3.9 OPTIONAL SERIAL OUTPUT

In case of an additional serial board is installed in the indicator the
connections are according the following drawings

COM5 serial connection

DSUB9 9-pin female connector.
PIN
1
2
3
4
5
6
7
8
9
NC = do not connect
ATTENTION
Max. conditions of usage under RS232 standard:
Max. transmission distance: 15 m
Maximum voltage at ends: ± 12 VDC
For connection to external devices, it is recommended to use a shielded cable, making sure to
connect the shield to the metal part of the 9-pin connector's casing.
If the COM5 serial port is configured in compliance with the RS422 standard, proceed as follows:
provide a jumper between pin 6 and pin 8 to connect the termination resistor.
ATTENTION
Max. conditions of usage under RS422 standard:
Max. transmission distance: 1200 m
Maximum voltage at ends: +/- 7V
For connection to external devices, it is recommended to use a twisted pair, making sure to
connect the shield to the metal part of the 9-pin connector's casing.
1-32
DESCRIPTION
RX422 -
RX232
TX232
NC
GND
RX422 +
TX422 -
TERMINATOR
TX422 +
DD2050