Professional ultrasonic nebuliser
13
1. Disassembling the nebulisation unit
Rotate the medication cup cover fixing lever
clockwise. Take the medication cup holder
ring) located below the cup cover and lift it up. In this
way, the nebulisation section is removed.
Note: Medication cup holder is a three-legged part
which keeps it steady on the table.
2. Filling the cooling water
Fill the cooling water in the water tank
amount is approximately 300ml. Pour some water
until the sensor level float in the water tank rises to its
upper position. Recommended water temperature is
approximately 25°C. To prevent scale buildup in the
water tank, use distilled or demineralised water instead
of tap water.
Sensor level float in upper position
3. Attaching the nebulisation unit
Attach the nebulisation unit to the water tank
sure that the air duct on the rear side of the nebulisation
unit is adjusted properly to the air filter cover
How to prepare before inhalation
(plastic
. The proper
Water level
Vibrator
. Be
.