Control Buttons • [Y/+] always controls left-hand response & increases numerical values • [N/-] controls right-hand response & moves to next digit • [MODE] controls center & returns to previous menu...
Page 6
Start-up: Turning On • Press and hold [MODE] • When display appears, release [MODE] key • Instrument performs self- test • Final display shows and unit is ready for zero calibration...
Page 7
Start-up: Programmable Zero Calibration • Unit performs Zero calibration after warm-up • “apply zero gas…” • If in fresh air, the UltraRAE can be calibrated, or a Zeroing kit with charcoal tube can be used...
Page 8
• To ensure that there are no leaks, periodically test the seals. • With the UltraRAE 3000 running, place your finger over the end of the inlet probe. • The alarm should sound and the pump- stall icon should flash on the display.
Page 9
Test Pump • If the pump does not alarm or show the stalled-pump icon, then check that all inlet parts are tight and inspect the O-ring for damage (replace it if necessary). • After inspection and/or replacing, test again. • If there is still no alarm, send instrument in for pump service.
Page 10
Start-up: Programmable Zero calibration...
Page 11
2 minutes – Wait 1 2 minutes • If “Lamp” message remains, turn off UltraRAE 3000 and restart – “Lamp” alarm clears and UltraRAE 3000 is ready for use • If after restart, “Lamp” message remains, the PID needs service...
Page 12
Start-up: Humidity Check • It is important to check if the instrument responds to moisture Cup hand over inlet or breathe into inlet for 10-20 seconds Do not block flow • If UltraRAE reads >2 ppm, then the sensor needs cleaning...
Page 13
Compound Specific vs VOC Monitor Two main operating modes: • Compound Specific Monitor: The factory default mode. After warm-up, the meter takes timed samples for Benzene or Butadiene • VOC Monitor: After warm-up, the meter works similar to the MiniRAE 3000 –...
Page 14
Compound Specific Monitor • Turn unit on and wait for warm-up to finish. • Press [N/-] to advance. • Before you start sampling, you must insert a RAE-Sep separation tube into the inlet/holder.
Page 15
Separation Tube Preparation • Open a package of RAE-Sep separation tubes and remove one. • Place the tip in the package’s tube tip breaker (the small hole on the front) and snap off the tip.
Page 16
Separation Tube Preparation • Turn the tube around and snap off the other end. • Handle tubes with care. Tube ends are sharp after ends are broken off.
Page 17
Inserting Separation Tube • Unscrew the front of the sampling probe from the base. • Slip the tube into the rubber holder in the front portion.
Page 18
Inserting Separation Tube • Insert the other end of the tube into the middle of the base while turning the front portion to tighten it onto the base’s threads. • Do not overtighten any portion of the sampling assembly.
Page 19
Measuring Compound Specific • Once the tube is in place, begin measuring by pressing [Y/+]. • The display shows a countdown – 60 seconds is standard, but sampling time depends on the type of separation tube selected and the temperature. •...
Page 20
Measuring Compound Specific • Once the countdown is complete, the gas reading is shown. • Press [Y/+] to continue sampling with the tube for 15 minutes to establish a STEL reading, or press [N/-] to return to the main menu...
Page 21
If not, the STEL value is not valid. • Abort the measurement and change the tube. Then do a snapshot test instead of a STEL test. • Note: If the STEL is exceeded, the UltraRAE 3000 goes into alarm.
Page 22
Measuring Compound Specific • Press [N/-] to advance to the next screen. • If you press [Y/+], you are asked, “Clear peak value! – Are You Sure?” • Press [Y/+] to clear the Peak value and exit to VOC operation.
Page 23
Measuring Compound Specific • If you press [N/-], this display is shown. • Remove the tube and put the inlet back together. Then press [N/-]. • After a few seconds, the UltraRAE 3000 enters VOC mode.
Page 24
VOC Mode • You can step through the rest of the screens by pressing [N/-] repeatedly until you reach the main menu again.
Page 25
VOC Mode • The instrument is programmed to operate in Basic User Level/Hygiene Mode as its default. • This gives you the most commonly needed features while requiring the fewest parameter adjustments. • Pressing [N/-] steps you from one screen to the next, and eventually return to the main display.
Page 26
VOC Monitor: Hygiene/Search Mode Two main operating modes: • Hygiene Mode: The factory default mode. After warm-up, the meter samples continuously. • Search Mode: After warm-up. the pump shuts off and “Ready” is displayed.
Page 27
Hygiene Mode/Basic User *Dashed (- - -) line is automatic return...
Page 28
Hygiene Mode/Basic User Screens • Current readings • TWA/STEL/PEAK • Date/Time/Temperature • Calibration Gas/Measurement Gas/Correction Factor • Enter PC Communications...
Page 29
Hygiene Mode/Advanced User Differences • Change reference gas without entering Programming Mode • Change measurement gas when reference gas is Isobutylene...
Hygiene Mode • Reading in ppm • Gas information • Calibration needed • Wireless status • Battery status • Datalog • Pump Status • Press [N/-] to move to next screen...
Page 31
TWA/STEL/PEAK • TWA reading • STEL reading • Peak reading • Press [Y/+] to clear • Press [N/-] to move to next screen...
Page 32
Date/Time/Temp • Date • Time • Internal temperature • Press [N/-] to move to next screen...
Page 33
Gas Display • Calibration gas • Measurement gas • Correction factor • Press [N/-] to move to next screen...
Page 34
PC Communication • Enter PC communication and stop sampling • Press [Y/+] to enter PC Comm mode – Press [Y/+] to exit after communication is complete • Press [N/-] to move to next screen...
Page 36
Search Mode • A discrete sampling mode can easily start/stop datalogging for many points • Perfect for drum or headspace sampling • Optional graphic screen...
Page 37
Optional Graphic Screen • Using ProRAE Studio, set display to graphic display • Leave display in ppm reading...
Page 38
Search Mode/Basic User *Dashed (- - -) line is automatic return...
Page 39
Search Mode/Basic User Screens • Ready?... Start Sampling • User ID/Site ID/Correction Factor • Current gas readings • Stop sampling? • AVG/PEAK • Date/Time/Temperature • Calibration Gas/Measurement Gas/Correction Factor • Enter PC Communications...
Page 40
Search Mode/Advanced User Differences • Change reference gas without entering Programming Mode • Change measurement gas when reference gas is Isobutylene...
Programming Mode • Navigate using interface of “Select,” “Back,” and “Next” • Press [MODE] to exit Programming • [Y/+] to increase numerical values or confirm a question • [N/-] to decrease numerical values...
Page 42
Programming Mode • Press and hold [N/-] and [MODE] to enter Programming • Input password “0000”...
Page 43
Calibration • Press [Y/+] to select Calibration menu and Calibration type • Press [N/-] to toggle between Zero calibration and Span calibration • Press [MODE] to exit Calibration menu...
Page 44
Two-point vs Three-point Calibration • Three-point calibration can be added for enhanced accuracy • Feature is added through ProRAE Studio • Second, higher concentration of calibration gas is required • After calibrating Span Gas 1, display moves to Span Gas 2...
Page 45
Calibration • You need: • Zero-grade air bottle or zeroing calibration • Bottle of 100ppm Isobutylene with .5L regulator and/or Tedlar bag, or matched flow regulator (VOC Monitor) • Bottle of 5ppm Benzene with .5L regulator and/or Tedlar bag, or matched flow regulator (Benzene tube)
Page 46
Matched Flow Calibration For best accuracy, a matched flow calibration is required! • Use matched flow regulator or Tedlar bag • Fill Tedlar bag with calibration gas and then draw down with UltraRAE 3000...
Page 47
Zero calibration • Press [Y/+] to select Zero Calibration • Press [MODE] to return to main display • Press [N/-] to select Span Calibration...
Zero calibration • If in fresh air, the UltraRAE 3000 can be calibrated or a Zeroing kit with charcoal tube can be used • Press [Y/+] to start calibration...
Page 49
Span Calibration • After Zero Calibration, menu returns to calibration selection with Span Calibration highlighted • Press [Y/+] to enter Span calibration. • Press [N/-] to skip Span calibration and return to Zero calibration. • Press [MODE] to exit and return to the top calibration menu.
Span Calibration • Turn on your span calibration gas. • Press [Y/+] to initiate calibration. • Press [N/-] to abort calibration.
Page 51
Span Calibration • Calibration is complete when instrument reads: Span 1 is done! Reading = xxx. ppm • Instrument then moves to request three- point calibration • Select [Y/+] to continue, [N/-] to skip • Press [MODE] to return to normal operations.
Page 52
Span Calibration • Calibration is complete when instrument reads: Span 2 is done! Reading = xxx. ppm • Press [MODE] to return to normal operations...
Page 53
Measurement Gas and Unit • Measurement gas has 4 lists – My list: maximum of 10, starting with Isobutylene – Last Ten: last ten used by instrument – Gas Library: all gases in TN-106 – Custom Gases: user modified via ProRAE Sudio...
Page 54
Measurement Lists • Press [N/-] to scroll through list • Press [Y/+] to select...
Page 55
Measurement Lists • Once list is selected, – Press [N/-] to scroll through options – Press [Y/+] to select option • Press [MODE] to exit • Press [Y/+] to save selection or [N/-] not to save...
Measurement List Abbreviation Unit MiniRAE ppbRAE 3000 3000 parts per million parts per billion mg/m3 milligrams per cubic meter ug/m3 micrograms per cubic meter...
Tube Selection • Scroll through the menu by pressing [N/-]. • Press [Y/+] to make a selection. • Press [MODE] when you are done. • Press [Y/+] to save your choice or [N/-] to undo your selection.
Alarm setting • Press [Y/+] to change High, Low, STEL, and/or TWA alarm • Scroll through the menu with [N/-] • Press [Y/+] to select alarm to change...
Page 59
Alarm setting • Cursor flashes over digits of alarm limit • Press [Y/+] to increase each digit value • Advance to the next number with [N/-] • Press [MODE] when done – Press [Y/+] to accept change – Press [N/-] to cancel change...
Alarm type • Latched alarm – although alarm condition might have cleared, [Y/+] must be pressed to acknowledge alarm • Automatic reset – once condition has cleared, alarm will stop (default setting)
Buzzer & Light • Program light and buzzer – Both on – Light only – Buzzer only – Both off • Press [N/-] to go to next option, Press [Y/+] to select and [MODE] to exit...
Page 62
Datalog • Press [Y/+] to enter Datalog menu – Clear Data – Interval – Data Selection – Datalog Type • Press [N/-] to move to next menu, and [MODE] to return to normal operations...
Clear Datalog • Erases all data stored in instrument • Press [Y/+] to clear data – Display asks “Are you sure?” • Press [Y/+] to confirm • Press [N/-] to keep data...
Page 64
Interval • Intervals are shown in seconds • Default is 60 seconds • Maximum is 3600 – Press [Y/+] to increase each digit – Press [N/-] to move to the next digit • Press [MODE] when done • Press [Y/+] to save and [N/-] to cancel changes...
Data Selection • Choose type of data to store – Average – Maximum – Minimum • Press [N/-] to move from one selection to next • Press [Y/+] to choose data, “X” shows which data is stored • Press [MODE] when done •...
Datalog Type Three Datalog Types • Automatic – Default mode, unit datalogs while running • Manual – unit datalog only when datalog is manualy turned on • Snapshot – datalogs only during single-event capture sampling...
Page 67
Datalog Type • To make changes to Datalog type – Press [N/-] to move to next option – Press [Y/+] to accept change • When done, press [MODE] • Press [Y/+] to save changes • Press [N/-] to discard changes...
Monitor Setup • Temperature Unit • Op Mode • Site ID • User ID • User Mode • Date • Time • Duty Cycle...
Page 69
Monitor Setup • Pump Speed • Language • Radio Power • Real Time Protocol • Power On Zero • Unit ID • LCD Contrast • Lamp ID...
Page 70
Radio Power • Radio connection can be turned on or off • Press [N/-] to highlight selection • Press [Y/+] to select radio setting • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Real Time Protocol • P2M – cable – Point to multipoint. – Data is transferred from the instrument to multiple locations using a wired connection • P2P – cable – Point to point. (default) – Data is transferred only between the instrument and one other location, such as a computer •...
Page 72
Real Time Protocol • Press [N/-] to highlight selection • Press [Y/+] to select Real time protocol setting • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Page 73
Op Mode • Switch between Hygiene or Search Mode by pressing [N/-] to highlight the option • Press [Y/+] to select mode • Press [MODE] to accept change • Press [Y/+] to save and [N/-] to discard change...
Page 74
Site ID • Enter 8-digit alpha/numeric/character Site ID • Press [Y/+] to display current Site ID • Press [Y/+] to change the number/ letter/ character – Note: the last four digits must be numbers...
Page 75
Site ID • Press [N/-] to move to the next digit • Press [MODE] to exit • If there has been a change, [Y/+] to save • [N/-] to discard any changes...
Page 76
User ID • Enter 8-digit alpha/numeric User ID • Press [Y/+] to display current User ID • Press [Y/+] to change the number/letter...
Page 77
User ID • Press [N/-] to move to the next digit • Press [MODE] to exit • If there has been a change, [Y/+] to save • [N/-] to discard any changes...
Page 78
User Mode • Basic Mode – User can only see and use a basic set of functions • Advanced Mode – Advanced users can see and use all screens and perform all available functions...
Page 79
User Mode • Press [N/-] to change option • Highlight will move to next selection each time [N/-] is pressed • Press [Y/+] to make selection • Press [MODE] to choose Basic or Advanced Mode • [Y/+] to accept change, [N/-] to discard change...
Page 80
Date • Date is expressed as Month/Day/Year, with two digits for each. • Press [Y/+] to display current date. • Note that the left-most digit flashes to indicate it is selected.
Page 81
Date • Press [Y/+] to change the number • Press [N/-] to move to the next digit • Press [MODE] once the date has been changed completely • [Y/+] to accept change, [N/-] to discard change...
Page 82
Time • Time is expressed as Hours/Minutes/Seconds, with two digits for each. • Time is in 24-hour (Military) format. • Press [Y/+] to display current time. • Note that the left-most digit flashes to indicate it is selected.
Page 83
Time • Press [Y/+] to change the number • Press [N/-] to move to the next digit • Press [MODE] once the time has been changed completely • [Y/+] to accept change, [N/-] to discard change...
Duty Cycle • The pump’s duty cycle is the ratio of its “on” time to ‘off’ time. • The duty cycle ranges from 50% to 100% (100% is default), and the period is 10 seconds. • Therefore, a duty cycle of 60% means that the pump is on for 6 seconds and off for four seconds and 100% is always on.
Page 85
Duty Cycle Important! Pump duty cycling is interrupted when the instrument senses a gas. • The pump’s duty cycle is disabled when the measurement is greater than the 2ppm threshold and is re-enabled when the reading falls below 90% of the threshold (1.8 ppm).
Page 86
Duty Cycle • Press [Y/+] to increase value • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Temperature Unit • Choose Temperature measurement in Fahrenheit or Celsius units • Press [N/-] to highlight selection • Press [Y/+] to select temperature unit • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Pump Speed • Choose high or low pump speed – Low speed conserves battery and is quieter • Press [N/-] to highlight selection • Press [Y/+] to select pump speed • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Page 89
Language • English is default language; 12 languages are available • Press [N/-] to highlight selection • Press [Y/+] to select Language • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Page 90
Power On Zero • Program instrument to request a Zero calibration on start-up • Press [N/-] to highlight selection • Press [Y/+] to select Power On Zero • When done, press [MODE] • [Y/+] to accept change, [N/-] to discard change...
Page 91
Unit ID • Three-digit number to separate wireless units on the same network – Units must have individual ID values • Press [Y/+] to change the number • Press [N/-] to move to the next digit • Press [MODE] once the Unit ID has been changed completely •...
LCD Contrast • Increase or decrease contrast of display – Minimum value is 20 – Maximum value is 60 • Press [Y/+] to increase, [N/-] to decrease the value • Press [MODE] once the contrast has been changed • [Y/+] to accept change, [N/-] to discard change...
Page 93
Lamp ID • Scroll through the menu by pressing [N/-]. • Press [Y/+] to make a selection. • Press [MODE] when you are done. • Press [Y/+] to save your choice or [N/-] to undo your selection.
Page 94
Monitor Set up • At any time, press [MODE] to return to normal operations • When inside a menu selection, press [MODE] to move to the top menu...
Power Off • Hold [MODE] for full 3 seconds • Audible alarm will beep and unit will countdown 5 seconds • Release [MODE] when you see “Unit Off…” • Unit will stay on if you release [MODE] before countdown completes...
Page 96
Humidity Filtering II Tubes Tube Tip Breaker Flex-I-Probe (p/n 021-2400-100) Tube Adapter Humidity Filtering II Tube (p/n 025-3002-000) (10-pack, p/n 025-2002-010) • Temporary relief for a dirty sensor • Dries sample gas for about ½ hour • Measure VOCs; multiple sample use OK •...
Page 97
Charging on Charging Cradle • To Charge unit – Push instrument down on charging cradle – Set back of unit against back wall of cradle • To remove unit from cradle – Push down on instrument – Pull unit towards you...
Page 98
Charging Battery on Travel charger • Line the back edge of Travel Charger with the back edge of unit • Carefully screw both plastic pins of charger into instrument • Plug wall adapter into travel charger once charger is fully attached to instrument...
Page 99
Charging Error • If the instrument or battery has been charging for more than 10 hours and you see the “Battery Charging Error” icon and a message that says, “Charging Too Long,” this indicates that the battery is not reaching a full charge. •...
Removing battery • Rubber boot stays on monitor • First push down on • Then pull battery from instrument...
Page 101
Charging Battery on Charging Cradle • Use back of charging station to charge battery externally – Push into contacts – Lay battery flat • To remove battery – Push battery towards station – Lift battery off of charging station...
Maintenance • Clean PID Lamp & Sensor – When display creeps upwards after good zero – When PID responds to moisture – When movement of PID results in response on display Clean Sensor Dirty Sensor Bias Electrode Bias Electrode Sensing Electrode Sensing Electrode No dirt build-up to Dirt build-up absorbs water and...
Page 103
Maintenance • How to clean the PID lamp – Use anhydrous methanol (lamp cleaning solution) – Use finger cots to hold lamp – Dip cotton swap in methanol and clean flat top surface of lamp – Dry lamp face with lens tissue...
Page 104
Maintenance • How to Clean PID Sensor – Always clean sample probe and replace or clean filters FIRST! If PID holds a stable zero after this step then further cleaning may not be necessary – Use anhydrous methanol (lamp cleaning solution) –...
Page 105
Configuration from Computer Datalogging Using ProRAE Studio...