patient and QC data. If possible, print the
patient log file and QC data before proceeding with a RAM reset.
1. Unplug the instrument (Model 504) or power off the instrument using the ON/OFF switch (Model 505)
and wait at least 5 seconds. While holding down the red
or power on the instrument. The following instrument display readings may differ slightly, depending
on the software version.
2. Instrument reads:
3. Instrument reads:
4. Instrument reads:
5. Instrument reads:
6. Instrument reads:
7. Instrument reads:
8. Instrument reads:
9. Instrument reads:
10. Instrument reads:
11. Instrument reads:
12. Instrument reads:
13. Instrument reads:
Page 1 of 4
"Setup, Reset System?"
'X'
Release the red
button and press the green
"Reset All to Default! Sure?"
''
Press the green
button.
"RAM Cleared Restart System."
Unplug or turn off the instrument and wait at least 5 seconds.
Plug in or turn on the instrument without holding the red
The software version will display, followed by "New Software, [OK] for Setup?"
''
Press the green
button.
"Motor Setup, Please Wait..."
The instrument will cycle and print out the system settings.
This step may take a couple of minutes.
"IR Auto Set, Remove Tubes [OK]"
Remove all tubes/test rack and press the green
"Calibrating..."
The instrument will print some settings.
"RAM Now Reset to Default. [OK]"
''
Press the green
button.
"Setup User Config?"
''
Press the green
button.
"Set Date & Time?"
''
Press the green
button. Default date format is YY-MM-DD.
''
Press the green
button to accept the date information on the display.
'X'
Press the red
button to change the date.
''
Press the green
button to advance to the time menu.
''
Press the green
button to accept the time information on the display.
'X'
Press the red
button to change the time.
''
Press the green
button to advance to the next prompt.
"Set ESR Method?"
''
Press the green
button.
"Set Tube Type?"
''
Press the green
button.
'X'
button on the instrument keypad, plug in
''
button.
'X'
button.
''
button.

Advertisement

loading
Need help?

Need help?

Do you have a question about the ESR-Auto Plus 504-V9 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Streck ESR-Auto Plus 504-V9

  • Page 1 ® ESR-Auto Plus RAM Reset Procedure For Model 504-V9,V10 and Model 505-V3 Sep 2014, 880004-4 Note: Performing a RAM reset will clear all existing patient and QC data. If possible, print the patient log file and QC data before proceeding with a RAM reset. 1.
  • Page 2 ® ESR-Auto Plus RAM Reset Procedure For Model 504-V9,V10 and Model 505-V3 Sep 2014, 880004-4 14. Instrument reads: “1.2ml (PAED)?” The available tube settings are “1.2ml” only, “2.0ml” only or “Both 1.2ml & 2.0ml”. ‘X’ Press the red button to scroll through the available options. ‘’...
  • Page 3 ® ESR-Auto Plus RAM Reset Procedure For Model 504-V9,V10 and Model 505-V3 Sep 2014, 880004-4 28. Instrument reads: “See Sys Info?” ‘X’ Press the red button. 29. Instrument reads: “Service Menu?” ‘’ Press the green button. 30. Instrument reads: “Restricted Entry, Give Code.” ‘’...
  • Page 4 ® ESR-Auto Plus RAM Reset Procedure For Model 504-V9,V10 and Model 505-V3 Sep 2014, 880004-4 42. Instrument reads: “Printouts?” ‘X’ If instrument is not connected to the LIS, press the red button. ‘’ When transmitting data to an LIS, press the green button.

This manual is also suitable for:

Esr-auto plus 504-v10Esr-auto plusmodel 505-v3