SAMWONTECH ※ This manual applies to TEMI2300, TEMI2500 and TEMI2700. ※ The model stated the manual content is TEMI2500. Contents Safety Precautions Handling overview Initial operation flow ………………………………………………………………………………… Setting Buttons ………………………………………………………………………………………· Input Keypad ………………………………………………………………………………………· OPERATION setting MAIN MENU screen …………………………………………………………………………………...
Page 3
SAMWONTECH SCREEN DISPLAY setting Screen display setting ……………………………………………………………………………… DI ERROR HISTORY ………………………………………………………………………………· RESERVE OPERATION setting …………………………………………………………………… USER SCREEN setting USER SCREEN viewer ……………………………………………………………………………… Creating BMP File for USER SCREEN image …………………………………………………· Communication Errors massage …………………………………………………………………· * Engineering Units 4st Edition of TEMI2000_Series IM : Aug. 18. 2009...
Read and understand this Operation Manual carefully before using the product. This Operation Manual describes functions and features of the product in detail, and SAMWONTECH can not guarantee against over applications would suit a customer’s particular purpose which is not described in this manual.
Page 5
SAMWONTECH does not guarantee or accept responsibility for this product other than the clauses stated in our warranty policy. SAMWONTECH assumes no liability to any party for any loss or damage, direct or indirect, caused by the use or any unpredictable defect of the product.
SAMWONTECH 2 Handling overview ▶ This product is programmable controller for temperature and humidity, designed with Touch Screen interface for user friendly easy access. 2.1 Initial operation flow ▶ Supplying the power after completing installation correctly, loading screen will appear for loading time then Initial Logo screen will be displayed.
SAMWONTECH 2.2 Setting Buttons ▶ [Table2-1] describes Basic Setting Buttons Table 2-1. Basic Setting Buttons Button Type Description Button to set target Set Point of temperature and pattern number to run, located at the upper left side of OPERATTION screen.
SAMWONTECH 2.3 Input Keypad ▶ When you pressing the button, the following Input Keypad appears and the required data value can be set by using this Input Keypad. ▶ If the setting value is out of the available range, ‘LIMIT ERROR’ message will appear with a ‘BEEP’...
Page 9
SAMWONTECH KEY LOCK Release ▶ Setting operation through Input Keypad can be performed after releasing KEY LOCK function switch ON to OFF ☞ For further information, refer to [4. FUNCTION & FIX OPERATION setting]. 2.3.1 Setting input validity test ▶ TEMI2500 confirms key input process and its validity of value by sound.
Page 10
SAMWONTECH Ex.) How to input Set Point ▶ The following describes how to change Set Point 49.40 to 12.5. → Press the button to pop up Numeric Input Keypad. → Press the numbers by the order (①→②→③→④→⑤). If setting is OK, press ENTER key(⑥).
SAMWONTECH 3. OPERATION setting 3.1 MAIN MENU screen ▶ MAIN MENU screen is organized into six sections. This MAIN MENU screen is the central gate way screen to move into lower parameter section. ▶ Press the (MAIN) button to move to this MAIN MENU screen ①...
SAMWONTECH 3.2 FIX Operation 3.2.1 FIX STOP screen ▶ Under setting ‘FIX’ operation for ‘OPERATION MODE’ on [4. FUNCTION & FIX OPERATION setting], when you press [OPER. SCREEN] button on the MAIN MENU screen, this FIX STOP screen will appear.
Page 13
SAMWONTECH ▶ Target Set Point of humidity for FIX operation can be set through below. [Figure 3-4] Input Keypad for humidity Set Point ▶ Press the blue button on the left lowerhand corner to pop up numeric Input Keypad [Figure 3-4].
Page 14
SAMWONTECH 3.2.2 The 1 FIX RUN screen ⑥ ① ② ⑦ ⑧ ③ ⑨ ④ ⑩ ⑤ ⑪ ⑫ [Figure 3-5] The 1 FIX RUN screen ① Displays current operation status (FIX operation RUN) ② Displays and Setting button for target Set Point of temperature.
Page 15
SAMWONTECH 3.2.3 The 2 FIX RUN screen ▶ Additional information of FIX operation status will be displayed in this screen, including PV, SP, control output MV (Manipulate Value) and activating state of other Digital Outputs. ③ ① ④ ② ⑤...
Page 16
SAMWONTECH 3.2.4 The 3 FIX Operations screen ▶ This screen is for viewing and recording data by trend graph. ▶ You can set items to display data by trend graph among 6 parameters – SP, PV, MV of temperature and humidity – shown at the left hand side of this screen by using each check box ( ▶...
Page 17
SAMWONTECH 3.2.5 FIX operation STOP screen ▶ Once the operation has completed normally as the time at ‘TIME OPERATION’ on [4. FUNCTION & FIX OPERATION setting], FIX operation will stop and ‘TIME OPERATION END’ message appears shown as [Figure3-8] ▶ However, this message will not appear if the operation is stopped by force pressing ‘STOP’ button.
SAMWONTECH 3.3 PROGRAM Operation 3.3.1 PROGRAM STOP screen ▶ Under setting ‘PROG’ operation at ‘OPERATION MODE’ on [4. FUNCTION & FIX OPERATION setting], when you press [OPER. SCREEN] section on the MAIN MENU screen, this PROGRAM STOP screen will appear.
Page 19
SAMWONTECH 3.3.2 The 1 PROGRAM RUN screen ⑥ ① ② ⑦ ⑧ ③ ⑨ ⑩ ⑪ ④ ⑫ ⑤ ⑬ ⑭ [Figure 3-10] The 1 PROGRAM RUN Screen ① Displays current operation status (PROG operation RUN) ② Displays current Set Point of temperature that will be reaching the target SP of current segment.
Page 20
SAMWONTECH 3.3.3 The 2 PROGRAM RUN Screen ▶ Additional information of PROGRAM operation status will be displayed in this screen, including PV, SP, control output MV (Manipulate Value) and activating state of other Digital Outputs. ① ⑤ ② ⑥ ③...
Page 21
SAMWONTECH 3.3.4 The 3 PROGRAM RUN Screen ▶ This screen is for viewing trend graph and recording and saving the data. ▶ You can set items to display by graph and record trend data among 6 parameters – SP, PV, MV of temperature and humidity–shown at the left hand side of this screen by using each check box (...
Page 22
SAMWONTECH 3.3.5 PROGRAM operation STOP Screen ▶ Once the operation has completed normally as all setting on processed pattern, PROGRAM operation will stop and ‘PATTERM OPERATION END’ message appears show as [Figure3-13] ▶ However, this message will not appear if the operation is stopped by force pressing ‘STOP’ button.
SAMWONTECH 3.4 AUTO TUNING and TUNING POINT ▶ The Hold and Step keys cannot be used while the program is in running with auto-tuning. ▶ Auto-Tuning is a strong function to establish optimal P.I.D value automatically by calculating the characteristics of the control system.
SAMWONTECH 4. FUNCTION & FIX OPEARTION setting ▶ Press [FUNCTION&FIX] button on MAIN MENU screen to enter into this function section. ▶ This screen is to set parameters concerned additional function for general operation and FIX operation. 4.1 Operation mode & Functions ①...
Page 25
SAMWONTECH Table 4-1. PROGRAM STOP Screen Parameters Parameters Range Units Default OPERATION MODE PROG, FIX Pattern POWER STOP STOP, COLD, HOT STOP MODE FUZZY SELECT OFF, ON Non-function KEY LOCK OFF, ON TEMP. SP SLOPE T.EUS(0.00 ~ 100.00%) / MIN T.EUS / MIN...
SAMWONTECH 4.2 FUZZY function ▶ FUZZY function can effectively control the overshoot that may occur when intense fluctuating load or frequent Set Point changing ▶ Before PV approaches to SP, automatically calculated SUPER SP can restrain overshoot. ▪ FUZZY ‘OFF’...
SAMWONTECH 4.3 SP SLOPE function ▶ SP SLOPE function to make the Set Point ascend or descent gradually overtime. ▶ When changing the Target SP during FIX run operation, NSP (current SP) will be changed from current PV to Target SP gradually by assigned ramping rate ▶...
SAMWONTECH 5. PROGRAM SETTING ▶ Press [PROGRAM SET] on MAIN MENU screen to enter into program setting section [Figure 5-1 PROGRAM SET screen]. ▶ This PROGRAM SET menu is composed with 6 groups concerned program operation setting. ① ③ ②...
SAMWONTECH 5.1 Program PATTERN setting ▶ This screen is to set programming segments of each pattern. ▶ Press [PATTERN SET] button on PROGRAM SET menu to enter into this screen ▶ Refer to [5.4 TIME SIGNAL] for how to use TIME SIGNAL.
Page 30
SAMWONTECH ▶ Press the setup box, and Input Keypad will appear. ▶ Press the setup box, and Option Keypad will appear. ▶ Screen when ‘TEMP SP’ setting of START CODE. Setup box for Start Set Point (SSP) of temperature 4st Edition of TEMI2000_Series IM : Aug. 18. 2009...
Page 31
SAMWONTECH ▶ Screen when ‘HUMI SP’ setting of START CODE. Setup box for Start Set Point (SSP) of humidity ▶ (INSERT) and (DELETE) buttons become active after activating specific segment pressing ▶ Press the (TEMP) setup box, and Input Keypad for TEMP SP will appear.
Page 32
SAMWONTECH ▶ Press the (HUMI) setup box, and Input Keypad for HUMI SP will appear. ▶ Press the (SEGMENT) setup box, and Input Keypad for TIME will appear. ▶ Press the (TIME SIGNAL) setup box, and Input Keypad for TS will appear.
Page 33
SAMWONTECH ☞ Press (Enter) button to complete time signal setting. ☞ Press (ESC) button to cancel input and disappear Input Keypad. ※ Four TIME SIGNALs, TS1 to TS4 can be set among 18 types from [5.4 TIME SIGNAL] by pressing TS1~TS4 button.
SAMWONTECH 5.2 REPEAT function ▶ Press [REPEAT SET] button on PROGRAM SET menu screen to enter into this screen. ▶ This screen is to set repeat operation of entire current pattern and partial segment operation after complete of current operation.
SAMWONTECH 5.3 Pattern FILE EDIT ▶ Press [FILE EDIT] button on PROGRAM SET menu screen to enter into this screen. ▶ This screen is to edit program by copy or delete pattern set on [5.1 Program PATTERN]. ▶ Currently processing pattern can not be deleted.
Page 36
SAMWONTECH ▶ ‘PARAMETER SETTING SET ERROR’ message will appear when editing with input error at original pattern number. [Figure 5-5] FILE EDIT screen - Error ▶ ‘THE SELECTED PATTERN IS EMPTY’ message will appear with no segment information at original pattern.
SAMWONTECH 5.4 TIME SIGNAL ▶ Press [TIME SIGNAL] button on PROTRAM SET menu screen to enter into this screen. ▶ This screen is to establish each 18 types of TIME SIGNAL setting. Refer to the [5.4.3 Description of TIME SIGNAL operation] ▶...
Page 38
SAMWONTECH ① ② [Figure 5-9] TIME SIGNAL SET screen – TS12 to TS17 ① DEALY TIME is to set the start time of TIME SIGNAL activating ‘ON’ at segment. TS event output will become ‘ON’ right after the DELAY TIME elapses.
Page 39
SAMWONTECH 5.4.3 Description of TIME SIGNAL operation Setting TIME SIGNAL Operation OPER.TIME 1. DELAY TIME = 000.00.00 TIME SIGNAL 2 N SEG TIME≥ SEGMENT (N-1) SEG TIME N SEG TIME (N+1) SEG TIME DELAY TIME +OPER.TIME OPER.TIME DELAY TIME 2. DELAY TIME ≠...
SAMWONTECH 5.5 WAIT function ▶ Press [WAIT SET] button on PROTRAM SET menu screen to enter into this screen. ▶ This screen is to set WAIT Function parameter. ▶ WAIT Function established on this screen can be applied to program segment setting on [5.1 Program PATTERN].
Page 41
SAMWONTECH ▶ The interaction between WAIT Operation and WAIT TIME is shown as below. ① Release WAIT Operation within WAIT TIME WAIT TIME Set point WAIT ZONE WAIT ZONE Release WAIT Operation Step to the next Segment WAIT Operation (n+1) progression...
SAMWONTECH 5.6 PATTERN NAME input ▶ Press [PATTERN NAME] button on PROTRAM SET menu screen to enter into this screen. ▶ This screen is to input the experiment name of each pattern. ▶ Refer to [Figure 3-10 The 1 PROGRAM RUN Screen].
SAMWONTECH 6. GRAPH & REC. setting 6.1 PATTERN GRAPH VIEW ▶ The following screens show the temperature/humidity pattern profile by graph when program operation. ▶ Press [GRAPH&REC.] on MAIN MENU screen to enter into this trend graph and data logging section, and [Figure 6-2 PATTERN GRAPH VIEW-Profile] will be displayed.
Page 44
SAMWONTECH ③ Check box to display pattern SP profile of humidity (blue color). If checking , Humidity pattern SP profile will be display by trend graph. ☞ Even during processing operation, Check box buttons are available. ④ Setup box to input the pattern number to display by trend graph ☞...
Page 45
SAMWONTECH ▶ Input Keypad for pattern number to display by trend graph. [Figure 6-4 Pattern number input keypad for graph] Table 6-1. PATTERN GRAPH VIEW setting Parameters Parameter Range Units Default PATTERN NUMBER 1 ~ 120 30 mins, 1 hour, 3 hours, 6 hours,...
SAMWONTECH 6.2 TREND GRAPH view & DATA processing ▶ This screen is to open specific recorded trend data file at the [3.3.4 The 3 PROGRAM RUN screen] or [3.2.4 The 3 FIX RUN screen] and show that trend data by graph.
Page 47
SAMWONTECH ▶ [Figure 6-6 Trend Graph screen – No item] is the screen when there is no checked item to display among 6 parameters. [Figure 6-6] Trend Graph screen – No item ▶ [Figure 6-7 Trend Graph screen – PV FILE list] shows how to open the trend data file in internal memory stored at [3.3.4 The 3...
SAMWONTECH 6.3 Recording TREND GRAPH on internal memory ▶ This screen is to establish parameters for recording data on [3.3.4 The 3 PROGRAM RUN screen] or [3.2.3 The 2 FIX RUN screen]. ① ② [Figure 6-3] Trend Graph setting screen ①...
SAMWONTECH 6.4 Data Transmission setting with SD memory card ▶ This screen is to set the parameters concern data logging and up / download with SD memory card. ☞ This screen will be hidden without SD card option. ① ②...
SAMWONTECH 7 SCREEN DISPLAY setting 7.1 Screen Display settings ▶ Press the DISPLAY SET on MAIN MENU screen to enter this display setting section. ▶ This screen is to set parameters concerned screen display such as character font, LED brightness, Buzzer and Backlight off time.
SAMWONTECH 7.2 DI ERROR HISTORY. ▶ This screen shows the history of date and time on DI ERROR occur. ▶ Max. recent 30 DI ERROR history recording displayed. ☞ If more than 30 errors, recent 30 data will be displayed.
Page 52
SAMWONTECH ▶ [Figure 7-3] and [Figure 7-4] are screens when occurring DI ERROR. ▶ DISPLAY METHOD for DI ERROR occur between TEXT and PICTURE can be set on [11. DI CONFIG] of the [Instruction Manual]. ▶ The (EXIT) button is used to return to the Operation screen.
SAMWONTECH 8 RESERVE OPERATION setting ▶ Press [RESERVE SET] on MAIN MENU screen to enter into this time setting section. ▶ This screen is to adjust the current time and set the time for RESERVE Operation. ③ ① ② [Figure 8-1] RESERVE SET Screen ①...
Page 54
SAMWONTECH Table 8-1. RESERVE OPERATION setting Parameters Parameter Range Units Default YEAR 2000~2099 MONTH 1~12 1~31 CURRENT TIME AM/PM AM, PM HOUR 1~12 0~59 YEAR 2000~2099 2008 MONTH 1~12 1~31 RESERVE TIME AM/PM AM, PM HOUR 1~12 0~59 RESERVE Press to activate RESERVE Operation...
SAMWONTECH 9 USER SCREEN 9.1 USER SCREEN viewer ▶ 16 individual images can be used as USER SCREEN. ▶ USER SCREEN viewer will be activated by no key input for established [START TIME] on [12. PICTURES VIEW setting] of [Installation Manual], and USER SCREEN images will display with rotating by [INTERVAL TIME], when using that function with more than one picture.
- When saving your image in [Photoshop] save as BMP. ☞ Name the folder inside the SD CARD as BMP. ※ More information on making a BMP file can be downloaded of SAMWONTECH's website. 4st Edition of TEMI2000_Series IM : Aug. 18. 2009...
Page 57
SAMWONTECH 10 COMMUNICATION ERROR message ▶ ‘CONTROL PART WAS DISCONNECTED!’ message will appear on Operation screen shown as [Figure 10-1 Communication Error with CONTROL UNIT] when communication error between the CONTROL UNIT and the DISPLAY PANEL occurs. ▶ ‘I/O BOARD WAS DISCONNECTED!’ message will appear on Operation screen shown as [Figure 10-2 Communication Error with I/O BOARD] when communication error between the I/O BOARD and the DISPLAY PANEL occurs.
Page 58
SAMWONTECH (ENGINEERING UNITS) EU and EUS are used for the scaling of the parameters of the controller. – EU, EUS ▶ If the sensor type or minimum/maximum input range (INRH, INRL) is adjusted, the EU(), EUS() parameters also change proportionally (minimum and maximum input ranges are reset).
Page 59
Gyeonggi-do, Korea 420-773 TEL : +82-32-326-9120 FAX : +82-32-326-9119 Website : http://www.samwontech.com e-mail : webmater@samwontech.com The contents of this document are subject to change without prior notice. Printed in Korea : 2008(A) All Rights Reserved. Copyright ⓒ 2008 SAMWONTECH CO.,LTD.